Alignment Solution for CT Image Reconstruction by Fixed Point and Virtual Rotation Axis (original) (raw)
Abstract:Since X-ray tomography is now widely adopted in many different areas, it becomes more crucial to find a robust routine of handling tomographic data to get quality reconstructed images. Though there are several existing techniques, it seems helpful to have a more automated method to remove the possible errors that hinder clearer image reconstruction. Here, we proposed an alternative method and new algorithm using the sinogram and the fixed point. A new physical concept of Center of Attenuation (CA) was also introduced to figure out how this fixed point is applied to the image reconstruction with errors we further categorized. Our technique showed a promising performance in restoring images with translation and vertical tilt errors.
Submission history
From: Kyungtaek Jun [view email]
[v1] Fri, 13 May 2016 08:22:43 UTC (1,238 KB)