A cell-based biosensor for real-time detection of cardiotoxicity using lensfree imaging (original) (raw)
* Corresponding authors
a Center for Biomedical Engineering, Department of Medicine, Brigham and Women's Hospital, Harvard Medical School, Boston, MA, USA
E-mail: alik@rics.bwh.harvard.edu
b Harvard-MIT Division of Health Sciences and Technology, Massachusetts Institute of Technology, Cambridge, MA, USA
c Electrical and Computer Engineering Department, Center for High Rate Nanomanufacturing, Northeastern University, Boston, MA, USA
d U.S. Army Corps of Engineers, Construction Engineering Research Laboratory, Champaign, IL, USA
e Wyss Institute for Biologically Inspired Engineering at Harvard University, Boston, MA, USA
Abstract
A portable and cost-effective real-time cardiotoxicity biosensor was developed using a CMOS imaging module extracted from a commercially available webcam. The detection system consists of a CMOS imaging module, a white LED and a pinhole. Real-time image processing was conducted by comparing reference and live frame images. To evaluate the engineered system, the effects of two different drugs, isoprenaline and doxorubicin, on the beating rate and beat-to-beat variations of ESC-derived cardiomyocytes were measured. The detection system was used to conclude that the beat-to-beat variability increased under treatment with both isoprenaline and doxorubicin. However, the beating rates increased upon the addition of isoprenaline but decreased for cultures supplemented with doxorubicin. Moreover, the response time for both the beating rates and the beat-to-beat variability of ESC-derived cardiomyocytes under treatment of isoprenaline was shorter than for doxorubicin, although the amount of isoprenaline used in the measurement was three orders of magnitude lower than that of doxorubicin. Given its ability to perform real-time cell monitoring in a simple and inexpensive manner, the proposed system may be useful for a range of cell-based biosensing applications.
You have access to this article
Please wait while we load your content... Something went wrong. Try again?
Supplementary files
- Movie S1 MOV (1204K)
- Movie S2 MOV (1041K)
- Movie S3 MOV (768K)
- Movie S4 MOV (785K)
- Movie S5 MOV (1117K)
Article information
DOI
https://doi.org/10.1039/C1LC20098D
Article type
Paper
Submitted
03 Feb 2011
Accepted
23 Mar 2011
First published
11 Apr 2011
Download Citation
Lab Chip, 2011,11, 1801-1807
Permissions
A cell-based biosensor for real-time detection of cardiotoxicity using lensfree imaging
S. B. Kim, H. Bae, J. M. Cha, S. J. Moon, M. R. Dokmeci, D. M. Cropek and A. Khademhosseini,Lab Chip, 2011, 11, 1801DOI: 10.1039/C1LC20098D
To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.
If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.
If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.