IN the prehospital setting, the incidence of unrecognized esophageal intubation has been reported to be as high as 1.8–2.0%. 1,2In cases of cardiopulmonary arrest, the critical failure to recognize an esophageal intubation continues to pose a serious safety problem.
Acceptable methods of validation of correct tracheal intubation are direct visualization of passage of the endotracheal tube through the vocal cords, fiberoptic bronchoscopy, and detection of expired carbon dioxide. 3–6Direct visualization may not be possible in some patients, however, and fiberoptic bronchoscopy may not be available in emergency settings. Expired end-tidal carbon dioxide can be detected with a capnograph, a mass spectrometer, or a single-use end-tidal colorimetric device. 7–9Even these devices may provide spurious results. For example, for a patient in cardiopulmonary arrest there may be no characteristic color change in the colorimetric device, and a waveform may be absent in a capnograph or spectrometer trace. 10–12Because the detection of expired carbon dioxide may not possible in a patient in cardiac arrest, a major safety issue for emergency intubations is the development of a means whereby one can confirm proper placement of the endotracheal tube quickly, assuring that oxygenation and ventilation can proceed. Because an unrecognized esophageal intubation can have devastating clinical consequences, the development of a definitive diagnostic device that is capable of distinguishing unequivocally , within seconds, between an endotracheal and an esophageal intubation in a patient in cardiopulmonary arrest, is of the highest importance. Such a diagnostic device should be compact, portable, and easy to use.
A recently developed airway adjunct is the esophageal detector device. This device consists of a syringe or suction bulb that, after intubation, attaches to an endotracheal tube circuit adapter. 13–15Studies have demonstrated the high reliability of the esophageal detector device in patients with normal airways. Reported drawbacks include, however, a slow inflation of the suction bulb, lasting as long as 30 s 16; failure to confirm tracheal intubation in the presence of airway obstruction (asthma, mediastinal mass) 17; and inability to use the device in children younger than 1 yr of age, owing to a failure rate of 25% even with a small modified syringe. 18
A method that has not been attempted to address the problem being discussed here is acoustic reflectometry (fig. 1) . 19,20A reflectometer generates a series of acoustic impulses and analyzes the reflections from the cavity of interest to generate an area–distance curve. In essence, one obtains a “one-dimensional image” of the cavity of interest. I hypothesized that acoustic reflectometry could be used to distinguish between an esophageal and a tracheal intubation.
An acoustic reflectometer. The loudspeaker within the wavetube emits a series of pulses that travel down the wavetube and into the airways of the lung. The microphone, located within the distal wavetube, is connected to a computer-based system that analyzes the reflections returning from the lungs.
Materials and Methods
Ten adult patients provided informed consent in accordance with institutional review board protocol. After intravenous induction of general anesthesia, patients were intubated, paralyzed, and mechanically ventilated. Conventional laryngoscopy and tracheal intubation with an endotracheal tube were performed. In females, an endotracheal tube 7.0 mm ID was positioned at 21 cm at the teeth; in males, an 8.0-mm tube was positioned at 23 cm at the teeth. After confirmation of proper endotracheal tube placement using a capnograph, the presence of bilateral and equal breath sounds was confirmed in all patients.
Acoustic reflectometric profiles were obtained via use of a customized, computer-based acoustic reflectometer (Hood Labs, Pembroke, MA), characterized by a maximal measurement distance of 50 cm from the distal end of the wavetube. The reflectometer impulse, 2 ms in duration, is characterized by a flat spectral range from 0–5,000 kHz (low-pass filter) and is repeated at the rate of 5 pulses/s.
The reflectometer profile is a semilogarithmic plot of area versus length. The horizontal coordinate in the profile corresponds to axial length, in centimeters, with the origin taken to be the distal end of the wavetube and the axial direction taken to be parallel to the longitudinal length of the wavetube. The vertical coordinate of the profile, indicated in square centimeters, corresponds to the coronal cross-sectional area of the structure at a given axial length.
To obtain a measurement, the endotracheal tube was disconnected momentarily from the breathing circuit at the junction between the endotracheal tube adapter and the breathing circuit. The wavetube was attached to the endotracheal tube adapter via a 1-in long connecting piece of medical-grade silicone tubing, and an acoustic reflectometric image of the endotracheal tube–airway system was obtained, typically within 15 s.
With an endotracheal tube already in place, assuring satisfactory patient ventilation, a second endotracheal tube of the same diameter was passed into the esophagus to the same depth (21 cm in females, 23 cm in males). Four breaths with a manual reservoir bag were given through this esophageally placed tube, in simulation of the usual ventilatory attempts that follow an intubation. An acoustic reflectometric profile of the endotracheal tube–esophagus system then was made, before and after the four reservoir-bag breaths.
Results
Acoustic Reflectometric Profile of an Endotracheal Intubation
The acoustic reflectometric profile for an tracheal intubation is shown in figure 2. As one proceeds in the direction of positive axial length, the following segments are encountered, in successive order:
Acoustic reflectometric profile for an tracheal intubation. In this semilogarithmic plot, the origin is at the distal tip of the wavetube, with the horizontal coordinate parallel to the axial length of the wavetube. The vertical coordinate is the coronal cross-sectional area of the airway at the given axial length. In the positive axial direction, four segments are encountered: the short 1-in connector segment (A); the endotracheal tube segment (7.0-mm-ID endotracheal tube, cut to a length of 26 cm, and positioned at 21 cm at the teeth; B); the tracheal segment, from the distal endotracheal tube to the carina, approximately 5 cm long (C); and from the carina and beyond into the lung, with a corresponding rapid rise in the area (D).
- 1. The approximately 1-in long connector segment of silicone tubing (15.8 mm ID; calculated cross-sectional area of 1.96 cm2).
- 2. A segment of constant cross-sectional area, which reflects the fixed area throughout the length of the endotracheal tube, in this case a tube of 7.0 mm ID, positioned at 21 cm at the teeth and cut to a length of 26 cm. Because the area of the silicone connector is greater than the inner cross-sectional area of the 7.0-mm endotracheal tube (calculated cross-sectional area 0.385 cm2), there is a decrease in area in going from the former to the latter.
- 3. A sudden rise in cross-sectional area followed by the first plateau. The vertical height of the plateau corresponds to the fixed area of the trachea beyond the tube, i.e. , for a distance of approximately 5 cm extending from the distal end of the endotracheal tube to the yet more distal carina. The length of this segment is best estimated from the initiation of rise to the end of the plateau.
- 4. The exponentially increasing area beyond the carina, with ensuing successive area rises and plateaus that are the result of successive bifurcations within the lung.
For a breathing tube correctly placed in the trachea, the dominant and characteristic feature of the profile is the constant cross-sectional area segment corresponding to the length of the endotracheal tube, followed by a segment exhibiting a rapid rise in the area beyond the carina, as shown in figure 2.
Acoustic Reflectometric Profile of an Esophageal Intubation
For an endotracheal tube in the esophagus, the characteristic reflectometric profile (fig. 3) consists of a segment of constant cross-sectional area corresponding to the length of the endotracheal tube (7.0 mm internal diameter, 34 cm length including adapter, positioned at 21 cm at the teeth), followed by an immediate decrease in the cross-sectional area to zero; this occurs because the esophageal lumen normally is closed immediately beyond the tip of the endotracheal tube. After the application of the four reservoir-bag breaths, a separate acoustic reflectometric profile of the endotracheal tube–esophagus was made, but there was no detectable difference in the resulting profile.
Acoustic reflectometric profile of an esophageal intubation. For the same patient as in figure 1, the first two segments encountered are the connector segment and the endotracheal tube segment (7.0-mm-ID endotracheal tube of length 34 cm, positioned at 21 cm at the teeth); however, slightly beyond the tip of the endotracheal tube, the area profile immediately goes to zero because of the usual closure of the esophageal lumen.
A spike artifact is noted in some of the esophageal intubation profiles. That this spike is an artifact is evident from the stretch of zero-area axial length (caused by complete closure of the esophageal lumen) proximal to the spike.
Figure 4shows three more pairs of endotracheal and esophageal intubation profiles, which are unequivocal as to what type of intubation has occurred. All 10 patients exhibited the same characteristic endotracheal and esophageal intubation profiles.
Three more pairs of patient endotracheal and esophageal intubations. The profiles are distinctive and characteristic.
Discussion
Distinction between a tracheal intubation and an esophageal intubation is possible through the use of acoustic reflectometry because the cartilaginous trachea is usually patent, whereas the nonrigid esophagus normally has a closed lumen. In the esophagus, the compliant walls close around the distal end of the endotracheal tube. The closed esophageal lumen prevents further transmission of the acoustic impulse down the cavity, and the reflectometer profile area value goes to zero. The observed spike in the esophageal profile, possibly because of aliasing (using too low a sampling frequency, such that the spectral detection device mistakes multiples of the sampling frequency as inherent to the system), should be removable by better analog filtering.
The creation of one-dimensional acoustic reflectometric images, which allow the distinction to be made between endotracheal and esophageal intubation, does not depend on ventilation, nor does it depend on the presence of exhaled carbon dioxide. The shape and size of the one-dimensional image depend solely on the physical dimensions of the cavity being studied, i.e. , on the underlying human anatomy, be it esophagus or trachea, to which the reflectometer–breathing tube system is attached. In view of this unusual capability, acoustic reflectometry has the potential to become a new gold standard, much like fiberoptic bronchoscopy, in determining whether proper tube placement has occurred.
In 2 of 10 patients, it was thought that cutting the endotracheal tube to a shorter length (and thus increasing the airway area seen in the profile) would enhance the decision-making process regarding proper endotracheal tube placement. This enhancing maneuver turned out to be unnecessary, however, because the decision could be made readily even with an uncut endotracheal tube, and the maneuver was not used in the other eight patients. Use of a reflectometer with only a 40-cm axial-length range would have been sufficient to identify the position of a 30-cm-long endotracheal tube.
A validation study with a much larger number of patients is necessary to determine the sensitivity and specificity of acoustic reflectometry in the detection and confirmation of proper endotracheal tube placement. Clearly, the vast majority of patients exhibit normal airways, and in these patients acoustic reflectometry would be expected to distinguish readily between an endotracheal and an esophageal intubation. A special effort would have to be made to recruit study patients with aberrant tracheal and esophageal anatomy (e.g. , esophageal dilatation), with markedly altered pathophysiology (e.g. , pulmonary edema), or who are otherwise at risk for airway injury or obstruction (e.g. , mediastinal mass compressing the trachea). Should such a study validate the reliability of acoustic reflectometry, which seems to allow the rapid determination of proper breathing tube placement with an extraordinary degree of certainty, the implications would be highly significant in terms of its potential use by paramedical personnel in the field in the setting of cardiopulmonary arrest, by emergency room physicians, and by anesthesiologists.
The reflectometer generates an impulse that propagates down the reflectometer wavetube, through the endotracheal tube, and into the cavity of interest. Using a computer-based data acquisition and analysis system, the pressure amplitudes of the acoustic reflections, which return from the cavity back to a microphone within the distal part of the wavetube, are analyzed by the Gopillaud–Ware–Aki algorithm, 19,20which breaks up the axial length into small segments, such that the travel time of the impulse is the same in each segment. The algorithm is thus a layer-stripping process in that it goes from one segment to the immediately more distal segment. The known impedance at one axial segment is used, in conjunction with the known pressure reflections at the microphone, to generate iteratively the impedance value at the next successive axial segment, and so forth. Through this process, an equivalent total cross-sectional area is determined for each point of characteristic impedance at a given axial length from the end of the wavetube.
Consequently, in this algorithm, errors tend to accumulate as one proceeds distally into the cavity. To assure that the cumulative error is minimal and that measurements are always valid, a limit must be preset for the maximal axial length (customized to 50 cm in this study) at which specific impedance measurements can be made.
Clinical studies with acoustic reflectometry, with the patient using a mouthpiece, have focused on the nasopharyngeal cavity, 21the glottal area, 22the difficult airway, 23the imaging of the upper airway and the trachea, 24area changes in asthma patients, 25,26and the narrowing of the distal endotracheal tube as a result of accumulation of mucus in chronically intubated patients. 27The reproducibility and accuracy of the reflectometric data reportedly exhibit a within-run variability of 10 ± 4% and a mean intrasubject day–today variability of 9 ± 4%. 28
The specific reflectometer used for this study is impractically long (1 m) and is unlikely to be accepted widely. A two-microphone reflectometer 29,30recently has been developed, however, which is 30 cm long and which has a maximal measurement distance of 60 cm. This two-microphone reflectometer may be adaptable for day-to-day clinical practice in the emergency and operating room settings. Nonetheless, further development of the reflectometer is needed to make the instrument smaller, lighter, and more user friendly.
In summary, in this limited pilot study, the technique of acoustic reflectometry generated characteristic signature images that allowed the unequivocal distinction between an endotracheal and an esophageal intubation to be made within seconds, and without resort to capnography. Pending further validation of the technique in a large series of patients, this pilot study suggests that acoustic reflectometry may have a useful role in the emergency imaging of the airway, particularly in the detection of esophageal intubations in the cardiopulmonary arrest setting if the usual techniques for confirmation of tube placement fail.
The author thanks Marion Portnoy, R.N., and Drs. Remu Chhokra, Inankereen Ghandivel, Manoj Shahane, Sajjida Ahmad, and Lawrence Kushins for assistance with data collection, and Drs. Jerome Parness and Sheldon Goldstein for useful comments.
References
1. Shea SR, MacDonald JR, Gruzinski G: Prehospital endotracheal tube airway or esophageal gastric tube airway: A critical comparison. Ann Emerg Med 1985; 14:102–112
2. Stewart RD, Paris PM, Winter PM, Pelton GH, Cannon GM: Field endotracheal intubation by paramedical personnel-Success rates and complications. Chest 1984; 85:341–5
3. Birmingham PK, Cheney FW, Ward RJ: Esophageal intubation: A review of detection techniques. Anesth Analg 1986; 65:886–91
4. Stehling L: Management of the airway, Clinical Anesthesia. Edited by Barash PG, Cullen BF, Stoelting RK. Philadelphia, JB Lippincott, 1992, pp 698–9
5. Stone DJ, Gal TJ: Airway management, Anesthesia, 4th edition. Edited by Miller RD. New York, Churchill Livingstone, 1994 p 1430
6. Standards for Basic Anesthetic Monitoring, American Society of Anesthesiologists (last amended October 23), Directory of Members, 1996; 1998:438–9
7. Goldberg JS, Rawle PR, Zehnder JL, Sladen RN: Colorimetric end-tidal carbon dioxide monitoring for tracheal intubation. Anesth Analg 1990; 70:191–4
8. Murray IP, Modell JH: Early detection of endotracheal tube accidents by monitoring carbon dioxide concentration in respiratory gas. A nesthesiology 1983; 59:344–6
9. Denman WT, Hayes M, Higgins D, Wilkinson DJ: The Fenem CO2detector device. Anaesthesia 1990; 45:465–7
10. Macleod BA, Heller MB, Gerard J, Yealy DM, Menegazzi JJ: Verification of endotracheal-tube placement with colorimetric end-tidal CO2detection. Ann Emerg Med 1991; 20:267–70
11. Bhende MS, Thompson AE, Cook DR, Saville AL: Validity of a disposable end-tidal CO2detector in verifying endotracheal tube placement in infants and children. Ann Emerg Med 1992; 21:142–5
12. Vukmir RB, Heller MB, Stein KL: Confirmation of endotracheal tube placement: A miniaturized infrared qualitative CO2detector. Ann Emerg Med 1991; 20:726–9
13. Jenkins WA, Verdile VP, Paris PM: The syringe aspiration technique to verify endotracheal tube position. Am J Emerg Med 1994; 12:413–6
14. Williams KN, Nunn JF: The oesophageal detector device. Anaesthesia 1989; 44:412–4
15. Wee MYK: The oesophageal detector device: Assessment of a new method to distinguish oesophageal from tracheal intubation. Anaesthesia 1988; 43:27–9
16. Zaleski L, Abello D, Gold M: The esophageal detector device. A nesthesiology 1993; 79:244–7
17. Baraka A: The oesophageal detector device (letter). Anaesthesia 1991; 46:697
18. Haynes SR, Norton NS: Use of the oesophageal detector device in children under one year of age. Anaesthesia 1990; 45:1067–9
19. Jackson AC, Butler JP, Miller EJ, Hoppin FG, Dawson SV: Airway geometry by analysis of acoustic pulse response measurements. J Appl Physiol 1977; 43:523–36
20. Gopillaud PL: An approach to inverse filtering of near-surface segment effects from sesimic records. Geophysics 1961; 26:754–60
21. Ware JA, Aki K: Continuous and discrete inverse-scattering problems in a stratified elastic medium: I. Plane waves at normal incidence. J Acoust Soc Am 1969; 45:911–21
22. Bradley TD, Brown IG, Grossman RF, Zamel N, Martinez D, Phillipson EA, Hoffstein V: Pharyngeal size in snorers, nonsnorers, and patients with obstructive sleep apnea. N Engl J Med 1986; 315:1327–31
23. D’Urzo AD, Rubinstein I, Lawson VG, Vassal KP, Rebuck SS, Slutsky AS, Hoffstein V: Comparison of glottic areas measured by acoustic reflections vs computerized tomography. J Appl Physiol 1988; 64:367–70
24. Eckmann D, Glassenberg R, Gavriely N: A pilot study: Acoustic reflectometry and the difficult airway. Anesth Analg 1996; 83:1084–9
25. Fredberg JJ, Wohl MEB, Glass GM, Dorkin HL: Airway area by acoustic reflections measured at the mouth. J Appl Physiol 1980; 48:749–58
26. Molfino NA, Slutsky AS, Hoffstein V, McLean PA, Rebuck AS, Drazen JM, Zamel N: Changes in cross-sectional area areas induced by methacholine, histamine, and LTC in asthmatic subjects. Am Rev Respir Dis 1992; 146:577–80
27. Van Surrell C, Louis B, Lofaso F, Beydon L, Brochard L, Harf A, Fredberg J, Isabey D: Acoustic method to estimate the longitudinal area profile of endotracheal tubes. Am J Respir Crit Care Med 1994; 149:28–33
28. Brooks LJ, Castile RG, Glass GM, Griscom NT, Wohl MEB, Fredberg JJ: Reproducibility and accuracy of airway area by acoustic reflection. J Appl Physiol 1984; 57:777–87
29. Louis B, Glass GM, Fredberg JJ: Pulmonary airway area by the two-microphone acoustic reflection method. J Appl Physiol 1994; 76:2234–40
30. Louis B, Glass G, Kresen B, Fredberg JJ: Airway area by acoustic reflection: The two-microphone method. J Biomechanical Engineering Trans Am Soc Mechanical Engineers 1993; 115:278–85