Experimental characterisation of CCD cameras for HREM at 300 kV
RR Meyer et al. Ultramicroscopy. 2000 Sep.
Abstract
The modulation transfer function, noise transfer function and detection quantum efficiency of both 1k2 and 2k2 cameras attached to the same 300 kV FEGTEM have been measured. The results show that both the MTF and the DQE of the 2k2 camera are lower than for the 1k2 camera at frequencies above a third of the Nyquist limit. This lower per-pixel performance of the 2k2 camera is primarily a result of the use of a thicker phosphor scintillator. However, MTF and DQE of the 2k2 camera operated as an effective 1k2 camera by using 2 x 2 pixel binning are superior to the 1k2 camera in the medium spatial frequency range.
Similar articles
- Characterisation of the signal and noise transfer of CCD cameras for electron detection.
Meyer RR, Kirkland AI. Meyer RR, et al. Microsc Res Tech. 2000 May 1;49(3):269-80. doi: 10.1002/(SICI)1097-0029(20000501)49:3<269::AID-JEMT5>3.0.CO;2-B. Microsc Res Tech. 2000. PMID: 10816267 - Study of a prototype high quantum efficiency thick scintillation crystal video-electronic portal imaging device.
Samant SS, Gopal A. Samant SS, et al. Med Phys. 2006 Aug;33(8):2783-91. doi: 10.1118/1.2216877. Med Phys. 2006. PMID: 16964854 - Solid-state fluoroscopic imager for high-resolution angiography: physical characteristics of an 8 cm x 8 cm experimental prototype.
Vedantham S, Karellas A, Suryanarayanan S, Onishi SK. Vedantham S, et al. Med Phys. 2004 Jun;31(6):1462-72. doi: 10.1118/1.1750992. Med Phys. 2004. PMID: 15259649 Free PMC article. - Physical evaluation of a needle photostimulable phosphor based CR mammography system.
Marshall NW, Lemmens K, Bosmans H. Marshall NW, et al. Med Phys. 2012 Feb;39(2):811-24. doi: 10.1118/1.3675403. Med Phys. 2012. PMID: 22320791 - Performance of a 2k CCD camera designed for electron crystallography at 400 kV.
Downing KH, Hendrickson FM. Downing KH, et al. Ultramicroscopy. 1999 Jan;75(4):215-33. doi: 10.1016/s0304-3991(98)00065-5. Ultramicroscopy. 1999. PMID: 9919710
Cited by
- Ranking TEM cameras by their response to electron shot noise.
Grob P, Bean D, Typke D, Li X, Nogales E, Glaeser RM. Grob P, et al. Ultramicroscopy. 2013 Oct;133:1-7. doi: 10.1016/j.ultramic.2013.01.003. Epub 2013 Jan 29. Ultramicroscopy. 2013. PMID: 23747527 Free PMC article. - Practical factors affecting the performance of a thin-film phase plate for transmission electron microscopy.
Danev R, Glaeser RM, Nagayama K. Danev R, et al. Ultramicroscopy. 2009 Mar;109(4):312-25. doi: 10.1016/j.ultramic.2008.12.006. Epub 2008 Dec 11. Ultramicroscopy. 2009. PMID: 19157711 Free PMC article. - Determination of atomic positions from time resolved high resolution transmission electron microscopy images.
Hussaini Z, Lin PA, Natarajan B, Zhu W, Sharma R. Hussaini Z, et al. Ultramicroscopy. 2018 Mar;186:139-145. doi: 10.1016/j.ultramic.2017.12.018. Epub 2017 Dec 28. Ultramicroscopy. 2018. PMID: 29324400 Free PMC article. - Electron counting and beam-induced motion correction enable near-atomic-resolution single-particle cryo-EM.
Li X, Mooney P, Zheng S, Booth CR, Braunfeld MB, Gubbens S, Agard DA, Cheng Y. Li X, et al. Nat Methods. 2013 Jun;10(6):584-90. doi: 10.1038/nmeth.2472. Epub 2013 May 5. Nat Methods. 2013. PMID: 23644547 Free PMC article. - Characterization of a direct detection device imaging camera for transmission electron microscopy.
Milazzo AC, Moldovan G, Lanman J, Jin L, Bouwer JC, Klienfelder S, Peltier ST, Ellisman MH, Kirkland AI, Xuong NH. Milazzo AC, et al. Ultramicroscopy. 2010 Jun;110(7):744-7. doi: 10.1016/j.ultramic.2010.03.007. Epub 2010 Mar 25. Ultramicroscopy. 2010. PMID: 20382479 Free PMC article.
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous