Validation of the Omron HEM-907 device for blood pressure measurement - PubMed (original) (raw)
Validation of the Omron HEM-907 device for blood pressure measurement
Mohamed A El Assaad et al. Blood Press Monit. 2002 Aug.
Abstract
Background: The aim of this study was to validate the Omron HEM-907 blood pressure (BP) measuring device according to the international validation protocol.
Design: The international validation protocol is divided into two phases: the first phase is performed on 15 selected subjects and if the device passes this phase, 18 more subjects are selected making a total number of 33 subjects on which the final validation is performed.
Methods: For each subject, BP measurements were performed simultaneously by two trained observers using mercury sphygmomanometers alternately with the Omron HEM-907 device. In all, 99 measurements were obtained for comparison. The difference between the BP value given by the device and that obtained by the two observers (mean of the two observers) was calculated for each measure.
Results: The difference between the two observers was -1 +/- 2 mmHg for the systolic BP (SBP) and for the diastolic BP (DBP). The Omron HEM-907 passed the first phase of the validation process. For the second phase, the average differences between the device and mercury sphygmomanometer readings were -1 +/- 7 and -5 +/- 6 mmHg for SBP and DBP respectively. Readings for the HEM-907 device differed by less than 5 mmHg for 61 of the systolic readings and 52 of the diastolic readings; by less than 10 mmHg for 85 of the systolic readings and 85 of the diastolic readings; and by less than 15 mmHg for 94 of the systolic readings and 96 of the diastolic readings.
Conclusions: The Omron HEM-907 device passes the two phases of the international validation protocol.
Copyright 2002 Lippincott Williams & Wilkins
Similar articles
- Validation of two automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension: the Omron M6 (HEM-7001-E) and the Omron R7 (HEM 637-IT).
Topouchian JA, El Assaad MA, Orobinskaia LV, El Feghali RN, Asmar RG. Topouchian JA, et al. Blood Press Monit. 2006 Jun;11(3):165-71. doi: 10.1097/01.mbp.0000209078.17246.34. Blood Press Monit. 2006. PMID: 16702826 - Evaluation of two devices for self-measurement of blood pressure according to the international protocol: the Omron M5-I and the Omron 705IT.
El Assaad MA, Topouchian JA, Asmar RG. El Assaad MA, et al. Blood Press Monit. 2003 Jun;8(3):127-33. doi: 10.1097/00126097-200306000-00006. Blood Press Monit. 2003. PMID: 12900590 - Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension.
Belghazi J, El Feghali RN, Moussalem T, Rejdych M, Asmar RG. Belghazi J, et al. Vasc Health Risk Manag. 2007;3(4):389-400. Vasc Health Risk Manag. 2007. PMID: 17969368 Free PMC article. - Determining which automatic digital blood pressure device performs adequately: a systematic review.
Wan Y, Heneghan C, Stevens R, McManus RJ, Ward A, Perera R, Thompson M, Tarassenko L, Mant D. Wan Y, et al. J Hum Hypertens. 2010 Jul;24(7):431-8. doi: 10.1038/jhh.2010.37. Epub 2010 Apr 8. J Hum Hypertens. 2010. PMID: 20376077 Free PMC article. Review. - Standardized protocol of blood pressure measurement and quality control program for the Korea National Health and Nutrition Examination Survey.
Kim HL, Park SM, Cho IJ, Kim YM, Kim DH, Kim SH, Kim KI, Sung KC, Ihm SH, Shin J, Kim Y, Oh K, Lee EM; Quality control team of blood pressure measurements for the Korea National Health, Nutrition Examination Survey (KNHANES). Kim HL, et al. Clin Hypertens. 2023 Oct 12;29(1):28. doi: 10.1186/s40885-023-00252-7. Clin Hypertens. 2023. PMID: 37821979 Free PMC article. Review.
Cited by
- Association of von Willebrand factor blood levels with exercise hypertension.
Nikolic SB, Adams MJ, Otahal P, Edwards LM, Sharman JE. Nikolic SB, et al. Eur J Appl Physiol. 2015 May;115(5):1057-65. doi: 10.1007/s00421-014-3087-3. Epub 2014 Dec 25. Eur J Appl Physiol. 2015. PMID: 25539599 - Associations and clinical relevance of aortic-brachial artery stiffness mismatch, aortic reservoir function, and central pressure augmentation.
Schultz MG, Hughes AD, Davies JE, Sharman JE. Schultz MG, et al. Am J Physiol Heart Circ Physiol. 2015 Oct;309(7):H1225-33. doi: 10.1152/ajpheart.00317.2015. Epub 2015 Aug 14. Am J Physiol Heart Circ Physiol. 2015. PMID: 26276816 Free PMC article. - Safety and feasibility of achieving lower systolic blood pressure goals in persons with type 2 diabetes: the SANDS trial.
Weir MR, Yeh F, Silverman A, Devereux RB, Galloway JM, Henderson JA, Howard WJ, Russell M, Wilson C, Ratner R, Sorkin J, Umans JG, Fleg JL, Stylianou M, Lee E, Howard BV. Weir MR, et al. J Clin Hypertens (Greenwich). 2009 Oct;11(10):540-8. doi: 10.1111/j.1751-7176.2009.00121.x. J Clin Hypertens (Greenwich). 2009. PMID: 19817934 Free PMC article. Clinical Trial. - Validation of the Instant Blood Pressure Smartphone App.
Plante TB, Urrea B, MacFarlane ZT, Blumenthal RS, Miller ER 3rd, Appel LJ, Martin SS. Plante TB, et al. JAMA Intern Med. 2016 May 1;176(5):700-2. doi: 10.1001/jamainternmed.2016.0157. JAMA Intern Med. 2016. PMID: 26938174 Free PMC article. No abstract available. - Clinical Profiles and Factors Associated with a Low Sodium Intake in the Population: An Analysis of the Swiss Survey on Salt.
Burnier M, Paccaud FM, Bochud M. Burnier M, et al. Nutrients. 2020 Nov 23;12(11):3591. doi: 10.3390/nu12113591. Nutrients. 2020. PMID: 33238516 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources