Microbubbles induce renal hemorrhage when exposed to diagnostic ultrasound in anesthetized rats
James H Wible Jr et al. Ultrasound Med Biol. 2002 Nov-Dec.
Abstract
The generation of ultrasound (US) bioeffects using a clinical imaging system is controversial. We tested the hypothesis that the presence of microbubbles in the US field of a medical imager induces biologic effects. Both kidneys of anesthetized rats were insonified for 5 min using a medical imaging system after the administration of microbubbles. One kidney was insonified using a continuous mode (30 Hz) and the opposite kidney was insonified using an intermittent (1 Hz) technique. The microbubbles were exposed to three different transducer frequencies and four transducer output powers. After insonification, the animals were euthanized, the kidneys were removed and their gross appearance scored under "blinded" conditions using a defined scale. After the administration of microbubbles, US imaging of the kidney caused hemorrhage in the renal tissue. The severity and area of hemorrhage increased with an increase in the transducer power and a decrease in the transducer frequency. Intermittent insonification in the presence of microbubbles produced a greater degree of renal hemorrhage than continuous imaging techniques.
Similar articles
- The Dependence of Glomerular Capillary Hemorrhage Induced by Contrast Enhanced Diagnostic Ultrasound on Microbubble Diameter.
Miller DL, Lu X, Dou C, Fabiilli ML, Church CC. Miller DL, et al. Ultrasound Med Biol. 2018 Mar;44(3):613-621. doi: 10.1016/j.ultrasmedbio.2017.11.007. Epub 2017 Dec 26. Ultrasound Med Biol. 2018. PMID: 29287999 Free PMC article. - Bioeffects of positive and negative acoustic pressures in mice infused with microbubbles.
Dalecki D, Child SZ, Raeman CH, Xing C, Gracewski S, Carstensen EL. Dalecki D, et al. Ultrasound Med Biol. 2000 Oct;26(8):1327-32. doi: 10.1016/s0301-5629(00)00297-0. Ultrasound Med Biol. 2000. PMID: 11120371 - Nephron injury induced by diagnostic ultrasound imaging at high mechanical index with gas body contrast agent.
Williams AR, Wiggins RC, Wharram BL, Goyal M, Dou C, Johnson KJ, Miller DL. Williams AR, et al. Ultrasound Med Biol. 2007 Aug;33(8):1336-44. doi: 10.1016/j.ultrasmedbio.2007.03.002. Epub 2007 May 16. Ultrasound Med Biol. 2007. PMID: 17507144 Free PMC article. - Bubble dynamics involved in ultrasonic imaging.
Postema M, Schmitz G. Postema M, et al. Expert Rev Mol Diagn. 2006 May;6(3):493-502. doi: 10.1586/14737159.6.3.493. Expert Rev Mol Diagn. 2006. PMID: 16706749 Review. - Overview of experimental studies of biological effects of medical ultrasound caused by gas body activation and inertial cavitation.
Miller DL. Miller DL. Prog Biophys Mol Biol. 2007 Jan-Apr;93(1-3):314-30. doi: 10.1016/j.pbiomolbio.2006.07.027. Epub 2006 Aug 22. Prog Biophys Mol Biol. 2007. PMID: 16989895 Review.
Cited by
- A pilot study to assess markers of renal damage in the rodent kidney after exposure to 7 MHz ultrasound pulse sequences designed to cause microbubble translation and disruption.
Johnson K, Cianciolo R, Gessner RC, Dayton PA. Johnson K, et al. Ultrasound Med Biol. 2012 Jan;38(1):168-72. doi: 10.1016/j.ultrasmedbio.2011.10.001. Epub 2011 Nov 21. Ultrasound Med Biol. 2012. PMID: 22104535 Free PMC article. - [Ultrasound contrast agents. Pharmaceutical drug safety and bioeffects].
Krix M, Jenne JW. Krix M, et al. Radiologe. 2007 Sep;47(9):800-7. doi: 10.1007/s00117-007-1544-2. Radiologe. 2007. PMID: 17876626 Review. German. - Capillary Hemorrhage Induced by Contrast-Enhanced Diagnostic Ultrasound in Rat Intestine.
Lu X, Dou C, Fabiilli ML, Miller DL. Lu X, et al. Ultrasound Med Biol. 2019 Aug;45(8):2133-2139. doi: 10.1016/j.ultrasmedbio.2019.04.012. Epub 2019 May 14. Ultrasound Med Biol. 2019. PMID: 31101449 Free PMC article. - Focused Ultrasound: Noninvasive Image-Guided Therapy.
Moonen CTW, Kilroy JP, Klibanov AL. Moonen CTW, et al. Invest Radiol. 2025 Mar 1;60(3):205-219. doi: 10.1097/RLI.0000000000001116. Epub 2024 Aug 21. Invest Radiol. 2025. PMID: 39163359 Free PMC article. Review. - Leveraging the power of ultrasound for therapeutic design and optimization.
Caskey CF, Hu X, Ferrara KW. Caskey CF, et al. J Control Release. 2011 Dec 20;156(3):297-306. doi: 10.1016/j.jconrel.2011.07.032. Epub 2011 Jul 30. J Control Release. 2011. PMID: 21835212 Free PMC article. Review.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical