Testing of work environments for electromagnetic interference
D Marco et al. Pacing Clin Electrophysiol. 1992 Nov.
Abstract
A challenge for pacemaker therapists is whether a patient working in an environment with the potential for electromagnetic interference (EMI) can return to their work after a pacemaker has been implanted. Common practice has been to prohibit pacemaker patients from using electric welding machines. Twelve work environments and a new method for monitoring the pacemaker rhythm in the presence of EMI were tested. The new method uses a special memory called the event record found in several Siemens Pacesetter pacemaker models. Surface ECGs with a marking system, intracardiac electrograms, and a digital monitor were used to verify the results with event records. The results from several sources of EMI are reported. Twenty-one in vivo and in vitro tests were performed in the work environments of 12 patients. Event records were useful and accurate both in vivo and in vitro. Electric are welding machines up to 225 A did not affect these pacemakers. Arc welding machines using 1,000 A or more inhibited the in vitro test system within 1 or 2 meters of the weld or power generator. Electric welding machines with high frequency voltage superimposed on the welding current affected the pacemaker when it was within 2 meters of the power unit and 1 meter of the weld. Very large industrial degaussing coils affected pacemakers within 2 meters. The test method using event records was found to be an effective addition to monitoring the pacemaker. These results are specific for the pacemaker models tested. Such testing allows the physician to make a knowledgeable decision regarding return to work for the pacemaker patient in a high EMI environment.
Similar articles
- Electromagnetic interference from welding and motors on implantable cardioverter-defibrillators as tested in the electrically hostile work site.
Fetter JG, Benditt DG, Stanton MS. Fetter JG, et al. J Am Coll Cardiol. 1996 Aug;28(2):423-7. doi: 10.1016/0735-1097(96)00147-7. J Am Coll Cardiol. 1996. PMID: 8800120 - Electromagnetic interference with cardiac pacemakers and implantable cardioverter-defibrillators from low-frequency electromagnetic fields in vivo.
Tiikkaja M, Aro AL, Alanko T, Lindholm H, Sistonen H, Hartikainen JE, Toivonen L, Juutilainen J, Hietanen M. Tiikkaja M, et al. Europace. 2013 Mar;15(3):388-94. doi: 10.1093/europace/eus345. Epub 2012 Nov 1. Europace. 2013. PMID: 23125355 - The influence of elevated 50 Hz electric and magnetic fields on implanted cardiac pacemakers: the role of the lead configuration and programming of the sensitivity.
Toivonen L, Valjus J, Hongisto M, Metso R. Toivonen L, et al. Pacing Clin Electrophysiol. 1991 Dec;14(12):2114-22. doi: 10.1111/j.1540-8159.1991.tb06482.x. Pacing Clin Electrophysiol. 1991. PMID: 1723194 - Implantable rhythm devices and electromagnetic interference: myth or reality?
Dyrda K, Khairy P. Dyrda K, et al. Expert Rev Cardiovasc Ther. 2008 Jul;6(6):823-32. doi: 10.1586/14779072.6.6.823. Expert Rev Cardiovasc Ther. 2008. PMID: 18570620 Review. - [Electrocoagulation in patients with cardiac pacemakers: what should be considered?].
Habicht JM, Stulz P. Habicht JM, et al. Chirurg. 1995 Aug;66(8):774-9. Chirurg. 1995. PMID: 7587540 Review. German.
Cited by
- Adding wisdom to 'smart' bioelectronic systems: a design framework for physiologic control including practical examples.
Gunduz A, Opri E, Gilron R, Kremen V, Worrell G, Starr P, Leyde K, Denison T. Gunduz A, et al. Bioelectron Med (Lond). 2019 Mar;2(1):29-41. doi: 10.2217/bem-2019-0008. Epub 2019 May 30. Bioelectron Med (Lond). 2019. PMID: 33868718 Free PMC article. - Electromagnetic interference on pacemakers.
Erdogan O. Erdogan O. Indian Pacing Electrophysiol J. 2002 Jul 1;2(3):74-8. Indian Pacing Electrophysiol J. 2002. PMID: 17006562 Free PMC article. - Effects of radiofrequency catheter ablation on patients with permanent pacemakers.
Sadoul N, Blankoff I, de Chillou C, Beurrier D, Messier M, Bizeau O, Magnin I, Dodinot B, Aliot E. Sadoul N, et al. J Interv Card Electrophysiol. 1997 Nov;1(3):227-33. doi: 10.1023/a:1009721023732. J Interv Card Electrophysiol. 1997. PMID: 9869976 - Electromagnetic interference in patients with implanted cardioverter-defibrillators and implantable loop recorders.
de Sousa M, Klein G, Korte T, Niehaus M. de Sousa M, et al. Indian Pacing Electrophysiol J. 2002 Jul 1;2(3):79-84. Indian Pacing Electrophysiol J. 2002. PMID: 17006563 Free PMC article. No abstract available. - Testing of common electromagnetic environments for risk of interference with cardiac pacemaker function.
Tiikkaja M, Aro AL, Alanko T, Lindholm H, Sistonen H, Hartikainen JE, Toivonen L, Juutilainen J, Hietanen M. Tiikkaja M, et al. Saf Health Work. 2013 Sep;4(3):156-9. doi: 10.1016/j.shaw.2013.06.002. Epub 2013 Jul 20. Saf Health Work. 2013. PMID: 24106646 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical