Biomechanical evaluation of a novel lumbosacral axial fixation device
Eric H Ledet et al. J Biomech Eng. 2005 Nov.
Abstract
Background: Interbody arthrodesis is employed in the lumbar spine to eliminate painful motion and achieve stability through bony fusion. Bone grafts, metal cages, composite spacers, and growth factors are available and can be placed through traditional open techniques or minimally invasively. Whether placed anteriorly, posteriorly, or laterally, insertion of these implants necessitates compromise of the anulus--an inherently destabilizing procedure. A new axial percutaneous approach to the lumbosacral spine has been described. Using this technique, vertical access to the lumbosacral spine is achieved percutaneously via the presacral space. An implant that can be placed across a motion segment without compromise to the anulus avoids surgical destabilization and may be advantageous for interbody arthrodesis. The purpose of this study was to evaluate the in vitro biomechanical performance of the axial fixation rod, an anulus sparing, centrally placed interbody fusion implant for motion segment stabilization.
Method of approach: Twenty-four bovine lumbar motion segments were mechanically tested using an unconstrainedflexibility protocol in sagittal and lateral bending, and torsion. Motion segments were also tested in axial compression. Each specimen was tested in an intact state, then drilled (simulating a transaxial approach to the lumbosacral spine), then with one of two axial fixation rods placed in the spine for stabilization. The range of motion, bending stiffness, and axial compressive stiffness were determined for each test condition. Results were compared to those previously reported for femoral ring allografts, bone dowels, BAK and BAK Proximity cages, Ray TFC, Brantigan ALIF and TLIF implants, the InFix Device, Danek TIBFD, single and double Harms cages, and Kaneda, Isola, and University plating systems.
Results: While axial drilling of specimens had little effect on stiffness and range of motion, specimens implanted with the axial fixation rod exhibited significant increases in stiffness and decreases in range of motion relative to intact state. When compared to existing anterior, posterior, and interbody instrumentation, lateral and sagittal bending stiffness of the axial fixation rod exceeded that of all other interbody devices, while stiffness in extension and axial compression were comparable to plate and rod constructs. Torsional stiffness was comparable to other interbody constructs and slightly lower than plate and rod constructs.
Conclusions: For stabilization of the L5-S1 motion segment, axial placement of implants offers potential benefits relative to traditional exposures. The preliminary biomechanical data from this study indicate that the axial fixation rod compares favorably to other devices and may be suitable to reduce pathologic motion at L5-S1, thus promoting bony fusion.
Similar articles
- Iliac screws may not be necessary in long-segment constructs with L5-S1 anterior lumbar interbody fusion: cadaveric study of stability and instrumentation strain.
Hlubek RJ, Godzik J, Newcomb AGUS, Lehrman JN, de Andrada B, Bohl MA, Farber SH, Kelly BP, Turner JD. Hlubek RJ, et al. Spine J. 2019 May;19(5):942-950. doi: 10.1016/j.spinee.2018.11.004. Epub 2018 Nov 9. Spine J. 2019. PMID: 30419290 - MIS Expandable Interbody Spacers: A Literature Review and Biomechanical Comparison of an Expandable MIS TLIF With Conventional TLIF and ALIF.
Cannestra AF, Peterson MD, Parker SR, Roush TF, Bundy JV, Turner AW. Cannestra AF, et al. Spine (Phila Pa 1976). 2016 Apr;41 Suppl 8:S44-9. doi: 10.1097/BRS.0000000000001465. Spine (Phila Pa 1976). 2016. PMID: 26825792 Review. - Biomechanical evaluation of lumbosacral reconstruction techniques for spondylolisthesis: an in vitro porcine model.
Cunningham BW, Lewis SJ, Long J, Dmitriev AE, Linville DA, Bridwell KH. Cunningham BW, et al. Spine (Phila Pa 1976). 2002 Nov 1;27(21):2321-7. doi: 10.1097/00007632-200211010-00004. Spine (Phila Pa 1976). 2002. PMID: 12438979 - Biomechanical properties of threaded inserts for lumbar interbody spinal fusion.
Tencer AF, Hampton D, Eddy S. Tencer AF, et al. Spine (Phila Pa 1976). 1995 Nov 15;20(22):2408-14. doi: 10.1097/00007632-199511001-00007. Spine (Phila Pa 1976). 1995. PMID: 8578391 - The axial transsacral approach to interbody fusion at L5-S1.
Issack PS, Kotwal SY, Boachie-Adjei O. Issack PS, et al. Neurosurg Focus. 2014 May;36(5):E8. doi: 10.3171/2014.2.FOCUS13467. Neurosurg Focus. 2014. PMID: 24785490 Review.
Cited by
- Spine biomechanical testing methodologies: The controversy of consensus vs scientific evidence.
Costi JJ, Ledet EH, O'Connell GD. Costi JJ, et al. JOR Spine. 2021 Jan 5;4(1):e1138. doi: 10.1002/jsp2.1138. eCollection 2021 Mar. JOR Spine. 2021. PMID: 33778410 Free PMC article. Review. - Minimally invasive trans-sacral approach to L5-S1 interbody fusion: Preliminary results from 1 center and review of the literature.
Bradley WD, Hisey MS, Verma-Kurvari S, Ohnmeiss DD. Bradley WD, et al. Int J Spine Surg. 2012 Dec 1;6:110-4. doi: 10.1016/j.ijsp.2011.12.005. eCollection 2012. Int J Spine Surg. 2012. PMID: 25694879 Free PMC article. - Low Back Pain: Current Surgical Approaches.
Baliga S, Treon K, Craig NJ. Baliga S, et al. Asian Spine J. 2015 Aug;9(4):645-57. doi: 10.4184/asj.2015.9.4.645. Epub 2015 Jul 28. Asian Spine J. 2015. PMID: 26240729 Free PMC article. Review. - Alternative Surgical Strategy for AxiaLIF Pseudarthrosis: A Series of Three Case Reports.
Louwerens JK, Groot D, van Duijvenbode DC, Spruit M. Louwerens JK, et al. Evid Based Spine Care J. 2013 Oct;4(2):143-8. doi: 10.1055/s-0033-1357357. Evid Based Spine Care J. 2013. PMID: 24436713 Free PMC article. - State of the art advances in minimally invasive surgery for adult spinal deformity.
Hussain I, Fu KM, Uribe JS, Chou D, Mummaneni PV. Hussain I, et al. Spine Deform. 2020 Dec;8(6):1143-1158. doi: 10.1007/s43390-020-00180-8. Epub 2020 Aug 6. Spine Deform. 2020. PMID: 32761477 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials