Safety of electronic apex locators and pulp testers in patients with implanted cardiac pacemakers or cardioverter/defibrillators
Brian L Wilson et al. J Endod. 2006 Sep.
Abstract
The purpose of this study was to determine if electronic apex locators (EAL) or electric pulp testers (EPT) interfere with the function of implanted cardiac pacemakers (ICP) or cardioverter/defibrillators (ICD). Twenty-seven patients with ICPs or ICDs had continuous electrocardiogram monitoring and device interrogation to detect interferences during the use of two types of EALs and one EPT. No interferences were detected by any ICP or ICD. In six patients, with intermittent pacing, a significant increase in pacing was observed during EAL/EPT stimulation (p < 0.05). Examination of RR intervals (a measure of intrinsic heart rate) demonstrated significantly longer RR intervals (slower intrinsic heart rate) during EAL/EPT stimulation (p < 0.05). Evaluation of the electrocardiograms for each patient failed to show any abnormalities in pacing during testing. These findings led us to conclude that the increased pacing frequency observed was related to a slower intrinsic heart rate and not electrical interference with the cardiac devices. In conclusion, the two EALs and one EPT used in this study did not interfere with the functioning of any of the cardiac devices tested.
Similar articles
- Interference of electronic apex locators with implantable cardioverter defibrillators.
Idzahi K, de Cock CC, Shemesh H, Brand HS. Idzahi K, et al. J Endod. 2014 Feb;40(2):277-80. doi: 10.1016/j.joen.2013.07.027. Epub 2013 Sep 11. J Endod. 2014. PMID: 24461418 - Effect of electronic apex locators on cardiac pacemaker function.
Garofalo RR, Ede EN, Dorn SO, Kuttler S. Garofalo RR, et al. J Endod. 2002 Dec;28(12):831-3. doi: 10.1097/00004770-200212000-00009. J Endod. 2002. PMID: 12489653 - Influence of electronic apex locators and a gutta-percha heating device on implanted cardiac devices: an in vivo study.
Moraes AP, Silva EJ, Lamas CC, Portugal PH, Neves AA. Moraes AP, et al. Int Endod J. 2016 Jun;49(6):526-32. doi: 10.1111/iej.12487. Epub 2015 Jul 15. Int Endod J. 2016. PMID: 26105663 - Effects of electromagnetic interference on implanted cardiac devices and their management.
Yerra L, Reddy PC. Yerra L, et al. Cardiol Rev. 2007 Nov-Dec;15(6):304-9. doi: 10.1097/CRD.0b013e31813e0ba9. Cardiol Rev. 2007. PMID: 18090066 Review. - Electronic apex locator.
Kim E, Lee SJ. Kim E, et al. Dent Clin North Am. 2004 Jan;48(1):35-54. doi: 10.1016/j.cden.2003.10.005. Dent Clin North Am. 2004. PMID: 15066506 Review.
Cited by
- Influence and safety of electronic apex locators in patients with cardiovascular implantable electronic devices: a systematic review.
AlRahabi MK, Ghabbani HM. AlRahabi MK, et al. Libyan J Med. 2019 Dec;14(1):1547071. doi: 10.1080/19932820.2018.1547071. Libyan J Med. 2019. PMID: 30458679 Free PMC article. - Small bowel capsule endoscopy in patients with cardiac pacemakers and implantable cardioverter defibrillators: Outcome analysis using telemetry review.
Cuschieri JR, Osman MN, Wong RC, Chak A, Isenberg GA. Cuschieri JR, et al. World J Gastrointest Endosc. 2012 Mar 16;4(3):87-93. doi: 10.4253/wjge.v4.i3.87. World J Gastrointest Endosc. 2012. PMID: 22442746 Free PMC article. - Dental pulp testing: a review.
Chen E, Abbott PV. Chen E, et al. Int J Dent. 2009;2009:365785. doi: 10.1155/2009/365785. Epub 2009 Nov 12. Int J Dent. 2009. PMID: 20339575 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous