Fluorescence photobleaching recovery measurement of protein absolute diffusion constants

B G Barisas et al. Biophys Chem. 1979 Sep.

Abstract

The technique of fluorescence photobleaching recovery [Axelrod et al., Biophys. J. 16 (1976) 1055] has been applied to the measurement of absolute diffusion constants of a number of fluoiescein isothiocyanate-labeled proteins. Measured diffusion constants agree to within +/- 7% of published values for the underivatized proteins. The method has sufficient sensitivity to reveal the concentration dependence at neutral pH of the diffusion constant of alpha-chymotrypsin. The rapidity with which the labelling and measurements can be performed and the small amount of material required suggest the technique may be useful in rapid characterization of small protein samples. Some developments in optical and electronic systems and in data processing for this technique are discussed.