The Dynesys lumbar spinal stabilization system: a preliminary report on positional magnetic resonance imaging findings
James Beastall et al. Spine (Phila Pa 1976). 2007.
Abstract
Study design: We present the positional magnetic resonance imaging findings of a prospective case series of patients undergoing surgery with the Dynesys spinal stabilization device (Zimmer, Inc., Warsaw, IN).
Objective: To explore the biomechanical impact of the Dynesys device in vivo.
Summary of background data: Spinal fusion surgery is widely used for painful degenerative conditions of the lumbar spine that have not responded to conservative measures. It often produces good outcomes but can be associated with adjacent segment hypermobility that may lead to further degeneration and pain. Previous cadaveric biomechanical studies claim that the Dynesys Dynamic Spinal Stabilization System allows some lumbar movement, behaving similar to a normal spine in extension but similar to rigid fixation in flexion.
Methods: Twenty-four patients with dominant low back pain, with or without leg pain, were treated with the Dynesys. All patients underwent positional magnetic resonance imaging before surgery and 9 months after surgery. Measurements were made to assess the differences at the operated level, adjacent level, and whole lumbar spine.
Results: There was a statistically significant reduction in flexion-extension range of movement of both the whole lumbar spine by 13.37 degrees (P = 0.002) and at the instrumented segments by 4.08 degrees (P < 0.001) following surgery. There was an insignificant reduction in range of movement at the level above instrumentation (P = 0.807). Mean anterior disc height at the instrumented level reduced by 0.7 mm following insertion of the Dynesys (P < 0.027). Mean posterior disc height reduced by 0.3 mm (P = 0.453). In a neutral posture, the Dynesys had no significant impact on lordosis or inclination of operated or adjacent levels. Contrary to cadaveric study findings, the Dynesys appears to restrict extension more than flexion with respect to a neutral posture.
Conclusions: In vivo, the Dynesys Stabilization System allows movement at the instrumented level, albeit reduced, with no significant increased mobility at the adjacent segments. There was reduction of the anterior disc height without a significant increase of the posterior disc height.
Similar articles
- Disc changes in the bridged and adjacent segments after Dynesys dynamic stabilization system after two years.
Kumar A, Beastall J, Hughes J, Karadimas EJ, Nicol M, Smith F, Wardlaw D. Kumar A, et al. Spine (Phila Pa 1976). 2008 Dec 15;33(26):2909-14. doi: 10.1097/BRS.0b013e31818bdca7. Spine (Phila Pa 1976). 2008. PMID: 19092623 - Dynamic bulging of intervertebral discs in the degenerative lumbar spine.
Zou J, Yang H, Miyazaki M, Morishita Y, Wei F, McGovern S, Wang JC. Zou J, et al. Spine (Phila Pa 1976). 2009 Nov 1;34(23):2545-50. doi: 10.1097/BRS.0b013e3181b32998. Spine (Phila Pa 1976). 2009. PMID: 19841611 - Biomechanical evaluation of total disc replacement arthroplasty: an in vitro human cadaveric model.
Cunningham BW, Gordon JD, Dmitriev AE, Hu N, McAfee PC. Cunningham BW, et al. Spine (Phila Pa 1976). 2003 Oct 15;28(20):S110-7. doi: 10.1097/01.BRS.0000092209.27573.90. Spine (Phila Pa 1976). 2003. PMID: 14560182 - [Controversies about instrumented surgery and pain relief in degenerative lumbar spine pain. Results of scientific evidence].
Robaina-Padrón FJ. Robaina-Padrón FJ. Neurocirugia (Astur). 2007 Oct;18(5):406-13. Neurocirugia (Astur). 2007. PMID: 18008014 Review. Spanish. - Biomechanics of lumbar arthroplasty.
Crawford NR. Crawford NR. Neurosurg Clin N Am. 2005 Oct;16(4):595-602, v. doi: 10.1016/j.nec.2005.06.002. Neurosurg Clin N Am. 2005. PMID: 16326282 Review.
Cited by
- Minimally invasive dynamic screw stabilization using cortical bone trajectory.
Chang CC, Kuo CH, Chang HK, Tu TH, Fay LY, Wu JC, Cheng H, Huang WC. Chang CC, et al. BMC Musculoskelet Disord. 2020 Sep 10;21(1):605. doi: 10.1186/s12891-020-03629-z. BMC Musculoskelet Disord. 2020. PMID: 32912278 Free PMC article. - Kinematic evaluation of the adjacent segments after lumbar instrumented surgery: a comparison between rigid fusion and dynamic non-fusion stabilization.
Morishita Y, Ohta H, Naito M, Matsumoto Y, Huang G, Tatsumi M, Takemitsu Y, Kida H. Morishita Y, et al. Eur Spine J. 2011 Sep;20(9):1480-5. doi: 10.1007/s00586-011-1701-1. Epub 2011 Feb 8. Eur Spine J. 2011. PMID: 21301893 Free PMC article. - The BioFlex System as a Dynamic Stabilization Device : Does It Preserve Lumbar Motion?
Zhang HY, Park JY, Cho BY. Zhang HY, et al. J Korean Neurosurg Soc. 2009 Nov;46(5):431-6. doi: 10.3340/jkns.2009.46.5.431. Epub 2009 Nov 30. J Korean Neurosurg Soc. 2009. PMID: 20041052 Free PMC article. - [Effect of lumbar hybrid instrumentation and rigid fusion on the treated and the adjacent segments. A biomechanical study].
Wiedenhöfer B, Akbar M, Fürstenberg CH, Carstens C, Hemmer S, Schilling C. Wiedenhöfer B, et al. Orthopade. 2011 Feb;40(2):169-77. doi: 10.1007/s00132-010-1717-y. Orthopade. 2011. PMID: 21279503 German. - Short-term effects of a dynamic neutralization system (Dynesys) for multi-segmental lumbar disc herniation.
Wang Q, Liu J, Shi Y, Chen Y, Yu H, Ma J, Ren W, Yang H, Wang H, Xiang L. Wang Q, et al. Eur Spine J. 2016 May;25(5):1409-1416. doi: 10.1007/s00586-015-4307-1. Epub 2015 Nov 17. Eur Spine J. 2016. PMID: 26577393
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials