Comment on time-varying eddy currents effects on diffusion-weighting echo-planar imaging - PubMed (original) (raw)
Comment
Comment on time-varying eddy currents effects on diffusion-weighting echo-planar imaging
Fernando Calamante et al. Neuroimage. 2012.
Abstract
This commentary addresses the recent paper entitled "Dynamic correction of artifacts due to susceptibility effects and time-varying eddy currents in diffusion tensor imaging" (Truong et al., 2011) and discusses their findings in relation to previous studies from the late 1990s that showed equivalent findings regarding the complex time-varying nature of eddy currents during the imaging readout. The similarities between the findings from these studies are highlighted, and their possible implications for present commonly used eddy current reduction methods are discussed.
Copyright © 2011 Elsevier Inc. All rights reserved.
Comment on
- Dynamic correction of artifacts due to susceptibility effects and time-varying eddy currents in diffusion tensor imaging.
Truong TK, Chen NK, Song AW. Truong TK, et al. Neuroimage. 2011 Aug 15;57(4):1343-7. doi: 10.1016/j.neuroimage.2011.06.008. Epub 2011 Jun 14. Neuroimage. 2011. PMID: 21689763 Free PMC article.
Similar articles
- Dynamic correction of artifacts due to susceptibility effects and time-varying eddy currents in diffusion tensor imaging.
Truong TK, Chen NK, Song AW. Truong TK, et al. Neuroimage. 2011 Aug 15;57(4):1343-7. doi: 10.1016/j.neuroimage.2011.06.008. Epub 2011 Jun 14. Neuroimage. 2011. PMID: 21689763 Free PMC article. - PSF mapping-based correction of eddy-current-induced distortions in diffusion-weighted echo-planar imaging.
In MH, Posnansky O, Speck O. In MH, et al. Magn Reson Med. 2016 May;75(5):2055-63. doi: 10.1002/mrm.25746. Epub 2015 Jun 22. Magn Reson Med. 2016. PMID: 26096666 - Correction for Eddy Current-Induced Echo-Shifting Effect in Partial-Fourier Diffusion Tensor Imaging.
Truong TK, Song AW, Chen NK. Truong TK, et al. Biomed Res Int. 2015;2015:185026. doi: 10.1155/2015/185026. Epub 2015 Aug 30. Biomed Res Int. 2015. PMID: 26413505 Free PMC article. - High-resolution diffusion imaging using a radial turbo-spin-echo sequence: implementation, eddy current compensation, and self-navigation.
Seifert MH, Jakob PM, Jellus V, Haase A, Hillenbrand C. Seifert MH, et al. J Magn Reson. 2000 Jun;144(2):243-54. doi: 10.1006/jmre.2000.2041. J Magn Reson. 2000. PMID: 10828192 - Geometric distortion correction of high-resolution 3 T diffusion tensor brain images.
Ardekani S, Sinha U. Ardekani S, et al. Magn Reson Med. 2005 Nov;54(5):1163-71. doi: 10.1002/mrm.20651. Magn Reson Med. 2005. PMID: 16187289 Clinical Trial.
Cited by
- An integrated approach to correction for off-resonance effects and subject movement in diffusion MR imaging.
Andersson JLR, Sotiropoulos SN. Andersson JLR, et al. Neuroimage. 2016 Jan 15;125:1063-1078. doi: 10.1016/j.neuroimage.2015.10.019. Epub 2015 Oct 20. Neuroimage. 2016. PMID: 26481672 Free PMC article.