Relationship between the use of new or used computed tomography scanners and image retake rates in South Korea
Jae-Seok Hong et al. Acta Radiol. 2013 May.
Abstract
Background: The use of high-cost imaging has increased worldwide, contributing to increased healthcare expenditures. Without proper quality verification, the installation of used imaging equipment may lead to wasteful increases in cost due to ineffective and poor-quality imaging that requires repeat scans.
Purpose: To examine the relationship between the use of new or used computed tomography (CT) scanners and image retake rates to evaluate the comparative quality of used and new CT scanners.
Material and methods: This was a retrospective cohort study of patients who first underwent CT from January 1 to June 30, 2008 (n = 258,572). Data were obtained by linking the Health Care Institution Registration Data with the Korean National Health Insurance Claims Database. Image retake rates within 30, 60, 90, and 180 days after the first CT scan were calculated, and differences in the image retake rate by new versus used CT scanners were assessed.
Results: After adjusting for confounders, use of a used CT scanner for the first scan increased the odds of retake within 30 days (odds ratio [OR]: 1.34, 95% confidence interval [CI]: 1.22-1.48), 60 days (OR: 1.59, 95% CI: 1.47-1.72), 90 days (OR: 1.48, 95% CI: 1.38-1.59), and 180 days (OR: 1.38, 95% CI: 1.30-1.46) compared with use of a new CT scanner. Such results were evident in scans of the spine, abdomen, chest, and face or skull base.
Conclusion: The quality control associated with import of used CT scanners should be improved. Moreover, regular and detailed quality inspections of used CT scanners currently operating in healthcare institutions are necessary.
Keywords: Computed tomography scanner; medical devices; quality of healthcare.
© 2013 The Foundation Acta Radiologica.
Comment in
- Evaluating the relative usefulness of old radiological equipment.
Abildgaard A. Abildgaard A. Acta Radiol. 2013 May;54(4):363. doi: 10.1177/0284185113479332. Acta Radiol. 2013. PMID: 24879876 No abstract available.
Similar articles
- Single-detector helical CT in PET-CT: assessment of image quality.
Römer W, Chung M, Chan A, Townsend DW, Torok F, McCook B, Federle MP, Avril N. Römer W, et al. AJR Am J Roentgenol. 2004 Jun;182(6):1571-7. doi: 10.2214/ajr.182.6.1821571. AJR Am J Roentgenol. 2004. PMID: 15150012 - Coronal reformations of the chest on 64-row multi-detector row CT: evaluation of image quality in comparison with 16-, 8- and 4-row multi-detector row CT.
Nishino M, Kubo T, Kataoka ML, Raptopoulos V, Hatabu H. Nishino M, et al. Eur J Radiol. 2006 Aug;59(2):231-7. doi: 10.1016/j.ejrad.2006.02.009. Epub 2006 Mar 29. Eur J Radiol. 2006. PMID: 16567075 - GE HiSpeed advantage computed tomography scanner.
[No authors listed] [No authors listed] Health Devices. 1997 Dec;26(12):457-70. Health Devices. 1997. PMID: 9416473 Review. - CT radiation dose and image quality.
Payne JT. Payne JT. Radiol Clin North Am. 2005 Nov;43(6):953-62, vii. doi: 10.1016/j.rcl.2005.07.002. Radiol Clin North Am. 2005. PMID: 16253656 Review.
Cited by
- Clinical and CT characteristics of Xpert MTB/RIF-negative pulmonary tuberculosis.
Lee HN, Kim JI, Kim YH. Lee HN, et al. PLoS One. 2021 May 3;16(5):e0250616. doi: 10.1371/journal.pone.0250616. eCollection 2021. PLoS One. 2021. PMID: 33939723 Free PMC article. - Excess of Radiation Burden for Young Testicular Cancer Patients using Automatic Exposure Control and Contrast Agent on Whole-body Computed Tomography Imaging.
Niiniviita H, Kulmala J, Pölönen T, Määttänen H, Järvinen H, Salminen E. Niiniviita H, et al. Radiol Oncol. 2017 Feb 26;51(2):235-240. doi: 10.1515/raon-2017-0012. eCollection 2017 Jun. Radiol Oncol. 2017. PMID: 28740460 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources