Controlled attenuation parameter for the detection of steatosis severity in chronic liver disease: a meta-analysis of diagnostic accuracy

Ke-Qing Shi et al. J Gastroenterol Hepatol. 2014 Jun.

Abstract

Background and aim: Controlled attenuation parameter (CAP) is a novel ultrasound-based elastography method for detection of steatosis severity. This meta-analysis aimed to assess the performance of CAP.

Methods: PubMed, the Cochrane Library, and the Web of Knowledge were searched to find studies, published in English, relating to accuracy evaluations of CAP for detecting stage 1 (S1), stage 2 (S2), or stage 3 (S3) hepatic steatosis which was diagnosed by liver biopsy. Sensitivities, specificities, and hierarchical summary receiver operating characteristic (HSROC) curves were used to examine CAP performance. The clinical utility of CAP was also evaluated.

Results: Nine studies, with 11 cohorts were analyzed. The summary sensitivities and specificities values were 0.78 (95% confidence interval [CI], 0.69-0.84) and 0.79 (95% CI, 0.68-0.86) for ≥ S1, 0.85 (95% CI, 0.74-0.92) and 0.79 (95% CI, 0.71-0.85) for ≥ S2, and 0.83 (95% CI, 0.76-0.89) and 0.79 (95% CI, 0.68-0.87) for ≥ S3. The HSROCs were 0.85 (95% CI, 0.81-88) for ≥ S1, 0.88 (95% CI, 0.85-0.91) for ≥ S2, and 0.87 (95% CI, 0.84-0.90) for ≥ S3. Following a "positive" measurement (over the threshold value) for ≥ S1, ≥ S2, and ≥ S3, the corresponding post-test probabilities for the presence of steatosis (pretest probability was 50%) were 78%, 80% and 80%, respectively; if the values were below these thresholds ("negative" results), the post-test probabilities were 22%, 16%, and 17%, respectively.

Conclusions: CAP has good sensitivity and specificity for detecting hepatic steatosis; however, based on a meta-analysis, CAP was limited in their accuracy of steatosis, which precluded widespread use in clinical practice.

Keywords: FibroScan; controlled attenuation parameter; hepatic steatosis; sensitivity; specificity.

© 2014 Journal of Gastroenterology and Hepatology Foundation and Wiley Publishing Asia Pty Ltd.