The optimum placement-site of the electrode in electric pulp testing of the 12 anterior teeth
I B Bender et al. J Am Dent Assoc. 1989 Mar.
Abstract
The present study of 2,387 recordings of 12 anterior teeth in 53 patients indicates that the incisal edge is the optimal placement-site for the electric pulp tester to determine the lowest response threshold. The results show significant individual variations in the lowest threshold responses of the cervical-third, middle-third, incisal one-third, and incisal edge sites on a tooth, with a confidence level of 99%, according to the analysis of variance. The maxillary teeth gave a higher response threshold than the mandibular teeth and different types of teeth (canines and incisors) had statistically significant different response thresholds. The application of the electric pulp tester to the incisal-edge region with exposed dentin produced the most significant decrease in the threshold response.
Similar articles
- Appropriate electrode placement site for electric pulp testing of anterior teeth in Nigerian adults: a clinical study.
Udoye CI, Jafarzadeh H, Okechi UC, Aguwa EN. Udoye CI, et al. J Oral Sci. 2010 Jun;52(2):287-92. doi: 10.2334/josnusd.52.287. J Oral Sci. 2010. PMID: 20587955 - Appropriate electrode placement site of electric pulp tester for the premolars: a clinical study.
Bargale SD, Davangere Padmanabh SK. Bargale SD, et al. J Indian Soc Pedod Prev Dent. 2015 Apr-Jun;33(2):138-42. doi: 10.4103/0970-4388.155128. J Indian Soc Pedod Prev Dent. 2015. PMID: 25872633 - Electrode Placing Sites affect Pulp Vitality Test of Human Incisors and Premolars.
Tian SY, Tang L, Zheng CY. Tian SY, et al. Chin J Dent Res. 2017;20(4):225-230. doi: 10.3290/j.cjdr.a39222. Chin J Dent Res. 2017. PMID: 29181460 - Does dental fluorosis affect electric pulp test response and/or optimum probe placement site?
Idon PI, Nwamba PN, Udoye CI, Jafarzadeh H. Idon PI, et al. Aust Endod J. 2021 Aug;47(2):307-313. doi: 10.1111/aej.12479. Epub 2021 Jan 13. Aust Endod J. 2021. PMID: 33438314 - Probe placement during electric pulp-testing procedures.
Jacobson JJ. Jacobson JJ. Oral Surg Oral Med Oral Pathol. 1984 Aug;58(2):242-7. doi: 10.1016/0030-4220(84)90145-2. Oral Surg Oral Med Oral Pathol. 1984. PMID: 6592520
Cited by
- Comparison of the vitality test with sensitivity tests in mature and immature teeth: clinical trial.
Çağırır Dindaroğlu F, Özay Güngör N. Çağırır Dindaroğlu F, et al. BMC Oral Health. 2024 May 27;24(1):613. doi: 10.1186/s12903-024-04317-3. BMC Oral Health. 2024. PMID: 38802767 Free PMC article. Clinical Trial. - Identification of the Ideal Reference Site and Pain Threshold Values for the Placement of Electric Pulp Testers in Permanent Upper and Lower Anterior Teeth: A Cross-Sectional Study.
Saikiran KV, Gurunathan D, Gayathri E, Nuvvula S. Saikiran KV, et al. Cureus. 2024 Apr 12;16(4):e58156. doi: 10.7759/cureus.58156. eCollection 2024 Apr. Cureus. 2024. PMID: 38741880 Free PMC article. - The effect of orthodontic tooth movement on the sensitivity of dental pulp: A systematic review and meta-analysis.
Golež A, Ovsenik M, Cankar K. Golež A, et al. Heliyon. 2023 Mar 24;9(4):e14621. doi: 10.1016/j.heliyon.2023.e14621. eCollection 2023 Apr. Heliyon. 2023. PMID: 37025792 Free PMC article. - Comparison of pulp sensibility test responses in normotensive and hypertensive individuals: A clinical study.
Saklecha P, Kishan KV, Shroff MG. Saklecha P, et al. J Conserv Dent. 2022 Sep-Oct;25(5):526-530. doi: 10.4103/jcd.jcd_105_22. Epub 2022 Sep 12. J Conserv Dent. 2022. PMID: 36506632 Free PMC article.