Abstract
We describe a simple, reproducible, and generally applicable method to assess the performance of log amplifiers by using a fluorescent sample that provides multiple peaks of different intensities. The channel differences between multiple peaks are used to evaluate the logarithmic behavior of the fluorescence signal amplifier on the flow cytometer. A calibration curve can be created to correct the channel numbers for deviations from true logarithmic behavior and then convert data into relative linear intensities. By using these linear fluorescent intensities, we compared the capacity of different antisera against HIV-1 (human immunodeficiency virus type 1) peptides to inhibit the binding of HIV-1 to CEM, a CD4-positive T-cell line. A wide range of applications for this calibration procedure can be envisioned and the method is valuable for monitoring instrument performance over time.
Similar articles
- Determination of linear fluorescence intensities from flow cytometric data accumulated with logarithmic amplifiers.
Muirhead KA, Schmitt TC, Muirhead AR. Muirhead KA, et al. Cytometry. 1983 Jan;3(4):251-6. doi: 10.1002/cyto.990030404. Cytometry. 1983. PMID: 6822146 - A simple and rapid method for determining the linearity of a flow cytometer amplification system.
Bagwell CB, Baker D, Whetstone S, Munson M, Hitchcox S, Ault KA, Lovett EJ. Bagwell CB, et al. Cytometry. 1989 Nov;10(6):689-94. doi: 10.1002/cyto.990100604. Cytometry. 1989. PMID: 2582958 - Computing the central location of immunofluorescence distributions: logarithmic data transformations are not always appropriate.
Coder DM, Redelman D, Vogt RF. Coder DM, et al. Cytometry. 1994 Jun 15;18(2):75-8. doi: 10.1002/cyto.990180204. Cytometry. 1994. PMID: 7924701 Review. - Calibration of flow cytometry detector systems.
Durand RE. Durand RE. Cytometry. 1981 Nov;2(3):192-3. doi: 10.1002/cyto.990020312. Cytometry. 1981. PMID: 7297355 - Quantitative flow cytometry.
Schwartz A, Fernández-Repollet E. Schwartz A, et al. Clin Lab Med. 2001 Dec;21(4):743-61. Clin Lab Med. 2001. PMID: 11770285 Review.
Cited by
- Ion Doped Hollow Silica Nanoparticles as Promising Oligonucleotide Delivery Systems to Mesenchymal Stem Cells.
Trayford C, Ibrahim DM, van Rijt S. Trayford C, et al. Int J Nanomedicine. 2024 Sep 20;19:9741-9755. doi: 10.2147/IJN.S461167. eCollection 2024. Int J Nanomedicine. 2024. PMID: 39329032 Free PMC article. - Relationship among serum levels of IL-6, sIL-6R, s gp130 and CD126 on T-cell in HIV-1 infected and uninfected men participating in the Los Angeles Multi-Center AIDS Cohort Study.
Aziz N, Shih R, Alexopoulos N, Jamieson BD, Mimiaga MJ, Martinez-Maza O, Detels R. Aziz N, et al. PLoS One. 2023 Oct 9;18(10):e0290702. doi: 10.1371/journal.pone.0290702. eCollection 2023. PLoS One. 2023. PMID: 37812611 Free PMC article. - Assessment of Telomere Length, Phenotype, and DNA Content.
Kelesidis T, Schmid I. Kelesidis T, et al. Curr Protoc Cytom. 2017 Jan 5;79:7.26.1-7.26.23. doi: 10.1002/cpcy.12. Curr Protoc Cytom. 2017. PMID: 28055113 Free PMC article. - Assessment of telomere length, phenotype, and DNA content.
Schmid I, Jamieson BD. Schmid I, et al. Curr Protoc Cytom. 2004 Sep;Chapter 7:Unit 7.26. doi: 10.1002/0471142956.cy0726s29. Curr Protoc Cytom. 2004. PMID: 18770803 Free PMC article. - Determination of DNA content of aquatic bacteria by flow cytometry.
Button DK, Robertson BR. Button DK, et al. Appl Environ Microbiol. 2001 Apr;67(4):1636-45. doi: 10.1128/AEM.67.4.1636-1645.2001. Appl Environ Microbiol. 2001. PMID: 11282616 Free PMC article.