High-efficiency liquid-scintillation counting of 14C-labelled material in aqueous solution and determination of specific activity of labelled proteins - PubMed (original) (raw)

High-efficiency liquid-scintillation counting of 14C-labelled material in aqueous solution and determination of specific activity of labelled proteins

T C Hall et al. Biochem J. 1965 Sep.

Abstract

1. Inexpensive scintillation mixtures are described which enable the detection of as little as 40mumuc of (14)C in aqueous solution with an efficiency of counting of over 80%. 2. A rapid method for the counting of alkaline, acidic and neutral aqueous solutions of up to 1ml. volume is described. Ethanol or 2-ethoxyethanol is used as blending agent. 3. The scintillation counting of alkaline solutions is applied to the accurate determination of the specific activity of (14)C-labelled proteins from plant tissues. 4. Attention has been paid to the importance of a standardized washing procedure for the removal of all traces of radioactive material from glassware.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Klin Wochenschr. 1955 Oct 1;33(37-38):879-84 - PubMed
    1. Biochem J. 1964 Apr;91(1):155-61 - PubMed
    1. Arch Biochem Biophys. 1962 Mar;96:459-64 - PubMed
    1. J Biol Chem. 1952 Apr;195(2):549-65 - PubMed
    1. Biochem J. 1959 Oct;73:287-92 - PubMed

MeSH terms

Substances

LinkOut - more resources