Is mode switching beneficial? A randomized study in patients with paroxysmal atrial tachyarrhythmias - PubMed (original) (raw)
Clinical Trial
. 1997 Aug;30(2):496-504.
doi: 10.1016/s0735-1097(97)00162-9.
Affiliations
- PMID: 9247524
- DOI: 10.1016/s0735-1097(97)00162-9
Free article
Clinical Trial
Is mode switching beneficial? A randomized study in patients with paroxysmal atrial tachyarrhythmias
K Kamalvand et al. J Am Coll Cardiol. 1997 Aug.
Free article
Abstract
Objectives: We sought 1) to compare three pacing modalities-DDDR with mode switching (DM), DDDR with conventional upper rate behavior (DR) and VVIR (VR)-in patients with a history of atrial tachyarrhythmias, and 2) to assess the efficacy of six mode-switching algorithms.
Background: A history of atrial tachyarrhythmias has been a relative contraindication to dual-chamber pacing. Several mode-switching algorithms have recently been developed to prevent rapid tracking of atrial tachyarrhythmias.
Methods: Forty-eight patients (mean age 64 years, 58% male) with a history of atrial tachyarrhythmias and heart block had a DM pacemaker implanted. Pacemakers were programmed to DM, DR and VR modes for 4 weeks each in a randomized crossover design. All subjects used a patient-activated electrocardiographic (ECG) recorder throughout the study and additionally underwent ambulatory ECG monitoring and a treadmill exercise test in each mode. They completed three symptom questionnaires at the end of each pacing period. At the end of the study, patients chose their preferred pacing period.
Results: DM was significantly better than VR mode objectively (exercise time DM 8.1 min, VR 7.0 min, p < 0.01) and subjectively (perceived well-being DM 69, VR 51, p < 0.001; functional class DM 2.2, VR 2.5, p < 0.05; subjective symptom score DM 21.2, VR 26.8, p = 0.01). Patient-perceived well-being was significantly better with DM than with DR mode (DM 69, DR 60, p = 0.02). DM mode was the preferred pacing period (DM 51%, DR 14%, VR 14%). Early termination of pacing because of adverse symptoms was requested by 33% of patients during VR, 19% during DR but only 3% during DM mode. A higher proportion of patients with a fast mode-switching device preferred DM mode (fast 55%, slow 49%), whereas no patients with a fast mode-switching device chose VR as the preferred mode (fast 0%, slow 19%). In the subgroup of patients who had had atrioventricular node ablation, DM was also preferred to VR mode (DM 53%, VR 27%). Overall, there were only two cases of inappropriate mode switching and one case of inappropriate tracking of an atrial tachyarrhythmia.
Conclusions: DM is the pacing mode of choice of patients with paroxysmal atrial tachyarrhythmias. With optimal programming, inappropriate mode switching and tracking of atrial tachyarrhythmias was very uncommon.
Similar articles
- [Mode-switching algorithms: programming and usefulness].
Israel CW. Israel CW. Herz. 2001 Feb;26(1):2-17. doi: 10.1007/pl00002003. Herz. 2001. PMID: 11258106 German. - DDD(R) pacing with automatic mode switch in patients with paroxysmal atrial fibrillation following AV nodal ablation.
Schuchert A, van Langen H, Michels K, Meinertz T. Schuchert A, et al. Cardiology. 1997 Jul-Aug;88(4):323-7. doi: 10.1159/000177353. Cardiology. 1997. PMID: 9197425 - DDIR versus VVIR pacing in patients with paroxysmal atrial tachyarrhythmias.
Vanerio G, Maloney JD, Pinski SL, Simmons TW, Castle LW, Trohman RG, Wilkoff BL. Vanerio G, et al. Pacing Clin Electrophysiol. 1991 Nov;14(11 Pt 1):1630-8. doi: 10.1111/j.1540-8159.1991.tb02739.x. Pacing Clin Electrophysiol. 1991. PMID: 1721153 - Dual demand pacing using retriggerable refractory periods for ventricular rate control during paroxysmal supraventricular tachyarrhythmias in patients with dual chamber pacemakers.
Jayaprakash S, Sparks PB, Kalman JM, Mond HG. Jayaprakash S, et al. Pacing Clin Electrophysiol. 2000 Jul;23(7):1156-63. doi: 10.1111/j.1540-8159.2000.tb00917.x. Pacing Clin Electrophysiol. 2000. PMID: 10914373 Review. - Analysis of mode switching algorithms in dual chamber pacemakers.
Israel CW. Israel CW. Pacing Clin Electrophysiol. 2002 Mar;25(3):380-93. doi: 10.1046/j.1460-9592.2002.00380.x. Pacing Clin Electrophysiol. 2002. PMID: 11990674 Review.
Cited by
- Inappropriate mode switching clarified by using a chest radiograph.
Marino B, Jaiswal A, Goldbarg S. Marino B, et al. J Arrhythm. 2015 Aug;31(4):246-8. doi: 10.1016/j.joa.2015.01.001. Epub 2015 Feb 14. J Arrhythm. 2015. PMID: 26336569 Free PMC article. - Predictors of Recurrent Atrial Fibrillation Using Mode Switch Quantification.
Ruisi P, Makaryus JN, Catanzaro JN, Ruisi M, Cedrone A, Makaryus AN, Altman E, Jadonath R, Beldner S. Ruisi P, et al. Cardiol Res. 2013 Oct;4(4-5):135-138. doi: 10.4021/cr292w. Epub 2013 Oct 15. Cardiol Res. 2013. PMID: 28352435 Free PMC article. - Prävalenz von Vorhofarrhythmien bei Schrittmacherpatienten im 6 Monats Follow-up Konzept und erste Ergebnisse der AIDA II Studie.
Jakob MJ, Fischer W, Hartkopf T, Himmrich E, Kluge P, Richter F, Scholz H. Jakob MJ, et al. Herzschrittmacherther Elektrophysiol. 2000 Jan;11 Suppl 1:85-6. doi: 10.1007/BF03042540. Herzschrittmacherther Elektrophysiol. 2000. PMID: 19495656 German. No abstract available. - Automatic mode switching in atrial fibrillation.
Stabile G, De Simone A, Romano E. Stabile G, et al. Indian Pacing Electrophysiol J. 2005 Jul 1;5(3):186-96. Indian Pacing Electrophysiol J. 2005. PMID: 16943867 Free PMC article. - [Pacemaker therapy in patients with atrial fibrillation].
Schuchert A, Meinertz T. Schuchert A, et al. Herz. 1998 Jun;23(4):260-8. doi: 10.1007/BF03044321. Herz. 1998. PMID: 9690113 Review. German.