Performance evaluation of a whole-body PET scanner using the NEMA protocol. National Electrical Manufacturers Association
G Brix et al. J Nucl Med. 1997 Oct.
Free article
Abstract
This study evaluates the performance of the newly developed high-resolution whole-body PET scanner ECAT EXACT HR+.
Methods: The scanner consists of four rings of 72 bismuth germanate block detectors each, covering an axial field of view of 15.5 cm with a patient port of 56.2 cm. A single block detector is divided into an 8 x 8 matrix, giving a total of 32 rings with 576 detectors each. The dimensions of a single detector element are 4.39 x 4.05 x 30 mm3. The scanner is equipped with extendable tungsten septa for two-dimensional two-dimensional measurements, as well as with three 68Ge line sources for transmission scans and daily quality control. The spatial resolution, scatter fraction, count rate, sensitivity, uniformity and accuracy of the implemented correction algorithms were evaluated after the National Electrical Manufacturers Association protocol using the standard acquisition parameters.
Results: The transaxial resolution in the two-dimensional mode is 4.3 mm (4.4 mm) in the center and increases to 4.7 mm (4.8 mm) tangential and to 8.3 mm (8.0 mm) radial at a distance of r = 20 cm from the center. The axial slice width measured in the two-dimensional mode varies between 4.2 and 6.6 mm FWHM over the transaxial field of view. In the three-dimensional mode the average axial resolution varies between 4.1 mm FWHM in the center and 7.8 mm at r = 20 cm. The scatter fraction is 17.1% (32.5%) for a lower energy discriminator level of 350 keV. The maximum true event count rate of 263 (345) kcps was measured at an activity concentration of 142 (26.9) kBq/ml. The total system sensitivity for true events is 5.7 (27.7) cps/Bq/ml. From the uniformity measurements, we obtained a volume variance of 3.9% (5.0%) and a system variance of 1.6% (1.7%). The implemented three-dimensional scatter correction algorithm reveals very favorable properties, whereas the three-dimensional attenuation correction yields slightly inaccurate results in low- and high-density regions.
Conclusion: The ECAT EXACT HR+ has an excellent, nearly isotropic spatial resolution, which is advantageous for brain and small animal studies. While the relatively low slice sensitivity may hamper the capability for performing fast dynamic two-dimensional studies, the scanner offers a sufficient sensitivity and count rate capacity for fully three-dimensional whole-body imaging.
Similar articles
- Performance characteristics of a newly developed PET/CT scanner using NEMA standards in 2D and 3D modes.
Mawlawi O, Podoloff DA, Kohlmyer S, Williams JJ, Stearns CW, Culp RF, Macapinlac H; National Electrical Manufacturers Association. Mawlawi O, et al. J Nucl Med. 2004 Oct;45(10):1734-42. J Nucl Med. 2004. PMID: 15471842 - Performance characteristics of a new 3-dimensional continuous-emission and spiral-transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU 2-2001 standard.
Matsumoto K, Kitamura K, Mizuta T, Tanaka K, Yamamoto S, Sakamoto S, Nakamoto Y, Amano M, Murase K, Senda M. Matsumoto K, et al. J Nucl Med. 2006 Jan;47(1):83-90. J Nucl Med. 2006. PMID: 16391191 - Performance characteristics of a whole-body PET scanner.
DeGrado TR, Turkington TG, Williams JJ, Stearns CW, Hoffman JM, Coleman RE. DeGrado TR, et al. J Nucl Med. 1994 Aug;35(8):1398-406. J Nucl Med. 1994. PMID: 8046501 - PET instrumentation: what are the limits?
Budinger TF. Budinger TF. Semin Nucl Med. 1998 Jul;28(3):247-67. doi: 10.1016/s0001-2998(98)80030-5. Semin Nucl Med. 1998. PMID: 9704366 Review. - Physical principles and technology of clinical PET imaging.
Townsend DW. Townsend DW. Ann Acad Med Singap. 2004 Mar;33(2):133-45. Ann Acad Med Singap. 2004. PMID: 15098626 Review.
Cited by
- The brain's "dark energy" puzzle: How strongly is glucose metabolism linked to resting-state brain activity?
Volpi T, Silvestri E, Aiello M, Lee JJ, Vlassenko AG, Goyal MS, Corbetta M, Bertoldo A. Volpi T, et al. J Cereb Blood Flow Metab. 2024 Aug;44(8):1433-1449. doi: 10.1177/0271678X241237974. Epub 2024 Mar 5. J Cereb Blood Flow Metab. 2024. PMID: 38443762 - Harmonization of [11C]raclopride brain PET images from the HR+ and HRRT: method development and validation in human subjects.
Hoye J, Toyonaga T, Zakiniaeiz Y, Stanley G, Hampson M, Morris ED. Hoye J, et al. EJNMMI Phys. 2022 Apr 13;9(1):27. doi: 10.1186/s40658-022-00457-z. EJNMMI Phys. 2022. PMID: 35416555 Free PMC article. - Fragile X Mental Retardation Protein and Cerebral Expression of Metabotropic Glutamate Receptor Subtype 5 in Men with Fragile X Syndrome: A Pilot Study.
Brašić JR, Goodman JA, Nandi A, Russell DS, Jennings D, Barret O, Martin SD, Slifer K, Sedlak T, Mathur AK, Seibyl JP, Berry-Kravis EM, Wong DF, Budimirovic DB. Brašić JR, et al. Brain Sci. 2022 Feb 26;12(3):314. doi: 10.3390/brainsci12030314. Brain Sci. 2022. PMID: 35326270 Free PMC article. - Role of amygdala in stress-induced upregulation of airway IL-1 signaling in asthma.
Rosenkranz MA, Esnault S, Gresham L, Davidson RJ, Christian BT, Jarjour NN, Busse WW. Rosenkranz MA, et al. Biol Psychol. 2022 Jan;167:108226. doi: 10.1016/j.biopsycho.2021.108226. Epub 2021 Nov 17. Biol Psychol. 2022. PMID: 34800561 Free PMC article. - Principal Component Analysis of Striatal and Extrastriatal D2 Dopamine Receptor Positron Emission Tomography in Manganese-Exposed Workers.
Criswell SR, Searles Nielsen S, Dlamini WW, Warden MN, Perlmutter JS, Sheppard L, Moerlein SM, Lenox-Krug J, Checkoway H, Racette BA. Criswell SR, et al. Toxicol Sci. 2021 Jul 16;182(1):132-141. doi: 10.1093/toxsci/kfab045. Toxicol Sci. 2021. PMID: 33881537 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources