Selective interference with pacemaker activity by electrical dental devices
C S Miller et al. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1998 Jan.
Abstract
Objective: We sought to determine whether electromagnetic interference with cardiac pacemakers occurs during the operation of contemporary electrical dental equipment.
Study design: Fourteen electrical dental devices were tested in vitro for their ability to interfere with the function of two Medtronics cardiac pacemakers (one a dual-chamber, bipolar Thera 7942 pacemaker, the other a single-chamber, unipolar Minix 8340 pacemaker). Atrial and ventricular pacemaker output and electrocardiographic activity were monitored by means of telemetry with the use of a Medtronics 9760/90 programmer.
Results: Atrial and ventricular pacing were inhibited by electromagnetic interference produced by the electrosurgical unit up to a distance of 10 cm, by the ultrasonic bath cleaner up to 30 cm, and by the magnetorestrictive ultrasonic scalers up to 37.5 cm. In contrast, operation of the amalgamator, electric pulp tester, composite curing light, dental handpieces, electric toothbrush, microwave oven, dental chair and light, ENAC ultrasonic instrument, radiography unit, and sonic scaler did not alter pacing rate or rhythm.
Conclusions: These results suggest that certain electrosurgical and ultrasonic instruments may produce deleterious effects in medically fragile patients with cardiac pacemakers.
Similar articles
- [Electromagnetic interference of electrical dental equipment with cardiac pacemakers].
Brand HS, van der Hoeff EV, Schrama TA, Entjes ML, van Nieuw AA. Brand HS, et al. Ned Tijdschr Tandheelkd. 2007 Sep;114(9):373-6. Ned Tijdschr Tandheelkd. 2007. PMID: 17937372 Dutch. - Interference of cardiac pacemaker and implantable cardioverter-defibrillator activity during electronic dental device use.
Roedig JJ, Shah J, Elayi CS, Miller CS. Roedig JJ, et al. J Am Dent Assoc. 2010 May;141(5):521-6. doi: 10.14219/jada.archive.2010.0224. J Am Dent Assoc. 2010. PMID: 20436099 - Interference of electrical dental equipment with implantable cardioverter-defibrillators.
Brand HS, Entjes ML, Nieuw Amerongen AV, van der Hoeff EV, Schrama TA. Brand HS, et al. Br Dent J. 2007 Nov 24;203(10):577-9. doi: 10.1038/bdj.2007.1043. Br Dent J. 2007. PMID: 18037843 - Ultrasonic dental scaler: associated hazards.
Trenter SC, Walmsley AD. Trenter SC, et al. J Clin Periodontol. 2003 Feb;30(2):95-101. doi: 10.1034/j.1600-051x.2003.00276.x. J Clin Periodontol. 2003. PMID: 12622849 Review. - Electrical interference of pacemaker activity by electrical dental devices: a short review.
Antunovic DM. Antunovic DM. J N Z Soc Periodontol. 2007;(90):7-11. J N Z Soc Periodontol. 2007. PMID: 18330208 Review. No abstract available.
Cited by
- Dental Management Considerations for Patients with Cardiovascular Disease-A Narrative Review.
Gupta K, Kumar S, Anand Kukkamalla M, Taneja V, Syed GA, Pullishery F, Zarbah MA, Alqahtani SM, Alobaoid MA, Chaturvedi S. Gupta K, et al. Rev Cardiovasc Med. 2022 Jul 20;23(8):261. doi: 10.31083/j.rcm2308261. eCollection 2022 Aug. Rev Cardiovasc Med. 2022. PMID: 39076626 Free PMC article. Review. - Neuromuscular electrical stimulation for a dysphagic stroke patient with cardiac pacemaker using magnet mode change: A case report.
Kim M, Park JK, Lee JY, Kim MJ. Kim M, et al. World J Clin Cases. 2021 Jul 6;9(19):5313-5318. doi: 10.12998/wjcc.v9.i19.5313. World J Clin Cases. 2021. PMID: 34307583 Free PMC article. - Can active signals of cellphone interfere with electronic working length determination of a root canal in a dental clinic? An in vivo study.
Gohil UK, Parekh VV, Kinariwala N, Oza KM, Somani MC. Gohil UK, et al. J Conserv Dent. 2017 May-Jun;20(3):170-173. doi: 10.4103/JCD.JCD_252_16. J Conserv Dent. 2017. PMID: 29279620 Free PMC article. - Management of Patients With Cardiovascular Implantable Electronic Devices in Dental, Oral, and Maxillofacial Surgery.
Tom J. Tom J. Anesth Prog. 2016 Summer;63(2):95-104. doi: 10.2344/0003-3006-63.2.95. Anesth Prog. 2016. PMID: 27269668 Free PMC article. Review. - Dental management of a patient fitted with subcutaneous Implantable Cardioverter Defibrillator device and concomitant warfarin treatment.
Shah AH, Khalil HS, Kola MZ. Shah AH, et al. Saudi Dent J. 2015 Jul;27(3):165-70. doi: 10.1016/j.sdentj.2014.11.015. Epub 2015 Apr 23. Saudi Dent J. 2015. PMID: 26236132 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical