Modelling of a holographic interferometry based calorimeter for radiation dosimetry (original) (raw)

NASA/ADS

;

Abstract

In this research work, a model for predicting the behaviour of holographic interferometry based calorimeters for radiation dosimetry is introduced. Using this technique for radiation dosimetry via measuring the variations of refractive index due to energy deposition of radiation has several considerable advantages such as extreme sensitivity and ability of working without normally used temperature sensors that disturb the radiation field. We have shown that the results of our model are in good agreement with the experiments performed by other researchers under the same conditions. This model also reveals that these types of calorimeters have the additional and considerable merits of transforming the dose distribution to a set of discernible interference fringes.

Publication:

Nuclear Instruments and Methods in Physics Research A

Pub Date:

August 2017

DOI:

10.1016/j.nima.2017.05.019

Bibcode:

2017NIMPA.864...40B

Keywords: