In vitro head-to-head comparison of anticoagulation properties of two heparin brands in a human blood miniature mock loop (original) (raw)
2018, Interactive CardioVascular and Thoracic Surgery
OBJECTIVES: The first aim was the development of a human blood miniature mock-loop system consisting of 2 identical extracorporeal circuits, which enable systematic head-to-head comparisons of test substances. In a second step, we evaluated the suitability of the mockloop system, by comparing 2 different brands of heparin (ROTEXMEDICA vs B.BRAUN), which have showed different anticoagulation capacities in the clinic. METHODS: For 1 experiment (18 in total), blood of the same healthy human donor was divided into 2 portions (2 Â 50 ml), heparinized with 37.5 IUÁml-1 of the competing test substances and diluted to a haematocrit value of 20-25%. Each mock loop was filled with 70 ml, and in vivo heparin degradation was simulated in 3 different groups by protamine application, representing 0%, 50% and 100% heparin antagonization. At baseline, 5, 60, 120, 240 and 360 min, blood samples were taken to perform thromboelastometry, flow cytometry, haemolysis and general haemostasis analysis.