Magnetic resonance imaging of artificial lumbar disks: safety and metal artifacts (original) (raw)

Chinese medical journal, 2009

Abstract

This study was to investigate the safety of two types of commercially available lumbar artificial discs (CHARITE and PRODISC-L) during a magnetic resonance imaging (MRI) procedure in a 1.5-Tesla MR system, and to evaluate the size of metal artifacts on the MR image for different sequences. A 1.5-Tesla clinical MR imaging system was used. The degree of deflection of the endplates of two artificial discs was evaluated by an angle-measurement instrument at the portals of the MRI scanner. The heating effect of the radio frequency (RF) magnetic field was evaluated by using "worst-case" imaging sequences on a human cadaver implanted with an artificial lumbar disc at the L5/S1 intervertebral disc location. The temperatures of the tissue adjacent to the implant, and of the L4/L5 intervertebral disc (used as a control) were measured, respectively, using a digital probe thermometer before and after the MRI scan sequence. A rectangular water phantom was designed to evaluate the metal...