An Automated Method for Ozonesonde Calibration: New Insights (original) (raw)

2008

Abstract

An automated method for preparation of the electrochemical concentration cell (ECC) ozonesonde is presented. Development of a computer-controlled system for preparation and calibration of the ECC is an improvement over the manual preparation method, and reduces subjectivity considerably. Preparation measurements in digital form aids analysis of the ECC before release and enhances post-flight data certification. Calibration of ozonesondes over a range of ozone concentrations between 0 mPA and 30 mPA is discussed. This presentation describes the automatic system, gives examples of calibrations. The automated system enables comparison of varying potassium iodide (KI) concentrations that should allow adjustment of earlier ozonesonde data obtained with different KT concentrations used since 1970, i.e., 2, 1.5, 1, and 0.5 percent. Preliminary results indicate ECC accuracy has a strong dependence on the electrolyte concentration and should not be considered linear with altitude.