Computer-stimulated test fitting of an implantable hearing aid using implantable hearing aid using three-dimensional CT scans of the temporal bone: preliminary study (original) (raw)
The American journal of otology, 1994
Abstract
In preparation for future implantation of the implantable middle ear transducer in patients, a method was sought for preoperatively test fitting a model of the device, using computer generated three-dimensional (3-D) temporal bone images derived from spiral computed tomography (CT) data. A 3-D model of the implantable middle ear transducer was designed using NIH Image software on a Macintosh computer. High resolution human temporal bone CT scans were obtained using a spiral CT scanner (Siemens Somatom Plus S). The 3-D transducer model was superimposed onto 3-D reconstructions of the temporal bone using ANALYZE software on a computer graphics workstation (Sun SPARCstation 10), showing the transducer "implanted" in the temporal bone. Measurements were validated using a cadaver temporal bone. This process produced images demonstrating the "fit" of the current transducer design in the mastoid region of the adult temporal bone. It permitted assessment of the proximity...