Neural probe design for reduced tissue encapsulation in CNS (original) (raw)

Evaluating the in vivo glial response to miniaturized parylene cortical probes coated with an ultra-fast degrading polymer to aid insertion

Objective. Despite the feasibility of short-term neural recordings using implantable microelectrodes, attaining reliable, chronic recordings remains a challenge. Most neural recording devices suffer from a long-term tissue response, including gliosis, at the device–tissue interface. It was hypothesized that smaller, more flexible intracortical probes would limit gliosis by providing a better mechanical match with surrounding tissue. Approach. This paper describes the in vivo evaluation of flexible parylene microprobes designed to improve the interface with the adjacent neural tissue to limit gliosis and thereby allow for improved recording longevity. The probes were coated with an ultrafast degrading tyrosine-derived polycarbonate (E5005(2K)) polymer that provides temporary mechanical support for device implantation, yet degrades within 2 h post-implantation. A parametric study of probes of varying dimensions and polymer coating thicknesses were implanted in rat brains. The glial tissue response and neuronal loss were assessed from 72 h to 24 weeks post-implantation via immunohistochemistry. Main results. Experimental results suggest that both probe and polymer coating sizes affect the extent of gliosis. When an appropriate sized coating dimension (100 µm × 100 µm) and small probe (30 µm × 5 µm) was implanted, a minimal post-implantation glial response was observed. No discernible gliosis was detected when compared to tissue where a sham control consisting of a solid degradable polymer shuttle of the same dimensions was inserted. A larger polymer coating (200 µm × 200 µm) device induced a more severe glial response at later time points, suggesting that the initial insertion trauma can affect gliosis even when the polymer shuttle degrades rapidly. A larger degree of gliosis was also observed when comparing a larger sized probe (80 µm × 5 µm) to a smaller probe (30 µm × 5 µm) using the same polymer coating size (100 µm × 100 µm). There was no significant neuronal loss around the implantation sites for most device candidates except the group with largest polymer coating and probe sizes. Significance. These results suggest that: (1) the degree of mechanical trauma at device implantation and mechanical mismatches at the probe-tissue interface affect long term gliosis; (2) smaller, more flexible probes may minimize the glial response to provide improved tissue biocompatibility when used for chronic neural signal recording; and (3) some degree of glial scarring did not significantly affect neuronal distribution around the probe.

Chronic tissue response to carboxymethyl cellulose based dissolvable insertion needle for ultra-small neural probes

2014

Implantable neural electrodes must drastically improve chronic recording stability before they can be translated into long-term human clinical prosthetics. Previous studies suggest that sub-cellular sized and mechanically compliant probes may result in improved tissue integration and recording longevity. However, currently these design features are restricted by the opposing mechanical requirements needed for minimally damaging insertions. We designed a non-cytotoxic, carboxymethylcellulose (CMC) based dissolvable delivery vehicle (shuttle) to provide the mechanical support for insertion of ultra-small, ultra-compliant microfabricated neural probes. Stiff CMC-based shuttles rapidly soften immediately after being placed ∼1 mm above an open craniotomy as they absorb vapors from the brain. To address this, we developed a sophisticated targeting, high speed insertion (∼80 mm/s), and release system to implant these shuttles. After implantation, the goal is for the shuttle to dissolve away leaving only the electrodes behind. Here we show the histology of chronically implanted shuttles of large (300 μm × 125 μm) and small (100 μm × 125 μm) size at discrete time points over 12 weeks. Early time points show the CMC shuttle expanded after insertion as it absorbed moisture from the brain and slowly dissolved. At later time points neuronal cell bodies populate regions within the original shuttle tract. The large CMC shuttles show that the CMC expansion can cause extended secondary damage. On the other hand, the smaller CMC shuttles show limited secondary damage, wound closure by 4 weeks, absence of activated microglia at 12 weeks, as well as evidence suggesting neural regeneration at the implant site. This shuttle, therefore, shows great promise facilitating the implantation of nontraditional ultra-small, and ultra-compliant probes.

Model Neural Prostheses With Integrated Microfluidics: A Potential Intervention Strategy for Controlling Reactive Cell and Tissue Responses

IEEE Transactions on Biomedical Engineering, 2004

Model silicon intracortical probes with microfluidic channels were fabricated and tested to examine the feasibility of using diffusion-mediated delivery to deliver therapeutic agents into the volume of tissue exhibiting reactive responses to implanted devices. Three-dimensional probe structures with microfluidic channels were fabricated using surface micromachining and deep reactive ion etching (DRIE) techniques. In vitro functional tests of devices were performed using fluorescence microscopy to record the transient release of Texas Red labeled transferrin (TR-transferrin) and dextran (TR-dextran) from the microchannels into 1% w/v agarose gel. In vivo performance was characterized by inserting devices loaded with TR-transferrin into the premotor cortex of adult male rats. Brain sections were imaged using confocal microscopy. Diffusion of TR-transferrin into the extracellular space and uptake by cells up to 400 m from the implantation site was observed in brain slices taken 1 h postinsertion. The reactive tissue volume, as indicated by the presence of phosphorylated mitogen-activated protein kinases (MAPKs), was characterized using immunohistochemistry and confocal microscopy. The reactive tissue volume extended 600, 800, and 400 m radially from the implantation site at 1 h, 24 h, and 6 weeks following insertion, respectively. These results indicate that diffusion-mediated delivery can be part of an effective intervention strategy for the treatment of reactive tissue responses around chronically implanted intracortical probes.

Implant Size and Fixation Mode Strongly Influence Tissue Reactions in the CNS

PLoS ONE, 2011

The function of chronic brain machine interfaces depends on stable electrical contact between neurons and electrodes. A key step in the development of interfaces is therefore to identify implant configurations that minimize adverse long-term tissue reactions. To this end, we here characterized the separate and combined effects of implant size and fixation mode at 6 and 12 weeks post implantation in rat (n = 24) cerebral cortex. Neurons and activated microglia and astrocytes were visualized using NeuN, ED1 and GFAP immunofluorescence microscopy, respectively. The contributions of individual experimental variables to the tissue response were quantified. Implants tethered to the skull caused larger tissue reactions than un-tethered implants. Small diameter (50 mm) implants elicited smaller tissue reactions and resulted in the survival of larger numbers of neurons than did large diameter (200 mm) implants. In addition, tethering resulted in an oval-shaped cavity, with a cross-section area larger than that of the implant itself, and in marked changes in morphology and organization of neurons in the region closest to the tissue interface. Most importantly, for implants that were both large diameter and tethered, glia activation was still ongoing 12 weeks after implantation, as indicated by an increase in GFAP staining between week 6 and 12, while this pattern was not observed for un-tethered, small diameter implants. Our findings therefore clearly indicate that the combined small diameter, un-tethered implants cause the smallest tissue reactions.

Intact Histological Characterization of Brain-implanted Microdevices and Surrounding Tissue

Journal of Visualized Experiments, 2013

Research into the design and utilization of brain-implanted microdevices, such as microelectrode arrays, aims to produce clinically relevant devices that interface chronically with surrounding brain tissue. Tissue surrounding these implants is thought to react to the presence of the devices over time, which includes the formation of an insulating "glial scar" around the devices. However, histological analysis of these tissue changes is typically performed after explanting the device, in a process that can disrupt the morphology of the tissue of interest.

Ultra-miniature ultra-compliant neural probes with dissolvable delivery needles: design, fabrication and characterization

Stable chronic functionality of intracortical probes is of utmost importance toward realizing clinical application of brain-machine interfaces. Sustained immune response from the brain tissue to the neural probes is one of the major challenges that hinder stable chronic functionality. There is a growing body of evidence in the literature that highly compliant neural probes with sub-cellular dimensions may significantly reduce the foreign-body response, thereby enhancing long term stability of intracortical recordings. Since the prevailing commercial probes are considerably larger than neurons and of high stiffness, new approaches are needed for

Foreign Body Response to Intracortical Microelectrodes Is Not Altered with Dip-Coating of Polyethylene Glycol (PEG)

Frontiers in Neuroscience, 2017

Poly(ethylene glycol) (PEG) is a frequently used polymer for neural implants due to its biocompatible property. As a follow-up to our recent study that used PEG for stiffening flexible neural probes, we have evaluated the biological implications of using devices dip-coated with PEG for chronic neural implants. Mice (wild-type and CX3CR1-GFP) received bilateral implants within the sensorimotor cortex, one hemisphere with a PEG-coated probe and the other with a non-coated probe for 4 weeks. Quantitative analyses were performed using biomarkers for activated microglia/macrophages, astrocytes, blood-brain barrier leakage, and neuronal nuclei to determine the degree of foreign body response (FBR) resulting from the implanted microelectrodes. Despite its well-known acute anti-biofouling property, we observed that PEG-coated devices caused no significantly different FBR compared to non-coated controls at 4 weeks. A repetition using CX3CR1-GFP mice cohort showed similar results. Our histological findings suggest that there is no significant impact of acute delivery of PEG on the FBR in the long-term, and that temporary increase in the device footprint due to the coating of PEG also does not have a significant impact. Large variability seen within the same treatment group also implies that avoiding large superficial vasculature during implantation is not sufficient to minimize inter-animal variability.

3D Parylene sheath neural probe for chronic recordings

Journal of Neural Engineering, 2013

Objective. Reliable chronic recordings from implanted neural probes remain a significant challenge; current silicon-based and microwire technologies experience a wide range of biotic and abiotic failure modes contributing to loss of signal quality. Approach. A multi-prong alternative strategy with potential to overcome these hurdles is introduced that combines a novel three dimensional (3D), polymer-based probe structure with coatings. Specifically, the Parylene C sheath-based neural probe is coated with neurotrophic and anti-inflammatory factors loaded onto a Matrigel carrier to encourage the ingrowth of neuronal processes for improved recording quality, reduce the immune response, and promote improved probe integration into brain tissue for reliable, long-term implementation compared to its rigid counterparts. Main results. The 3D sheath structure of the probe was formed by thermal molding of a surface micromachined Parylene C microchannel, with electrode sites lining the interior and exterior regions of the lumen. Electrochemical characterization of the probes via cyclic voltammetry and electrochemical impedance spectroscopy was performed and indicated suitable electrode properties for neural recordings (1 kHz electrical impedance of ∼200 k in vitro). A novel introducer tool for the insertion of the compliant polymer probe into neural tissue was developed and validated both in vitro using agarose gel and in vivo in the rat cerebral cortex. In vivo electrical functionality of the Parylene C-based 3D probes and their suitability for recording the neuronal activity over a 28-day period was demonstrated by maintaining the 1 kHz electrical impedance within a functional range (<400 k ) and achieving a reasonably high signal-to-noise ratio for detection of resolvable multi-unit neuronal activity on most recording sites in the probe. Immunohistochemical analysis of the implant site indicated strong correlations between the quality of recorded activity and the neuronal/astrocytic density around the probe. Significance. The provided electrophysiological and immunohistochemical data provide strong support to the viability of the developed probe technology. Furthermore, the obtained data provide insights into further optimization of the probe design, including tip geometry, use of neurotrophic and anti-inflammatory drugs in the Matrigel coating, and placement of the recording sites.

Novel flexible Parylene neural probe with 3D sheath structure for enhancing tissue integration

Lab on a Chip, 2013

A Parylene C neural probe with a three dimensional sheath structure was designed, fabricated, and characterized. Multiple platinum (Pt) electrodes for recording neural signals were fabricated on both inner and outer surfaces of the sheath structure. Thermoforming of Parylene was used to create the three dimensional sheath structures from flat surface micromachined microchannels using solid microwires as molds. Benchtop electrochemical characterization was performed on the thin film Pt electrodes using cyclic voltammetry and electrochemical impedance spectroscopy and showed that electrodes possessed low impedances suitable for neuronal recordings. A procedure for implantation of the neural probe was developed and successfully demonstrated in vitro into an agarose brain tissue model. The electrode-lined sheath will be decorated with eluting neurotrophic factors to promote in vivo neural tissue ingrowth postimplantation. These features will enhance tissue integration and improve recording quality towards realizing reliable chronic neural interfaces.

An ultra-compliant, scalable neural probe with molded biodissolvable delivery vehicle

2012

This paper describes an ultra-compliant parylene-platinum neural probe embedded in a biodissolvable delivery vehicle. High probe compliance is achieved using thin wires (width of 10.0 μm and thickness of 2.7 μm) and by meandering the probe. The insertion of the ultra-compliant probe is achieved by encasing it in a dissolvable delivery vehicle made from molded carboxy-methylcellulose. In vivo implantations of delivery vehicles with 1.5 mm long shanks, widths of 100 μm and 300 μm and a targeted thickness of 135 μm have been done through the dura in the cortex of Sprague-Dawley rats at a speed of 80 mm-s-1. The delivery vehicle becomes a gel over a period of less than three minutes, after which the handling portions of the delivery vehicle are removed leaving the shanks embedded in the brain.