Malposition and revision rates of different imaging modalities for percutaneous iliosacral screw fixation following pelvic fractures: a systematic review and meta-analysis (original) (raw)
2013, Archives of Orthopaedic and Trauma Surgery
of the screw, language of the article English or German. Exclusion criteria were osteoporotic fracture, tumor, reviews, epidemiological studies, biomechanical/cadaveric studies, studies about operative technique. For statistical analysis the random effect model was used. Results A total of 51 studies fulfilled the inclusion requirements describing 2,353 percutaneous screw implantations following pelvic trauma in 1,731 patients. The estimated rate of malposition was 0.1 % for 262 screws using CT navigation. This rate was significantly lower (p < 0.0001) than for the conventional technique with malposition rate of 2.6 % (total 1,832 screws). Using 2D and 3D imagebased navigation and reconstruction techniques, the malposition rate was 1.3 % (total 445 screws). No significance was observed between the conventional and the 2D and 3D image-based navigation and reconstruction techniques. The rates of revision were not statistically significant with 2.7 % (1,832 implantations) in the conventional group, 1.3 % (445 implantations) in the group of 2D and 3D image-based navigation and reconstruction techniques and 0.8 % (262 implantations) using the CT navigation.