Iodine based radiopacity of experimental blood clots for testing of mechanical thrombectomy devices (original) (raw)

2013, Radiology and oncology

Barium sulfate powder used for radiopacity of experimental blood clots (EBCs) for testing mechanical thrombectomy devices (MTD) has negative effects on EBCs mechanical properties. In vitro and in vivo exploration was performed to determine if the iodine based contrast medium will have less negative effects on the EBCs than barium. Fresh blood from 2 swine was used to create fibrinogen enhanced and thrombin initiated EBC in tubes. Iodine radiopacity was achieved by mixing the blood with 65% Iohexol or by soaking the EBCs for 2 or 24 hours in Iohexol. The EBCs opacified with barium served as controls. In vitro study: The EBCs were subjected to four tests, manual elongation, catheter injection, radiopacity and contrast wash out tests. In vivo study: The common carotid arteries of 2 swine were embolized by either barium EBC or EBC soaked for 24 hours in Iohexol. The duration of radiopacity of the different EBCs was compared. The EBCs opacified with Iohexol initially had higher radiopaci...