Accuracy of the Spot and Plusoptix photoscreeners for detection of astigmatism (original) (raw)

Journal of AAPOS : the official publication of the American Association for Pediatric Ophthalmology and Strabismus / American Association for Pediatric Ophthalmology and Strabismus, 2015

Abstract

To evaluate the accuracy of the Spot (V2.0.16) and Plusoptix S12 (ROC4, V6.1.4.0) photoscreeners in detecting astigmatism meeting AAPOS referral criteria in students from a population with high prevalence of astigmatism. Students attending grades 3-8 on the Tohono O'odham reservation were examined. Screening was attempted with both the Spot and Plusoptix photoscreeners. Results were compared to cycloplegic refraction. Screening attempts providing no estimate of refractive error were considered fail/refer. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) for detection of refractive errors were determined using AAPOS referral criteria and receiver operating characteristic area under the curve (ROC AUC) analysis was conducted for measures of astigmatism. Agreement between screening and cycloplegic refraction measurements of astigmatism, spherical equivalent, and anisometropia were assessed using t tests and correlation analyses. A total...