Direct analysis of pharmaceuticals, their metabolites and transformation products in environmental waters using on-line TurboFlow™ chromatography–liquid chromatography–tandem mass spectrometry (original) (raw)
Related papers
2020
A simple on-line method was developed for the analysis of pharmaceuticals, pesticides and some metabolites in drinking, surface and wastewater samples. The technique is based on the use of on-line solid-phase extraction combined with liquid chromatography electrospray tandem mass spectrometry with positive electrospray ionization (LC-ESI(PI)-MS/MS). The injection of only 1 mL of filtered water sample is used with a total analysis time of 20 min, including the period required to flush the SPE cartridge with organic solvent and reconditioning the LC column. Method detection limits were in the range of 2 to 24 ng L À1 for the compounds of interest, with recoveries from 87 to 110% in surface as well as wastewater samples. Matrix effects were observed for some compounds without exceeding more than 25%. All results displayed a good degree of reproducibility, with relative standard deviations (RSD) of less than 12% for all compounds. Moreover, at least 200 samples were analyzed without alt...