[Measurement of tympanic temperature by infrared and thermocouple thermometry. A comparative study] (original) (raw)
Annales françaises d'anesthèsie et de rèanimation, 1995
Abstract
To compare tympanic temperatures obtained with an infrared thermometer Core Check 2090A (IVAC), non in contact with the tympanic membrane, to values obtained with a thermocouple thermometer Mon-A-Therm (Mallinckrodt) in contact with it, and to assess whether the differences depend on the person measuring the temperature. Prospective comparative open study. The study included 150 adults in whom the tympanic temperatures were measured at their admission to recovery room by 10 nurse anaesthetists. Each thermometer was inserted in the same patient into an external acoustic meatus. The median and range of the temperature differences were used to assess the bias and the accuracy of the measures. Their reliability was analyzed with consideration of the measuring person and the presence of irrelevant values, defined as a temperature difference over 0.3 degrees C. The temperatures obtained with both devices were similar (36.2 +/- 0.7 degrees C with IVAC thermometer vs 36.2 +/- 0.8 degrees C ...