Electrochemical Corrosion Behavior of Dental/Implant Alloys in Artificial Saliva (original) (raw)

2008, Journal of Materials Engineering and Performance

The corrosion behavior and passive film characteristics of various dental alloys such as Co-Cr, Ni-Cr, Cu-Ni-Al, and commercially pure Ti (c.p. Ti) were evaluated in artificial saliva medium by utilizing electrochemical impedance spectroscopy (EIS), Tafel polarization, and cyclic polarization studies. EIS studies were carried out for various durations viz. 1 h, 1 day, and 7 days to evaluate the stability of passive film and change in corrosion characteristics with respect to time. Electrochemical parameters such as Ecorr, icorr, corrosion rate, passive film characteristics with respect to time were obtained from various studies mentioned above. The corrosion resistance decreased in the order Cu-Ni-Al > cp Ti > Co-Cr (Commercial) > Ni-Cr > Co-Cr (DRDO developed) in artificial saliva solution.