A new Method for Myoelectric Signal Acquisition: Preparing the Patients to Efficiently Use an Artificial Arm (original) (raw)
This paper presents a new methodology in amputation surgery that is dedicated to patients who intend to use a myoelectric forearm prosthesis. The control signals for a myoelectric prosthesis are the surface EMG signals. In the case of an amputation stump, these signals are weak and difficult to use for effective control, due to the amputation methodology and the lack of activity after amputation. The proposed method aims the osteomyoplastic suturing of the muscles in circumferentially offset positions in order to obtain clear myoelectric signals, non-overlapping and with less correlated information. Following the surgery, a specific training is used, based on a biofeedback device, in order to help the patient to gradually obtain better control of stump’s muscles. The proposed methodology was evaluated experimentally at CMH, in a stump retouch surgery (on a patient who lost his hand almost 20 years ago), allowing comparison of the signals before surgery with ones after surgery and al...