Metallurgical Characterization of the Failed Motor Shaft Component from an Electric Surgical Motor used in Orthopedic Surgery (original) (raw)
Key Engineering Materials, 2015
Abstract
The paper presents the results of research conducted to determine the causes that led to the fracture of a motor shaft component from an electric surgical motor used in orthopedic surgery.The chemical and structural analysis emphasized a structure rich in complex silicate and sulfides inclusions. They form areas of weakness, acting like stress concentrators which induce fatigue shaft breakage. Their high density and non-uniform shape and distribution suggest a non-adequate technology of making and casting of steel, destinated to work in difficult conditions and having an important operational role.
Sohaciu Mirela hasn't uploaded this paper.
Let Sohaciu know you want this paper to be uploaded.
Ask for this paper to be uploaded.