Cholesterol biosensor based on N-(2-aminoethyl)-3-aminopropyl-trimethoxysilane self-assembled monolayer (original) (raw)
2007, Analytical Biochemistry
Cholesterol oxidase (ChOx) has been covalently immobilized onto two-dimensional self-assembled monolayer (SAM) of N-(2-aminoethyl)-3-aminopropyl-trimethoxysilane (AEAPTS) deposited on the indium-tin oxide (ITO) coated glass plates using N-ethyl-N 0 -(3-dimethylaminopropyl) carbodiimide and N-hydroxysuccinimide (EDC/NHS) chemistry. These ChOx/AEAPTS/ITO bioelectrodes are characterized using contact angle (CA) measurements, UV-visible spectroscopy, atomic force microscopy (AFM), electrochemical impedance technique, and Fourier transform infrared (FT-IR) technique. The covalently immobilized ChOx-modified AEAPTS bioelectrodes are used for the estimation of cholesterol in solution using UV-visible technique. These cholesterol sensing bioelectrodes show linearity as 50 to 500 mg/dl for cholesterol solution, detection limit as 25 mg/dl, sensitivity as 4.499 · 10 À5 Abs (mg/dl) À1 , K m value as 58.137 mg/dl (1.5 mM), apparent enzyme activity as 1.81 · 10 À3 U cm À2 , shelf life of approximately 10 weeks, and electrode reusability as 10 times.