What Is the Optimal Setting for a Continuous-Flow Left Ventricular Assist Device in Severe Mitral Regurgitation? (original) (raw)

Artificial organs, 2016

Abstract

Excessive left ventricular (LV) volume unloading can affect right ventricular (RV) function by causing a leftward shift of the interventricular septum in patients with mitral regurgitation (MR) receiving left ventricular assist device (LVAD) support. Optimal settings for the LVAD should be chosen to appropriately control the MR without causing RV dysfunction. In this study, we assessed the utility of our electrocardiogram-synchronized rotational speed (RS) modulation system along with a continuous-flow LVAD in a goat model of MR. We implanted EVAHEART devices after left thoracotomy in six adult goats weighing 66.4 ± 10.7 kg. Severe MR was induced through inflation of a temporary inferior vena cava filter placed within the mitral valve. We evaluated total flow (TF; the sum of aortic flow and pump flow [PF]), RV fractional area change (RVFAC) calculated by echocardiography, left atrial pressure (LAP), LV end-diastolic pressure (LVEDP), LV end-diastolic volume (LVEDV), and LV stroke wo...