In vitro thrombogenesis resulting from decreased shear rate and blood coagulability (original) (raw)
The International journal of artificial organs, 2016
Abstract
In vitro antithrombogenic testing with mock circulation is a useful type of pre-evaluation in ex vivo testing of mechanical assist devices. For effective in vitro testing, we have been developing a clear quantitative thrombogenesis model based on shear stress and blood coagulability. Bovine blood was used as the test medium. The activating clotting time (ACT) was adjusted with trisodium citrate and calcium chloride from 200 to 1,000 seconds. The blood was then applied to a rheometer and subjected to shear at 50 to 2,880 s-1. Blood coagulation time and degree of thrombogenesis were measured by the torque sensor of the rheometer. Prothrombin time (PT) and activated partial thromboplastin time (APTT) of the test blood were also measured after the application of shear. Blood coagulation time increased, and the degree of thrombogenesis decreased, with increases in shear rate to between 50 and 2,880 s-1. for test bloods with ACTs of 200 to 250 seconds. An ACT of 200 to 250 seconds is thus...