Marginal Fit of Implant-Supported All-Ceramic Zirconia Frameworks (original) (raw)
Journal of Oral Implantology, 2013
Abstract
This study evaluated the effect of fabrication techniques and cyclic loading on the vertical marginal fit of implant-supported fixed partial denture (FPD) frameworks. Thirty implant-supported 3-unit FPD frameworks were fabricated on a model system, divided into 3 equal groups (n = 10). The first group (control) was constructed from base metal alloy; the other 2 test groups were constructed from all-ceramic zirconia using a computer-aided design/computer-aided manufacturing (CAD/CAM) Cerec 3 system and a copy milling (Zirkonzahn) system. A cyclic load of 200 N was applied to each framework for up to 50,000 cycles. Linear measurements were made in micrometers of the vertical gap between the framework and the implant-supported abutment at 16 predetermined points before and after cyclic loading. The frameworks were viewed using scanning electron microscopy to inspect any fractographic features. One-way analysis of variance was performed to compare the marginal discrepancy values of the ...