A Method of Flexible Micro-wire Electrode Insertion in Rodent for Chronic Neural Recording and a Device for Electrode Insertion (original) (raw)
Related papers
Method of Thin Flexible Microelectrode Insertion in Deep Brain Region for Chronic Neural Recording
2019
Reliable chronic neural recording from focal deep brain structures is impeded by insertion injury and foreign body response, the magnitude of which is correlated with the mechanical mismatch between the electrode and tissue. Thin and flexible neural electrodes cause less glial scarring and record longer than stiff electrodes. However, the insertion of flexible microelectrodes in the brain has been a challenge. A novel insertion method is proposed, and demonstrated, for precise targeting deep brain structures using flexible micro-wire electrodes. A novel electrode guiding system is designed based on the principles governing the buckling strength of electrodes. The proposed guide significantly increases the critical buckling force of the microelectrode. The electrode insertion mechanism involves spinning of the electrode during insertion. The spinning electrode is slowly inserted in the brain through the electrode guide. The electrode guide does not penetrate into cortex. The electrod...
Ultrasoft microwire neural electrodes improve chronic tissue integration
Chronically implanted neural multi-electrode arrays (MEA) are an essential technology for recording electrical signals from neurons and/or modulating neural activity through stimulation. However, current MEAs, regardless of the type, elicit an inflammatory response that ultimately leads to device failure. Traditionally, rigid materials like tungsten and silicon have been employed to interface with the relatively soft neural tissue. The large stiffness mismatch is thought to exacerbate the inflammatory response. In order to minimize the disparity between the device and the brain, we fabricated novel ultrasoft electrodes consisting of elastomers and conducting polymers with mechanical properties much more similar to those of brain tissue than previous neural implants. In this study, these ultrasoft microelectrodes were inserted and released using a stainless steel shuttle with polyethyleneglycol (PEG) glue. The implanted microwires showed functionality in acute neural stimulation. When implanted for 1 or 8 weeks, the novel soft implants demonstrated significantly reduced inflammatory tissue response at week 8 compared to tungsten wires of similar dimension and surface chemistry. Furthermore, a higher degree of cell body distortion was found next to the tungsten implants compared to the polymer implants. Our results support the use of these novel ultrasoft electrodes for long term neural implants. Statement of Significance One critical challenge to the translation of neural recording/stimulation electrode technology to clinically viable devices for brain computer interface (BCI) or deep brain stimulation (DBS) applications is the chronic degradation of device performance due to the inflammatory tissue reaction. While many hypothesize that soft and flexible devices elicit reduced inflammatory tissue responses, there has yet to be a rigorous comparison between soft and stiff implants. We have developed an ultra-soft microelectrode with Young's modulus lower than 1 MPa, closely mimicking the brain tissue modulus. Here, we present a rigorous histological comparison of this novel ultrasoft electrode and conventional stiff electrode with the same size, shape and surface chemistry, implanted in rat brains for 1-week and 8-weeks. Significant improvement was observed for ultrasoft electrodes, including inflammatory tissue reaction, electrode-tissue integration as well as mechanical disturbance to nearby neurons. A full spectrum of new techniques were developed in this study, from insertion shuttle to in situ sectioning of the microelectrode to automated cell shape analysis, all of which should contribute new methods to the field. Finally, we showed the electrical functionality of the ultrasoft electrode, demonstrating the potential of flexible neu-ral implant devices for future research and clinical use.
Fluidic Microactuation of Flexible Electrodes for Neural Recording
Ultra-flexible microelectrodes that can bend and flex with the natural movement of the brain reduce the inflammatory response and improve the stability of long-term neural recordings. However, current methods to implant these highly flexible electrodes rely on temporary stiffening agents that increase the electrode size thus aggravating neural damage during implantation, which leads to cell loss and glial activation that persists even after the stiffening agents are removed or dissolve. A method to deliver thin, ultra-flexible electrodes deep into neural tissue without increasing the stiffness or size of the electrodes will enable minimally invasive electrical recordings from within the brain. Here we show that specially designed microfluidic devices can apply a tension force to ultra-flexible electrodes that prevents buckling without increasing the thickness or stiffness of the electrode during implantation. Additionally, these fluidic microdrives allow us to precisely actuate the ...
Flexible and stretchable micro-electrodes for in vitro and in vivo neural interfaces
Medical & Biological Engineering & Computing, 2010
Microelectrode arrays (MEAs) are designed to monitor and/or stimulate extracellularly neuronal activity. However, the biomechanical and structural mismatch between current MEAs and neural tissues remains a challenge for neural interfaces. This article describes a material strategy to prepare neural electrodes with improved mechanical compliance that relies on thin metal film electrodes embedded in polymeric substrates. The electrode impedance of micro-electrodes on polymer is comparable to that of MEA on glass substrates. Furthermore, MEAs on plastic can be flexed and rolled offering improved structural interface with brain and nerves in vivo.
2016
Micromotion between the brain and implanted electrodes is a major contributor to the failure of invasive microelectrodes. Movements of the electrode tip cause recording instabilities while spike amplitudes decline over the weeks/months post-implantation due to glial cell activation caused by sustained mechanical trauma. We compared the glial response over a 26-96 week period following implantation in the rabbit cortex of microwires and a novel flexible electrode. Horizontal sections were used to obtain a depth profile of the radial distribution of microglia, astrocytes and neurofilament. We found that the flexible electrode was associated with decreased gliosis compared to the microwires over these long indwelling periods. This was in part due to a decrease in overall microgliosis and enhanced neuronal density around the flexible probe, especially at longer periods of implantation.
Targeted delivery of neurochemicals and biomolecules for neuromodulation of brain activity is a powerful technique that, in addition to electrical recording and stimulation, enables a more thorough investigation of neural circuit dynamics. We have designed a novel flexible neural implant capable of controlled, localized chemical stimulation and high-density electrophysiology recording. To minimize tissue damage and response, the neural probe was implemented with a small cross-sectional dimension using planar micromachining processes on Parylene C, a mechanically flexible, biocompatible substrate. The probe shank features two large microelectrodes (chemical sites) for drug loading and sixteen small microelectrodes for electrophysiology recording to monitor neuronal response to drug release. To reduce the impedance while keeping the size of the microelectrodes small, poly(3,4-ethylenedioxythiophene) (PEDOT) was electrochemically coated on recording microelectrodes. In addition, PEDOT ...
Magnetic insertion system for flexible electrode implantation
Journal of neuroscience …, 2009
Chronic recording electrodes are a vital tool for brain research and neural prostheses. Despite decades of advances in recording technology, probe structures and implantation methods have changed little over time. Then as now, compressive insertion methods require probes to be constructed from hard, stiff materials, such as silicon, and contain a large diameter shank to penetrate the brain, particularly for deeper structures. The chronic presence of these probes results in an electrically isolating glial scar, degrading signal quality over time. This work demonstrates a new magnetic tension-based insertion mechanism that allows for the use of soft, flexible, and thinner probe materials, overcoming the materials limitations of modern electrodes. Probes are constructed from a sharp magnetic tip attached to a flexible tether. A pulsed magnetic field is generated in a coil surrounding a glass pipette containing the electrode. The applied field pulls the electrode tip forward, accelerating the probe into the neural tissue with a penetration depth that is calibrated against the charge voltage. Mathematical modeling and agar gel insertion testing demonstrate that the electrode can be implanted to a predictable depth given system specific parameters. Trial rodent implantations resulted in discernible single-unit activity on one of the probes. The current prototype demonstrates the feasibility of a tension based, magnetically driven implantation system and opens the door to a wide variety of new minimally invasive probe materials and configurations.
Design and Microfabrication Considerations for Reliable Flexible Intracortical Implants
arXiv (Cornell University), 2015
Current microelectrodes designed to record chronic neural activity suffer from recording instabilities due to the modulus mismatch between the electrode materials and the brain. We sought to address this by microfabricating a novel flexible neural probe. Our probe was fabricated from parylene-C with a WTi metal, using contact photolithography and reactive ion etching, with three design features to address this modulus mismatch: a sinusoidal shaft, a rounded tip and a polyimide anchoring ball. The anchor restricts movement of the electrode recording sites and the shaft accommodates the brain motion. We successfully patterned thick metal and parylene-C layers, with a reliable device release process leading to high functional yield. This novel reliably microfabricated probe can record stable neural activity for up to two years without delamination, surpassing the current state-of-the-art intracortical probes. This challenges recent concerns that have been raised over the long-term reliability of chronic implants when Parylene-C is used as an insulator, for both research and human applications The microfabrication and design considerations provided in this manuscript may aid in the future development of flexible devices for biomedical applications.
Implantable neurotechnologies: a review of micro- and nanoelectrodes for neural recording
Medical & biological engineering & computing, 2016
Electrodes serve as the first critical interface to the biological organ system. In neuroprosthetic applications, for example, electrodes interface to the tissue for either signal recording or tissue stimulation. In this review, we consider electrodes for recording neural activity. Recording electrodes serve as wiretaps into the neural tissues, providing readouts of electrical activity. These signals give us valuable insights into the organization and functioning of the nervous system. The recording interfaces have also shown promise in aiding treatment of motor and sensory disabilities caused by neurological disorders. Recent advances in fabrication technology have generated wide interest in creating tiny, high-density electrode interfaces for neural tissues. An ideal electrode should be small enough and be able to achieve reliable and conformal integration with the structures of the nervous system. As a result, the existing electrode designs are being shrunk and packed to form sma...