A novel technique for detection efficiency determination of HPGe (original) (raw)
2017, Radiation Physics and Chemistry
In this work, we present an experimental method to determine the detection efficiency of HPGe when the reference source according to the geometry of interest is not accessible. We use known activity point sources (PS) of 152 Eu, 137 Cs, 241 Am and 133 Ba to find the detection efficiency for disc source (DS) geometry. It can be assumed that a DS consists of several PS's. Mapping the detector surface by means of 137 Cs PS shows that there is radial symmetry for detection efficiency vs. energy. Each radial distance on the detector surface contains some points, which can be considered as a PS. By selecting two points in two different radii and central point, the DS efficiency is obtained. To ensure that the method is correct, we measure the activity of a known activity DS considering DS efficiency obtained by PS's. The DS comprises 137 Cs, 133 Ba and 60 Co. The relative difference between the measured and the reported activity of DS in most energies is less than 5%.
Related papers
A hybrid method to determine efficiency curve of HPGe detectors
Applied Radiation and Isotopes, 2009
A method was developed for determining the efficiency calibration curve of HPGe detectors between 26 and 1770 keV. A simple simulation based on the attenuation of gamma-rays in pure germanium was used together with efficiency calibrations performed for three detector-source distances: 5, 12 and 30 cm. The simple simulation model required an energy dependent term with three parameters and these parameters were determined from the fit of the simulated data to the measured efficiency of the detector. The efficiency response of the detector determined this way was then tested with other certified sources to verify the validity of the method.
Analytical approach for radioactivity correlation of disc sources with HPGe detector efficiency
2014
Article in Applied radiation and isotopes: including data, instrumentation and methods for use in agriculture, industry and medicine · We examine changes in the detector efficiency depended on source diameter and source detector separation. The concept of a double point detector model approach (DPDM) was used. Used the efficiency transfer factor for cases of sources with different geometry, transfer from a point source to a disc source. The dependence of the count rates on the efficiency estimate.
Study of Some Parameters Affecting Efficiency of HpGe Detectors for Accurate Radionuclides Analysis
The International Conference on Chemical and Environmental Engineering
In fact, optimized and adjusted parameters of analysis techniques are critical issues and lead strongly to accuracy and high confidence of results. This study is mainly focus on investigation (validation and verification; V&V) of some parameters affecting the efficiency and efficacy of HpGe gamma spectrometers; voltage variation, shaping time, type of detectors (n-type or p-type), geometry (size) of detector, source to detector distance and density of the sample matrix. Three different HpGe detectors with their electronics were used to carry out this work. The standard sources; 137 Cs, 60 Co, 57 Co, 133 Ba were used. The source-to-detector distance of 25cm was optimized to avoid the coincidence summing and dead-time correction .The efficiency curves of the spectrometers have been measured and presented at the above-mentioned parameters. A simulated reference efficiency curve at different parameters using standard point sources was also produced by Angel-3 software. The noises of electronics due to the electromechanical cooling system were also investigated. The optimum parameters were selected for precise analysis of some samples (Black sand and El-Dabaa site) at 3300V and 4000V for p-type and n-type detector respectively. It was observed that n-type detector has better efficiency than p-type especially at low energy below 120 KeV. It covers energies from 10 keV to 3 MeV. The analyses found also that the density has a strong effect on efficiency curve below energy 250 keV. The results obtained at different parameters are discussed, interpreted and presented as figures
Determination of full energy peak efficiency of HpGe detector from 59.5 to 1332.5 keV
Indian Journal of Pure & Applied Physics, 2006
To detect very low radio activities in the environmental samples, it is sometimes necessary to know the full energy peak efficiency of the detector. This work proposes a full energy peak efficiency simulation which represents adequately the high purity germanium coaxial detector responses in the energy range 59.5-1332.5 keV and at distances of 5, 10 and 25 cm.
Calibration of closed-end HPGe detectors using bar (Parallelepiped) sources
2008
The National Institute of Standards and Technology (NIST, Gaithersburg, MD 20878, USA) uses different extended sources, such as bar (parallelepiped) sources of high activity to calibrate the hyper pure germanium (HPGe) cylindrical detectors. The absolute full-energy peak efficiency and the source self-absorption have been investigated in this work by the use of compact analytical expressions for closedend HPGe detectors and bar sources. The validity of these expressions is tested through extensive comparisons with the experimental values carried out in the NIST. The comparisons indicate that the direct mathematical method used in the present work is useful in the efficiency calibration of HPGe detectors with bar sources.
Optimization of HPGe detector response using fast and reliable method
Nuclear Technology and Radiation Protection, 2020
Safeguards system with high effectiveness and efficiency must comprise a set of measurements with capabilities satisfactory for the verification of nuclear materials. In this paper, we present key parameter measurements of detector modeling in a commercial n-type low energy germanium detector of a planar crystal with a relative efficiency of nearly 15 %. The detector optimization will hold a significant function in measuring nuclear materials for safeguards application. Standard nuclear materials with diverse enrichment (depleted and low enriched) of uranium and point-like sources (137Cs, 60Co) and mixed radioactive source for Eu isotopes (152Eu, 154Eu, and 155Eu) were benefited to explore the energy resolution and detector efficiency. The energy resolution is measured over a wide range of rise time and flattop. In addition to the experimental work, the Monte Carlo simulation code is used for modeling the setup configuration to obtain the absolute efficiency at different energies. A...
Science and Technology Development Journal, 2015
Construction of detector is necessary. However, on large energy range the manufacturers could not also support the explicit function of relative and absolute efficiencies of detectors. One of the reasons is a restriction of energy range of gamma sources (normally < 3 MeV). This paper presents the results of construction of relative and absolute efficiency functions within a range from 122 keV to 8.5 MeV. The sources are used combining 152Eu point source and 36Cl activated isotope by thermal neutron captured reaction 35Cl of Dalat nuclear reactor (DNR) by 35Cl(n, γ)36Cl reaction. This result can be applied in determining quantitative analysis of samples of neutron activation and radioactivity chemistry
Evaluation of HPGe detector efficiency for point sources using virtual point detector model
Applied Radiation and Isotopes, 2011
The concept of a virtual point detector (VPD) has been developed and validated in the past for Ge(Li) and HPGe detectors. In the present research, a new semi-empirical equation involving photon energy and source-virtual point detector distance for the efficiency of point sources by HPGe detectors is introduced , which is based on the VPD model. The calculated efficiencies for both coaxial and off-axis geometries by this equation are in good agreement with experimental data. The estimated uncertainties are less than 4%.