Abstract
In this paper, we introduce a technique for double-sealed ceramic packages for the long-term protection of implanted electronics against body fluids. A sequential sealing procedure consisting of a first step, during which the package is sealed with epoxy, protecting the implant electronics from aggressive flux fumes. These result from the application of the actual moisture barrier which is a metal seal applied in a second step by soft soldering. Epoxy sealing is carried out in helium atmosphere for later fine leak testing. The solder seal is applied on the laboratory bench. After the first sealing step, a satisfactory barrier for moisture is already achieved with values for helium leakage of usually LHe = 6·10(-8) mbar 1 s(-1). After solder sealing, a very low leakage rate of LHe ≤ 1·10(-12) mbar 1 s(-1) was found, which was the lower detection limit of the measurement setup, suggesting excellent hermeticity and hence moisture barrier. Presuming an implant package volume of V ≥ 0.5 ...
Martin Schuettler hasn't uploaded this paper.
Let Martin know you want this paper to be uploaded.
Ask for this paper to be uploaded.