Energy Calibration of a Silicon-Strip Detector for Photon-Counting Spectral CT by Direct Usage of the X-ray Tube Spectrum (original) (raw)

Spectral response model for a multibin photon-counting spectral computed tomography detector and its applications

Journal of Medical Imaging, 2015

Variations among detector channels in computed tomography can lead to ring artifacts in the reconstructed images and biased estimates in projection-based material decomposition. Typically, the ring artifacts are corrected by compensation methods based on flat fielding, where transmission measurements are required for a number of material-thickness combinations. Phantoms used in these methods can be rather complex and require an extensive number of transmission measurements. Moreover, material decomposition needs knowledge of the individual response of each detector channel to account for the detector inhomogeneities. For this purpose, we have developed a spectral response model that binwise predicts the response of a multibin photoncounting detector individually for each detector channel. The spectral response model is performed in two steps. The first step employs a forward model to predict the expected numbers of photon counts, taking into account parameters such as the incident x-ray spectrum, absorption efficiency, and energy response of the detector. The second step utilizes a limited number of transmission measurements with a set of flat slabs of two absorber materials to fine-tune the model predictions, resulting in a good correspondence with the physical measurements. To verify the response model, we apply the model in two cases. First, the model is used in combination with a compensation method which requires an extensive number of transmission measurements to determine the necessary parameters. Our spectral response model successfully replaces these measurements by simulations, saving a significant amount of measurement time. Second, the spectral response model is used as the basis of the maximum likelihood approach for projection-based material decomposition. The reconstructed basis images show a good separation between the calcium-like material and the contrast agents, iodine and gadolinium. The contrast agent concentrations are reconstructed with more than 94% accuracy.

Modeling the performance of a photon counting x-ray detector for CT: Energy response and pulse pileup effects

Medical Physics, 2011

Recently, photon counting x-ray detectors ͑PCXDs͒ with energy discrimination capabilities have been developed for potential use in clinical computed tomography ͑CT͒ scanners. These PCXDs have great potential to improve the quality of CT images due to the absence of electronic noise and weights applied to the counts and the additional spectral information. With high count rates encountered in clinical CT, however, coincident photons are recorded as one event with a higher or lower energy due to the finite speed of the PCXD. This phenomenon is called a "pulse pileup event" and results in both a loss of counts ͑called "deadtime losses"͒ and distortion of the recorded energy spectrum. Even though the performance of PCXDs is being improved, it is essential to develop algorithmic methods based on accurate models of the properties of detectors to compensate for these effects. To date, only one PCXD ͑model DXMCT-1, DxRay, Inc., Northridge, CA͒ has been used for clinical CT studies. The aim of that study was to evaluate the agreement between data measured by DXMCT-1 and those predicted by analytical models for the energy response, the deadtime losses, and the distorted recorded spectrum caused by pulse pileup effects. Methods: An energy calibration was performed using 99m Tc ͑140 keV͒, 57 Co ͑122 keV͒, and an x-ray beam obtained with four x-ray tube voltages ͑35, 50, 65, and 80 kVp͒. The DXMCT-1 was placed 150 mm from the x-ray focal spot; the count rates and the spectra were recorded at various tube current values from 10 to 500 A for a tube voltage of 80 kVp. Using these measurements, for each pulse height comparator we estimated three parameters describing the photon energy-pulse height curve, the detector deadtime , a coefficient k that relates the x-ray tube current I to an incident count rate a by a = k ϫ I, and the incident spectrum. The mean pulse shape of all comparators was acquired in a separate study and was used in the model to estimate the distorted recorded spectrum. The agreement between data measured by the DXMCT-1 and those predicted by the models was quantified by the coefficient of variation ͑COV͒, i.e., the root mean square difference divided by the mean of the measurement. Results: Photon energy versus pulse height curves calculated with an analytical model and those measured using the DXMCT-1 were in agreement within 0.2% in terms of the COV. The COV between the output count rates measured and those predicted by analytical models was 2.5% for deadtime losses of up to 60%. The COVs between spectra measured and those predicted by the detector model were within 3.7%-7.2% with deadtime losses of 19%-46%.

Spectral and spatial resolution of semiconductor detectors in medical X- and gamma ray imaging

2008 IEEE Nuclear Science Symposium Conference Record, 2008

In X-and gamma ray based medical systems, detector performance is a key driver for diagnostic quality. Over the last years and decades, indirect conversion scintillator detectors have become the standard for many medical applications including X-ray Radiography, Computed Tomography and SPECT. Recently, direct conversion semiconductor detectors based on CdTe and CdZnTe (CZT) have come into focus, as they might offer improved or additional performance for specific applications. In this paper we use generic physical models to compare the spatial and spectral resolution of both detector types. The spatial resolution is quantified by the Modulation Transfer Function (MTF). We find that the direct conversion of quanta leads to an approximately sinc-like MTF. In comparison, the MTF of a pixelized scintillator detector shows a mid-frequency drop due to optical signal cross-talk.

Preliminary study of the advantages of X-ray energy selection in CT imaging

2007

It is well known that a monochromatic X-ray source with an energy optimized for the organ thickness to be imaged could result in a better image quality in transmission radiology. In this paper we present the preliminary investigation for the implementation of this technique in computer tomography (CT) imaging. The detection system is based on a 1 mm thick silicon pixel detector bump bonded to a VLSI read-out, Medipix2. This detector ensures a good detection efficiency (46%) in the used energy range (60 kVp) with a good spatial resolution that arises from a 55 mm square pixel. The Medipix2 read-out electronics is not only a single photon counting system, but has also the capability of dual-energy threshold, that allows us to detect only photons that are in a chosen energy window. In this paper we present the results obtained in CT imaging of small samples, by selecting various energy windows within a standard X-ray tube spectrum so as to maximize the differentiation between significant attenuation coefficients. This study is preliminary for a future development of a dual-energy CT that could add functional information to the morphological information that is obtained in a CT examination. r

Energy Calibration of the Pixels of Spectral X-ray Detectors

IEEE Transactions on Medical Imaging, 2014

The user has requested enhancement of the downloaded file. All in-text references underlined in blue are added to the original document and are linked to publications on ResearchGate, letting you access and read them immediately.

Preliminary evaluation of a silicon strip detector for photon-counting spectral CT

Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2012

An edge-on silicon strip detector designed for photon-counting spectral computed tomography (CT) is presented. Progress on the development of an application specific integrated circuit (ASIC) to process the pulses and sort them into energy bins is reported upon. The ASIC and detector are evaluated in terms of electronic noise, energy resolution, count rate linearity under high-frequency periodic pulses, threshold variation and gain. The high-frequency periodic pulses are injected both by means of an external pulse generator and a pulsed laser illuminating the silicon diode. The pulsed laser system has $ 100 ps pulse width and thus generates near instantaneous pulses in the diode, thus mimicking real X-ray conversions.

Material separation in x-ray CT with energy resolved photon-counting detectors

Medical Physics, 2011

Purpose: The objective of the study was to demonstrate that, in x-ray computed tomography ͑CT͒, more than two types of materials can be effectively separated with the use of an energy resolved photon-counting detector and classification methodology. Specifically, this applies to the case when contrast agents that contain K-absorption edges in the energy range of interest are present in the object. This separation is enabled via the use of recently developed energy resolved photoncounting detectors with multiple thresholds, which allow simultaneous measurements of the x-ray attenuation at multiple energies. Methods: To demonstrate this capability, we performed simulations and physical experiments using a six-threshold energy resolved photon-counting detector. We imaged mouse-sized cylindrical phantoms filled with several soft-tissue-like and bone-like materials and with iodine-based and gadolinium-based contrast agents. The linear attenuation coefficients were reconstructed for each material in each energy window and were visualized as scatter plots between pairs of energy windows. For comparison, a dual-kVp CT was also simulated using the same phantom materials. In this case, the linear attenuation coefficients at the lower kVp were plotted against those at the higher kVp. Results: In both the simulations and the physical experiments, the contrast agents were easily separable from other soft-tissue-like and bone-like materials, thanks to the availability of the attenuation coefficient measurements at more than two energies provided by the energy resolved photon-counting detector. In the simulations, the amount of separation was observed to be proportional to the concentration of the contrast agents; however, this was not observed in the physical experiments due to limitations of the real detector system. We used the angle between pairs of attenuation coefficient vectors in either the 5-D space ͑for non-contrast-agent materials using energy resolved photon-counting acquisition͒ or a 2-D space ͑for contrast agents using energy resolved photon-counting acquisition and all materials using dual-kVp acquisition͒ as a measure of the degree of separation. Compared to dual-kVp techniques, an energy resolved detector provided a larger separation and the ability to separate different target materials using measurements acquired in different energy window pairs with a single x-ray exposure. Conclusions: We concluded that x-ray CT with an energy resolved photon-counting detector with more than two energy windows allows the separation of more than two types of materials, e.g., soft-tissue-like, bone-like, and one or more materials with K-edges in the energy range of interest. Separating material types using energy resolved photon-counting detectors has a number of advantages over dual-kVp CT in terms of the degree of separation and the number of materials that can be separated simultaneously.

A simple method for CT-scanner calibration against effective photon energy

Instruments and Experimental Techniques, 2000

An original method for CT-scanner calibration against effective photon energy is presented. It is based on optimization of the CT-scanner output characteristic (the dependence of the measured linear attenuation coefficient on the calculated value). A significant gain in accuracy is achieved with respect to the calibration techniques which are conventionally used in commercial computed tomography. The method can be recommended for use in dual-energy computed tomography. As an example of its application, measurements of the effective atomic number and electron density of organic plastics are described. It is shown that the method provides acceptable results even for an increased noise level in CT images.

Characterization of a novel photon counting detector for clinical CT: count rate, energy resolution, and noise performance

Medical Imaging 2009: Physics of Medical Imaging, 2009

We report on a characterization study of a multi-row direct-conversion x-ray detector used to generate the first photon counting clinical x-ray computed tomography (CT) patent images. In order to provide the photon counting detector with adequate performance for low-dose CT applications, we have designed and fabricated a fast application specific integrated circuit (ASIC) for data readout from the pixellated CdTe detectors that comprise the photon counting detector. The cadmium telluride (CdTe) detector has 512 pixels with a 1 mm pitch and is vertically integrated with the ASIC readout so it can be tiled in two dimensions similar to those that are tiled in an arc found in 32-row multi-slice CT systems. We have measured several important detector parameters including the maximum output count rate, energy resolution, and noise performance. Additionally the relationship between the output and input rate has been found to fit a non-paralyzable detector model with a dead time of 160 nsec. A maximum output rate of 6 × 10 6 counts per second per pixel has been obtained with a low output x-ray tube for CT operated between 0.01 mA and 6 mA at 140 keV and different source-to-detector distances. All detector noise counts are less that 20 keV which is sufficiently low for clinical CT. The energy resolution measured with the 60 keV photons from a 241 Am source is ~12%. In conclusion, our results demonstrate the potential for the application of the CdTe based photon counting detector to clinical CT systems. Our future plans include further performance improvement by incorporating drift structures to each detector pixel.

Characterization of a novel photon counting detector for clinical CT: count rate, energy resolution, and noise performance

2009

We report on a characterization study of a multi-row direct-conversion x-ray detector used to generate the first photon counting clinical x-ray computed tomography (CT) patent images. In order to provide the photon counting detector with adequate performance for low-dose CT applications, we have designed and fabricated a fast application specific integrated circuit (ASIC) for data readout from the pixellated CdTe detectors that comprise the photon counting detector. The cadmium telluride (CdTe) detector has 512 pixels with a 1 mm pitch and is vertically integrated with the ASIC readout so it can be tiled in two dimensions similar to those that are tiled in an arc found in 32-row multi-slice CT systems. We have measured several important detector parameters including the maximum output count rate, energy resolution, and noise performance. Additionally the relationship between the output and input rate has been found to fit a non-paralyzable detector model with a dead time of 160 nsec. A maximum output rate of 6 × 106 counts per second per pixel has been obtained with a low output x-ray tube for CT operated between 0.01 mA and 6 mA at 140 keV and different source-to-detector distances. All detector noise counts are less that 20 keV which is sufficiently low for clinical CT. The energy resolution measured with the 60 keV photons from a 241Am source is ~12%. In conclusion, our results demonstrate the potential for the application of the CdTe based photon counting detector to clinical CT systems. Our future plans include further performance improvement by incorporating drift structures to each detector pixel.