Ferromagnetism and magnetic resonance artifacts of platinum embolization microcoils - PubMed (original) (raw)
Case Reports
. 1991 May-Jun;14(3):163-6.
doi: 10.1007/BF02577720.
Affiliations
- PMID: 1878906
- DOI: 10.1007/BF02577720
Case Reports
Ferromagnetism and magnetic resonance artifacts of platinum embolization microcoils
M W Marshall et al. Cardiovasc Intervent Radiol. 1991 May-Jun.
Abstract
We performed in vitro tests of ferromagnetism of platinum microcoils of two manufacturers. In addition, we performed in vitro magnetic resonance imaging (MRI) on these two devices to determine their levels of magnetic susceptibility artifact production. We found that both devices were nonferromagnetic and produced a very low level of artifact during MRI. In vivo MRI of a dog and 2 patients having previous placement of platinum microcoils was performed without coil migration or significant artifact production. Platinum microcoils are highly MR-compatible.
References
- Radiology. 1983 Apr;147(1):139-48 -PubMed
- Radiology. 1988 Sep;168(3):713-9 -PubMed
- Cardiovasc Intervent Radiol. 1989 May-Jun;12(3):125-7 -PubMed
- Radiology. 1986 Jan;158(1):57-61 -PubMed
- AJR Am J Roentgenol. 1988 May;150(5):1174-6 -PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources