wavemeters (original) (raw)

Author: the photonics expert

Definition: devices for precise measurements of laser wavelengths

Alternative term: wavelength meters

More general term: optical metrology instruments

Categories: article belongs to category photonic devices photonic devices, article belongs to category light detection and characterization light detection and characterization, article belongs to category optical metrology optical metrology

DOI: 10.61835/8qw Cite the article: BibTex plain textHTML Link to this page share on LinkedIn

A wavemeter (or wavelength meter) is a kind of interferometer which is used for precise wavelength measurements for light which is delivered in the form of a laser beam. There are different variants, including scanning wavemeters and static devices with no moving parts.

Scanning Michelson Interferometer

scanning Michelson wavemeter

Figure 1: Scanning Michelson wavemeter.

An often used type of wavemeter is based on a Michelson interferometer. Figure 1 shows the principle setup. Light from the optical source under test is sent into the Michelson interferometer, while the length of one of interferometer arms is scanned through a certain range. The period of the output power recorded by the photodetector with respect to the arm length changes then reflects the wavelength. The control of the measurement process and the data analysis are usually accomplished with a microprocessor device.

The principle of such a wavemeter can be extended for the measurement of the optical spectrum of non-monochromatic (= polychromatic) sources (→ spectrometers). Essentially, the spectrum is obtained by applying a Fourier transform to the detected power versus arm length difference. This method is called Fourier transform spectroscopy.

Various types of errors can affect the measurement accuracy:

Depending on the quality of the device, the achieved wavelength accuracy can be 0.01 nm, for example. Due to the scanning, such devices are not suitable for pulsed radiation.

Static Fizeau Interferometer

static Fizeau wavemeter

Figure 2: Static wavemeter based on a Fizeau interferometer.

A Fizeau interferometer (Figure 2) uses two plane reflecting surfaces with a slight deviation from exact parallelity. For example, it can contain a glass wedge with a small angular mismatch of a few arcseconds, for example, where the front surface is partially reflecting and the back surface is fully reflecting. There are also implementations with discrete mirrors.

Effectively, two copies of the input beam are superimposed with a slight relative angle, leading to an interference pattern, the period of which depends on the wavelength. Usually, the input beam first gets through some spatial filter, and a collimated beam with large diameter is sent onto the Fizeau interferometer.

The shape of the interference pattern is measured with a CCD array, for example, and the data are processed using a microprocessor.

This technique does not require moving parts and can achieve quite high wavelength resolution. It may also be applied to pulsed lasers.

Vacuum Wavelength, or Wavelength in Air?

There is a significant difference between wavelengths in air or in vacuum, although the refractive index of air is quite close to 1. That raises the question what type of wavelength a wavemeter specifies.

The interferometric measurements are usually done in air, but one cannot conclude from that that wavelength values in air will result. For example, a wavemeter may measure wavelengths relative to the wavelength of some reference source, and if that operates in a wavelength region with similar refractive index, one may get the vacuum wavelength of the probed light by using the vacuum wavelength for the reference source in the calculation. It is also possibly to apply a sophisticated compensation calculation, which may even take into account the measured air pressure and humidity (as these both influence the refractive index).

For highly accurate wavemeters, it makes more sense to deliver vacuum wavelengths, avoiding the dependence on detailed ambient conditions.

Aspects for Selecting a Wavemeter

Wavemeters of different types can differ in many respects:

Alternative Measurement Techniques

Other types of wavemeters can be based on Fabry–Pérot interferometers, for example.

The accuracy of wavelength measurements is affected by a number of factors such as wavefront distortions of the beam. A much higher precision can be achieved with optical frequency measurements (→ frequency metrology), which are immune to such effects.

Using a wavemeter is usually more precise than measuring a wavelength with a spectrometer. The advantage of a spectrometer, however, is that it also delivers information on relative powers of different spectral components. There are also wavemeters which can act as spectrometers, thus offering both functions with high accuracy.

More to Learn

Encyclopedia articles:

Suppliers

Bibliography

[1] M. B. Morris et al., “Fizeau wavemeter for pulsed laser wavelength measurement”, Appl. Opt. 23 (21), 3862 (1984); https://doi.org/10.1364/AO.23.003862
[2] R. Castell et al., “The accuracy of laser wavelength meters”, Appl. Phys. B 38 (1), 1 (1985); https://doi.org/10.1007/BF00691763
[3] D. F. Gray et al., “Simple compact Fizeau wavemeter”, Appl. Opt. 25 (8), 1339 (1986); https://doi.org/10.1364/AO.25.001339
[4] C. Reiser and R. B. Lopert, “Laser wavemeter with solid Fizeau wedge interferometer”, Appl. Opt. 27 (17), 3656 (1988); https://doi.org/10.1364/AO.27.003656
[5] P. J. Fox et al., “A reliable, compact, and low-cost Michelson wavemeter for laser wavelength measurement”, Am. J. Phys. 67 (7), 624 (1999); https://doi.org/10.1119/1.19335
[6] J. J. Snyder and T. W. Hänsch, “Laser Wavemeters”, in Topics in Applied Physics: Dye Lasers (ed. F. P. Schäfer), Springer, Berlin (1973)

(Suggest additional literature!)

Questions and Comments from Users

Here you can submit questions and comments. As far as they get accepted by the author, they will appear above this paragraph together with the author’s answer. The author will decide on acceptance based on certain criteria. Essentially, the issue must be of sufficiently broad interest.

Please do not enter personal data here. (See also our privacy declaration.) If you wish to receive personal feedback or consultancy from the author, please contact him, e.g. via e-mail.

By submitting the information, you give your consent to the potential publication of your inputs on our website according to our rules. (If you later retract your consent, we will delete those inputs.) As your inputs are first reviewed by the author, they may be published with some delay.