Measurement microphone calibration (original) (raw)

About DBpedia

In order to take a scientific measurement with a microphone, its precise sensitivity must be known (in volts per pascal). Since this may change over the lifetime of the device, it is necessary to regularly calibrate measurement microphones. This service is offered by some microphone manufacturers and by independent testing laboratories. Microphone calibration by certified laboratories should ultimately be traceable to primary standards a (National) Measurement Institute that is a signatory to International Laboratory Accreditation Cooperation. These could include the National Physical Laboratory in the UK, PTB in Germany, NIST in the USA and the National Measurement Institute, Australia, where the reciprocity calibration (see below) is the internationally recognised means of realising th

Property Value
dbo:abstract In order to take a scientific measurement with a microphone, its precise sensitivity must be known (in volts per pascal). Since this may change over the lifetime of the device, it is necessary to regularly calibrate measurement microphones. This service is offered by some microphone manufacturers and by independent testing laboratories. Microphone calibration by certified laboratories should ultimately be traceable to primary standards a (National) Measurement Institute that is a signatory to International Laboratory Accreditation Cooperation. These could include the National Physical Laboratory in the UK, PTB in Germany, NIST in the USA and the National Measurement Institute, Australia, where the reciprocity calibration (see below) is the internationally recognised means of realising the primary standard. Laboratory standard microphones calibrated using this method are used in-turn to calibrate other microphones using comparison calibration techniques (‘secondary calibration’), referencing the output of the ‘test’ microphone against that of the reference laboratory standard microphone. A microphone’s sensitivity varies with frequency (as well as with other factors such as environmental conditions) and is therefore normally recorded as several sensitivity values, each for a specific frequency band (see frequency spectrum). A microphone’s sensitivity can also depend on the nature of the sound field it is exposed to. For this reason, microphones are often calibrated in more than one sound field, for example a pressure field and a free field. Depending on their application, measurement microphones must be tested periodically (every year or several months, typically), and after any potentially damaging event, such as being dropped or exposed to sound levels beyond the device’s operational range. (en)
dbo:wikiPageID 28022494 (xsd:integer)
dbo:wikiPageLength 9881 (xsd:nonNegativeInteger)
dbo:wikiPageRevisionID 1057421368 (xsd:integer)
dbo:wikiPageWikiLink dbr:Primary_standard dbr:Volts dbr:International_Laboratory_Accreditation_Cooperation dbr:Mass dbr:Ohm dbr:Frequency dbr:NIST dbr:Critical_distance dbr:Anechoic_chamber dbr:Calibration dbc:Sound_technology dbr:Brüel_&_Kjær dbr:Adiabatic dbr:Time dbr:Dissemination dbc:Acoustics dbr:Pascal_(unit) dbr:Diaphragm_(acoustics) dbr:Length dbr:Frequency_spectrum dbr:Acoustic_impedance dbr:Physikalisch-Technische_Bundesanstalt dbr:Piston dbr:Microphone dbr:National_Measurement_Institute,_Australia dbr:National_Physical_Laboratory_(United_Kingdom) dbr:Sound_pressure_level dbr:Sensitivity_(electronics) dbr:Volt dbr:Condenser_microphones dbr:Specific_heat dbr:Uncertainty_of_measurement dbr:Adiabiatic dbr:A_weighting
dbp:wikiPageUsesTemplate dbt:No_footnotes dbt:Refbegin dbt:Refend dbt:Visible_anchor
dct:subject dbc:Sound_technology dbc:Acoustics
rdfs:comment In order to take a scientific measurement with a microphone, its precise sensitivity must be known (in volts per pascal). Since this may change over the lifetime of the device, it is necessary to regularly calibrate measurement microphones. This service is offered by some microphone manufacturers and by independent testing laboratories. Microphone calibration by certified laboratories should ultimately be traceable to primary standards a (National) Measurement Institute that is a signatory to International Laboratory Accreditation Cooperation. These could include the National Physical Laboratory in the UK, PTB in Germany, NIST in the USA and the National Measurement Institute, Australia, where the reciprocity calibration (see below) is the internationally recognised means of realising th (en)
rdfs:label Measurement microphone calibration (en)
owl:sameAs freebase:Measurement microphone calibration wikidata:Measurement microphone calibration dbpedia-fi:Measurement microphone calibration https://global.dbpedia.org/id/3cU59
prov:wasDerivedFrom wikipedia-en:Measurement_microphone_calibration?oldid=1057421368&ns=0
foaf:isPrimaryTopicOf wikipedia-en:Measurement_microphone_calibration
is dbo:wikiPageRedirects of dbr:Pistonphone dbr:Measurement_Microphone_Calibration
is dbo:wikiPageWikiLink of dbr:Pistonphone dbr:Reverberation_room dbr:Microphone dbr:Plane_wave_tube dbr:Measurement_Microphone_Calibration dbr:Noise_dosimeter
is foaf:primaryTopic of wikipedia-en:Measurement_microphone_calibration