Ultrasound—biophysics mechanisms (original) (raw)

Prog Biophys Mol Biol. Author manuscript; available in PMC 2007 Sep 28.

Published in final edited form as:

PMCID: PMC1995002

NIHMSID: NIHMS17904

Bioacoustics Research Laboratory, Department of Electrical and Computer Engineering, University of Illinois, 405 N. Mathews, Urbana, IL 61801, USA

*Tel.: +1 217 333 2407; fax: +1 217 244 0105. E-mail address: ude.cuiu@odw.

Abstract

Ultrasonic biophysics is the study of mechanisms responsible for how ultrasound and biological materials interact. Ultrasound-induced bioeffect or risk studies focus on issues related to the effects of ultrasound on biological materials. On the other hand, when biological materials affect the ultrasonic wave, this can be viewed as the basis for diagnostic ultrasound. Thus, an understanding of the interaction of ultrasound with tissue provides the scientific basis for image production and risk assessment. Relative to the bioeffect or risk studies, that is, the biophysical mechanisms by which ultrasound affects biological materials, ultrasound-induced bioeffects are generally separated into thermal and nonthermal mechanisms. Ultrasonic dosimetry is concerned with the quantitative determination of ultrasonic energy interaction with biological materials.

Whenever ultrasonic energy is propagated into an attenuating material such as tissue, the amplitude of the wave decreases with distance. This attenuation is due to either absorption or scattering. Absorption is a mechanism that represents that portion of ultrasonic wave that is converted into heat, and scattering can be thought of as that portion of the wave, which changes direction. Because the medium can absorb energy to produce heat, a temperature rise may occur as long as the rate of heat production is greater than the rate of heat removal. Current interest with thermally mediated ultrasound-induced bioeffects has focused on the thermal isoeffect concept. The non-thermal mechanism that has received the most attention is acoustically generated cavitation wherein ultrasonic energy by cavitation bubbles is concentrated. Acoustic cavitation, in a broad sense, refers to ultrasonically induced bubble activity occurring in a biological material that contains pre-existing gaseous inclusions. Cavitation-related mechanisms include radiation force, microstreaming, shock waves, free radicals, microjets and strain. It is more challenging to deduce the causes of mechanical effects in tissues that do not contain gas bodies. These ultrasonic biophysics mechanisms will be discussed in the context of diagnostic ultrasound exposure risk concerns.

Keywords: Ultrasonic biophysics, Ultrasonic bioeffects, Ultrasonic dosimetry, Thermal mechanism, Non-thermal mechanism

1. Ultrasonic biophysics

Ultrasonic biophysics is the study of mechanisms responsible for how ultrasound and biological materials interact. As shown in Fig. 1, the study of how ultrasound affects biological materials can be viewed as bioeffect studies that can lead to therapeutic applications and risk assessments for diagnostic ultrasound applications. On the other hand, the study of how tissue affects the ultrasound wave can be viewed as the basis for diagnostic ultrasound imaging. Thus, an understanding of the interaction of ultrasound with tissue provides the scientific basis for image production, therapeutic applications and risk assessment.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0001.jpg

Schematic diagram of ultrasonic biophysics.

Ultrasonic dosimetry (O'Brien, 1978, 1986, 1992b) is concerned with the quantitative determination of the interaction of ultrasonic energy with biological materials, that is, defining the quantitative relationship between some physical agent and the biological effect it produces. To understand more fully ultrasonic dosimetry and ultrasonic interaction mechanisms, it is appropriate to first introduce basic ultrasonic quantities, and then develop common nomenclature. Then, general dosimetric concepts will be presented because a large body of literature and history exists to quantitate the interaction of various propagated energies and biological materials. Ultrasonic dosimetry and its current status will be presented. To conclude, interaction mechanisms, both thermal and non-thermal, are discussed.

2. Basic ultrasonic quantities

Sound is the rapid motion of molecules. These molecular vibrations transport energy from a transmitter, a sound source like our voice, to a receiver like our ear. Sound travels in waves that transport energy from one location to another. When the molecules get closer together, this is called compression, and when they separate, this is called rarefaction. This mechanical motion, the rapid back and forth motion, is the basis for calling sound a mechanical wave or a mechanically propagated wave.

2.1. Acoustic spectrum

We have many perceptions of the nature of sound. The idea of pitch refers to our perception of frequency, that is, the number of times a second that air vibrates in producing sound that we hear. Voices are classified according to pitch in which the lowest frequency is a bass voice and the highest frequency is a soprano voice. This description of frequency, however, is limited to the frequency range, or spectrum, over which human beings can hear sounds. There are sound frequencies below and above what human beings can hear. The acoustic spectrum is shown in Fig. 2a. The lowest frequency classification in the acoustic spectrum is infrasound that has a frequency range less than about 20 Hz. Audible sound is what human beings hear and has an approximate frequency range between 20 Hz and 20 kHz. The ultrasound frequency range starts at a frequency of about 20 kHz. Examples of devices that emit frequencies at the lower frequency end of the ultrasonic spectrum are a dog whistle and industrial ultrasonic cleaners. The frequency designations of the infrasound-audio boundary and the audio-ultrasound boundary are a bit arbitrary because the frequency range over which human beings hear sounds is different between people and additionally changes as a function of age.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0002.jpg

(a) Acoustic spectrum and (b) medical ultrasound spectrum.

Most medical ultrasound equipment operates in the ultrasonic frequency range between 1 and 15 MHz (Fig. 2b). Therapeutic (physical therapy, high-frequency focused ultrasound and ablation) applications operate around 1 MHz. For most diagnostic applications in abdominal, obstetrical and gynecological ultrasound, and in echocardiography, the frequency range is generally between 2.5 and 7.5 MHz. For superficial body parts, such as the thyroid and the eye, and peripheral vascular applications where ultrasound does not have to penetrate very deeply into the body, higher ultrasonic frequencies in the range of 7.5–15 MHz can be used because ultrasonic attenuation increases with increasing frequency.

2.2. Acoustic wave types

The classification of sound waves is based on the type of motion that is induced in the medium by the propagating sound wave. For purposes of ultrasonic physics, the lowest level of organization within a fluid (gas and liquid) is called a particle (Pierce, 1981; Kinsler et al., 1982). The particle is represented in Fig. 3 as dots and can be thought of as a volume of material. Each of these dots consists of millions of molecules and yet each has dimensions of a fraction of an ultrasonic wavelength.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0003.jpg

Representations of longitudinal and shear waves: (a) longitudinal wave representation, (b) shear wave representation and (c) sine wave representation.

When an ultrasonic wave is propagated within material, the type of wave is classified in terms of the direction the ultrasonic energy is traveling relative to the direction the particle is moving. A longitudinal wave occurs when the particles move back and forth (that is, left to right and back—horizontally in Fig. 3a) relative to the direction of the wave energy that is also moving horizontally. Propagated longitudinal waves travel through all kinds of materials: gases, liquids and solids.

In the case of shear waves, the particles move at right angles to the direction of the wave propagation as shown in Fig. 3b. In this figure, the particles are moving vertically up and down while the wave energy is moving horizontally. Shear waves exist only in solid materials, not in fluids. Shear waves do not exist in soft tissues because soft tissues are approximated as a liquid. Shear waves do, however, travel in harder biological materials such as bone.

The physical and thus ultrasonic properties of tissue are influenced by and composed of water, ions, macromolecules and cells and are a consequence of the chemical structures of fibrous and non-fibrous components. Tissues are divided into various kinds, including epithelial, muscular, connective, nervous, blood, etc. Each of these tissue types has different physical properties. Common to all tissues is a large amount of water. Selected physical properties of pure water at 37 °C (98.6 °F) are listed in Table 1 (Nyborg, 1975). The physical properties of tissue depend strongly upon water because water makes up almost three-quarters of the entire mass of the human body. The water concentration varies from tissue to tissue with vitreous humor quite high at around 99%, liver at 70%, skin at 60%, cartilage at 30% and adipose as low as 10%.

Table 1

Selected physical properties of pure water at 37 °C

Compressibility (Pa−1) 4.4×10−10
Bulk modulus (Pa) 2.3×109
Density (kg/m3) 990
Speed (m/s) 1527

2.3. Acoustic waveforms

The nomenclature of acoustic waveforms is used to define and quantify the ultrasonic wave that interacts with biological materials.

Ultrasound travels in waves that emanate from a source. The high crests and low troughs represent specific amplitude values of the wave and correspond to peak compressional and peak rarefactional values. The distance from one crest to the next, or from one trough to the next, has a particular distance associated with it and is called the wavelength and denoted by λ in Fig. 4a. The time that it takes for one cycle to occur is called the period (Fig. 4b). The period (T) is the reciprocal of frequency (f = 1/T).

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0004.jpg

Schematic representations of an acoustic waveform: (a) amplitude vs. distance and (b) amplitude vs. time.

The horizontal axis can illustrate either distance (Fig. 4a) or time (Fig. 4b). This is an important concept in diagnostic ultrasonic instrumentation. Distance information can be converted to time values, and time converted to distance information. Ultrasonic instruments are constantly performing these conversions in order to display sonographic images. The space (or distance) over which one cycle travels is called the wavelength and the time which one cycle occupies is called the period, that is, wavelength is “distance/cycle” and period is “time/cycle.” Speed is the constant that relates wavelength (λ) to period (T = 1/f ): c =λ_f_. For medical applications, the propagation speed, c, in tissue is typically assumed to be constant at 1540m/s.

There are two basic generation modes of ultrasound used in medical ultrasound (Fig. 5). Generation mode means the way in which the ultrasonic wave is “shaped” when it is transmitted from the ultrasonic transducer, that is, the waveform's temporal characteristics. One generation mode is to continuously excite the ultrasonic transducer with an electrical sine wave at constant amplitude. This produces a continuous ultrasonic wave at the same frequency as that of the electrical frequency and is termed continuous wave ultrasound (CW mode or CW ultrasound), as shown in Fig. 5a.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0005.jpg

Schematic representations of continuous wave and pulsed wave ultrasound waveforms: (a) continuous wave representation and (b) pulsed wave representation.

Another generation mode is to turn on the ultrasound for a short time duration and turn it off for a much longer time duration and then to repeat this process. This generation mode is accomplished by exciting or shocking the ultrasonic transducer with very short electrical signals, waiting for some time and then repeating the electrical shocking. The ultrasonic waves that are generated are termed pulse wave ultrasound (PW mode or PW ultrasound), as shown in Fig. 5b. If the number of cycles per pulse is N, then the pulse duration (τ) is

The ratio of the pulse duration to the pulse repetition period (PRP) is called the duty factor (DF). The DF is the fractional amount of time that the pulse is activated, and given by

where PRF is the pulse repetition frequency. For example, if the pulse duration is 1 ms and the pulse repetition period is 1 ms (PRF = 1 kHz), then the duty factor is 0.001, or 0.1%.

A measured 2.5-MHz center frequency ultrasound pulse is shown in Fig. 6; note how nonlinear the waveform appears. The maximum value of the ultrasonic waveform, at a magnitude of about 10 MPa, occurs when the particles are compressed (at positive pressure). Thus, this value is called the peak compressional pressure amplitude. The minimum value of the ultrasonic waveform, at a magnitude of about −4 MPa, occurs when the particles are rarefied (at negative pressure). Thus, this value is called the peak rarefactional pressure amplitude, and has the value of 4 MPa (no minus sign).

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0006.jpg

Measured 2.5-MHz center frequency pulsed wave ultrasound waveform.

2.4. Acoustic propagation speed, impedance and attenuation

An understanding of the issues related to propagation speed, impedance and attenuation in biological materials are directly applicable to the mathematical descriptions of biophysics mechanisms.

The propagation properties generally used to describe quantitatively the propagation of ultrasound in materials are speed, impedance and attenuation. The propagation of ultrasound is assumed to be an adiabatic process, that is, a process in which heat conduction does not occur. Therefore, the speed at which ultrasonic energy propagates in a fluid is (Pierce, 1981; Kinsler et al., 1982; Hall, 1987)

where the elastic modulus for an isotropic fluid is BAD, the adiabatic bulk modulus and the medium's density is ρ. As such, ultrasonic waves propagated in fluids are longitudinal waves. For a gas, _B_AD = γ_P_0, where γ is the ratio of specific heats and _P_0 is the ambient (equilibrium) pressure, and thus

For a liquid, the elastic modulus is _B_AD = γ_B_T, where _B_T is the isothermal bulk modulus and, therefore,

In an isotropic solid, both longitudinal and shear waves are supported wherein their respective propagation speeds are

cL=Y(1−σ)ρ(1+σ)(1−2σ)

(6a)

and

where Y is the Young's modulus and σ is the Poisson's ratio. Because σ is less than 0.5, _c_L is greater than cS.

The specific acoustic impedance of the wave is defined as the ratio of the acoustic pressure to particle velocity. For plane waves, the specific acoustic impedance is

where + is for the positive-going wave and − is for the negative-going wave. For other than plane waves, _Z_S generally different, that is, Z_S depends upon both the medium and the wave type (plane, cylindrical, spherical, etc). The ρ_c product is encountered frequently in analytic acoustics and is called the characteristic acoustic impedance (Z) of the medium or simply the acoustic impedance. Only for a plane wave are these two impedances numerically the same. The unit of the acoustic impedance is the rayl (kg/m2 s), after Lord Rayleigh. Table 2 summarizes the numerical ranges of ρ, c and Z for the various media.

Table 2

Typical density, propagation speed and characteristic acoustic impedance values for isotropic media

ρ (kg/m3) c (m/s) Z (rayl)
Gas 1 100–1000 100–1000
Liquid 1000 1000-2000 1–2×106
Solid 1000–10,000 2000–10,000 (L) 10–100×106 (L)
1500–5000 (S) 4–50×106 (S)

L denotes a longitudinal wave and S denotes a shear wave.

The classical engineering trade-off of diagnostic ultrasound instrumentation is that between resolution and the depth of the image (or penetration). Both are directly affected by the ultrasonic frequency and attenuation. As frequency is increased, resolution improves and penetration decreases. Resolution improves because the ultrasonic wavelength in tissue decreases (becomes a smaller number). Wavelength is inversely related to frequency; increase one and the other decreases.

As frequency increases, the ultrasonic attenuation also increases. Penetration is directly affected by tissue attenuation because it is approximately linearly related to frequency. At an ultrasonic frequency of 1 MHz, the attenuation coefficient is approximately 0.7 dB/cm whereas at 2 MHz, it is 1.4 dB/cm. Thus, attenuation coefficient is related to frequency; increase one and the other increases. It can be expressed mathematically by the expression 0.7 dB/cm MHz, a value that approximately represents soft tissue.

2.5. Acoustic waveform quantities in a medium

A necessary concept to understand axial (or range) resolution is the distance one cycle (and hence one pulse) occupies in a medium. The distance one cycle occupies in a medium is the wavelength, λ = c/f , where c is the propagation speed in the medium. For a repeated pulse waveform (Fig. 5b), the distance one pulse occupies in a medium is called the spatial pulse length (SPL), that is, the number of wavelengths per pulse where

Axial resolution is the ability to resolve discrete structures along the beam axis. Quantitatively, it is represented as the minimum distance between two structures at different ranges at which both can just be discretely identified as two separate structures. The best axial resolution is represented by the expression

best axial resolution=SPL2=Nλ2.

(9)

The transducer design affects the minimum number of cycles. More highly damped transducers (also referred to as low Q transducers) produce very few cycles of ultrasound when excited by the pulser voltage. As the frequency increases, and other quantities remain constant, axial resolution improves. The term “best axial resolution” has been employed because, in practice, the receiving and processing electronics affect axial resolution, as does the quality of the video monitor. The electronics and monitor are often lumped into the term “system Q.” Low-valued system Q's provide better axial resolution than do high-valued ones. As a “rule of thumb,” there are Q/2 cycles of pressure contained in the pulse, that is, N = _Q_=2, which yields

best axial resolution=Nλ2=N2cf=Q4cf=c4Δf,

(10)

where the quality factor Q is defined as the ratio of the center frequency, f, to the system bandwidth, Δ_f_. For a propagation speed of 1540 m/s,

best axial resolution(inmm)=0.77Δf,

(11)

where Δ_f_ is in MHz.

However, ultrasound images are speckle images and therefore a more representative expression for axial resolution is

where FWHMA is the axial full-width half-maximum length of the pulse in millimeters and Δ_f_ is in MHz. This expression is also only a function of the system bandwidth but yields a numerical value for axial resolution about 1.8 times greater than the best axial resolution. Thus, the axial resolution improves (its numerical value decreases) when the bandwidth increases.

Lateral resolution is the ability to resolve discrete structures perpendicular, or lateral, to the beam axis. Quantitatively, it is represented as the minimum distance between two side-by-side structures at the same range at which both can just be discretely identified as two separate structures. The best lateral resolution is equal to the minimum beam width; the best lateral resolution term is employed here for the same reasons as that of the term best axial resolution. For a focused ultrasonic field, the beam width (BW) at the focus is

where ROC is radius of curvature (in measurement practice ROC is the distance between the source and the center of the focal region, the focal length) and D is the diameter for a circular source or linear end-to-end lengths for a rectilinear source. In imaging terminology, the term “f_-number” or “_f#” is often used to quantitative focusing where the lower the _f_-number value, the better is the focusing. In terms of the full-width half-maximum length, the beam width at the focus is

where FWHML is the lateral full-width half-maximum length and L is the focal length (basically the same as ROC.

3. First- and second-order quantities

There are many buzzwords to describe a general class of events such as the terms _first_- and second-order quantity. Quantity represents what is measured and the value of a quantity is generally expressed as the product of a number and a unit (Table 3).

Table 3

Typical ultrasonic quantities and units

Quantity Unit
Charge Coulomb (C)
Current Ampere (A = C/s)
Displacement Meter (m)
Energy Joule (J = Ws)
Energy density Joule per meter cubed (J/m3 = N/m2)
Force Newton (N)
Frequency Hertz (Hz)
Intensity Watt per centimeter squared (W/cm2)
Length Meter (m)
Mass Kilogram (kg)
Power Watt (W)
Speed Meter per second (m/s)
Temperature Degree celsius (°C)
Time Second (s)
Ultrasonic pressure Pascal (Pa = N/m2)
Voltage Volt (V)
Wavelength Meter (m)

First-order quantities are known as amplitude quantities and second-order quantities as energy-based quantities (Table 4). The basic ideas of first- and second-order quantities are (1) both first- and second-order quantities deal with the transport of energy, (2) all first-order quantities are directly proportional to each other, (3) all second-order quantities are directly proportional to each other, and (4) the product of any two first-order quantities is directly proportional to any second-order quantity.

Table 4

List of first-order and second-order quantities used in ultrasound

First-order quantities Second-order quantities
Current Energy
Particle acceleration Energy density
Particle displacement Intensity
Particle velocity Power
Ultrasonic pressure
Voltage

Acoustic wave propagation, and the development of its wave and other equations (Morse and Ingard, 1968; Nyborg, 1978; Pierce, 1981; Kinsler et al., 1982; Hall, 1987; Ensminger, 1988; O'Brien, 1992a; Blackstock, 2000), can be approached from the Equation of State which relates the change in density to the change in pressure, the Continuity Equation which relates particle motion to the change in density by invoking conservation of mass and the Momentum Equation (becomes Euler's Equation for a lossless medium at rest) which relates the change in pressure to particle motion through Newton's Second Law of Dynamics by invoking conservation of momentum. These equations and their various forms are: Equation of State

p=c02δρ{1+B2!Aδρρ0+C3!A(δρρ0)2+⋯},

(15a)

p=c02δρ{1+B2!As+C3!As2+⋯}.

(15b)

Continuity Equation

Momentum Equation

ρ0∂u→∂t+∇p=0(linear Euler’s Equation).

(17b)

The total or material derivative is

where the first term on the right-hand side is the time rate of change of q (any first-order acoustic variable) the fluid particle would experience if it were at rest (u→=0), and the second term is the additional rate of change caused by the particle's movement. Also, p is the acoustic pressure (instantaneous pressure P = _P_0 + p), δρ is the excess density (instantaneous density ρ = ρ0 + δρ), B/A is the coefficient of the first nonlinear parameter (Beyer, 1997), s is the condensation, the fractional change in density (δρ/ρ0) and u→ is the particle velocity of a fluid element.

Pressure P, velocity u→, density ρ can be expressed as

P = _P_0 + _p_1 + _p_2 + ,

(19a)

ρ = ρ0 + ρ1 + ρ2 + ,

(19c)

where the subscripts indicate the order. For example, _P_0 is the zero-order contribution to pressure, _p_1 is the first-order contribution that varies sinusoidally for a harmonic (CW) wave at frequency ω, and _p_2 is the second-order contribution that has both a temporal-dependent component at frequency 2ω and a temporal-independent component. Because the fluid is assumed to be at rest, the zero-order contribution to u→ is zero.

3.1. First-order quantities

The Equation of State, the Continuity Equation and the Euler's Equation for first-order contributions become, respectively,

In water and tissue, n B(∂s_1/∂_t)[unk]_Bs_1. Thus, the Equation of State becomes _p_1 = _Bs_1. Eliminating the order 1 subscripts and noting that u→=dξ→∕dt, where ξ→ is the particle displacement, by combining these equations for a one-dimensional wave propagating in the x direction yields the one-dimensional lossless wave equation

The one-dimensional lossless wave equation can be described by the particle displacement ξ(x, t), or can likewise be described by the particle velocity u(x, t), the particle acceleration a(x, t), or the acoustic pressure p(x, t). In terms of the particle displacement, the one-dimensional lossless wave equation traveling in the positive x direction is represented as

ξ(x, t) = ξ0 cos(ω_t_ − kx),

(24)

where ξ0 is the particle displacement amplitude, ω is the angular frequency, t is time and k is the wave number (also called the propagation constant).

For plane waves, particle velocity, particle acceleration and acoustic (ultrasonic) pressure are determined, respectively, from

p(x,t)=−ρc2∂ξ(x,t)∂x,

(27)

where Eq. (27) is determined by combining the Equation of State and the Continuity Equation to yield p = ρ0_c_2_s_, where s = −∂ξ/∂x for a plane wave. All first-order plane wave ultrasonic amplitude quantities are directly proportional to each other. These quantities are

where _U_0, _A_0 and _p_0 are the particle velocity amplitude, particle acceleration amplitude and acoustic (ultrasonic) pressure amplitude, respectively.

For the lossy one-dimensional wave equation, the medium's attenuation coefficient is part of the solution wherein

ξ(x, t) = ξ0_e_−Ax_cos(ω_tkx),

(29)

where A is attenuation coefficient.

3.2. Second-order quantities

When an ultrasound wave propagates in tissue, a mechanical strain is induced, where strain refers to the relative change in dimensions or shape of the body that is subjected to stress. From the second-order contribution to the Momentum Equation (Eq. (17)), the gradient of P, ∇P, a force quantity, is

where F→ is a temporal and spatial varying force per volume (in N/m3), the volume being a fluid element.

Also, ultrasonic wave propagation transports and dissipates energy, and second-order quantities are proportional to energy. Quantitatively, energy is represented in terms of energy density (a scalar) and intensity (a vector). For a plane wave propagating in the x direction, the instantaneous kinetic and potential energies are, respectively,

where u and p are the respective instantaneous values of particle velocity and acoustic pressure.

To evaluate the temporal-average energy density, the one-dimensional, harmonically varying particle velocity is assumed to be

u(x, t) = U_op cos(ω_tkx) + U_on cos(ω_t + kx),

(33)

where _U_op and _U_on are the particle velocity amplitudes for the positive and negative directed components, respectively, and the one-dimensional, harmonically varying ultrasonic pressure is

p(x, t) = p_op cos(ω_tkx) + p_on cos(ω_t + kx),

(34)

where _p_op = ρ_cU_op and _p_on = −ρ_cU_on. Therefore, the average energy density is

〈E〉=1T∫0TE(x,t)dt=ρ2(Uop2+Uon2).

(35)

Intensity is an extremely useful ultrasonic quantity that represents a measure of ultrasonic power flowing (temporal-average rate of flow of energy) at normal incidence to a specified unit area. The intensity concept is generally applied in connection with a traveling plane wave. Further, it is a vector quantity but, because the development herein is confined to a fluid and to the one-dimensional wave equation, vector notation is not used; the direction is known. The instantaneous intensity is defined as the dot product of the ultrasonic pressure and particle velocity but because these two quantities are in phase, the dot product is pu. Its temporal-average representation is given by

I=1T∫0Tpudt=ρc2(Uop2−Uon2).

(36)

It should be noted that for a standing wave where Uop2=Uon2, the temporal-average intensity (a vector) is zero whereas the temporal-average energy density (a scalar) is not.

For a progressive ultrasonic plane wave propagating in only the positive x direction, Uon2=0, Eqs. (35) and (36) become, letting Uop2=U02,

and

I=ρc2U02=12ρcp02=p0U02,

(38)

where 〈_E_〉 is the temporal-average energy density (in J/m3 or N/m2) and I is the temporal-average intensity (in W/m2 or, more conventionally in ultrasonic biophysics, W/cm2). Combining these results yields

which is a useful expression in terms of measuring ultrasonic intensity and ultrasonic power with radiation force techniques, but only valid for plane waves.

The temporal-average energy density is equivalent to the radiation force (in N) for a perfect absorber in that

where A is the area of the absorber. The product term IA is acoustic power W, thus,

For a perfect reflecting surface, the radiation force is twice that of an absorbing target.

If the medium is lossy, and the loss is assumed to be purely absorptive with an absorption coefficient α, then for a one-dimensional wave propagating in the positive x direction, the particle velocity (from Eqs. (25) and (29)) and temporal-average intensity can be described by

u(x, t) = U_0e−α_x sin(ω_t_ − kx),

(42)

where _I_0 is the intensity at x = 0. From these two expressions, and the temporal-average value of F→ (Eq. (30)), the temporal-average radiation force on the medium is

4. Dosimetry

Dosimetry is the determination of a dose, or similar type of physical quantity, that characterizes the physical agent as to its potential or actual interaction with the biological material of interest. Ultrasonic dosimetry's objective is to relate magnitudes of specific quantities, such as intensity, acoustic pressure, particle displacement, etc., or perhaps some quantity yet to be developed, to the likelihood of occurrence of a biological alteration. To accomplish this, it is necessary (1) to quantify the output quantity or quantities of the source, (2) to determine the effect of the material on the propagating energy, viz., reflections, refraction, scattering, absorption, etc., and (3) to relate quantitatively the first two items at the site of interest.

Typically, dose connotes something that is given or imparted in a quantitative manner. The history of other radiation forms has documented that defining dose, or dose-like concepts, is difficult, especially when the objective is to include all possible physical and biological variables. More commonly, however, special quantities are developed for the biological action under consideration. In ionizing radiation, for example, dose generally refers to the quantity absorbed dose that has been defined as the energy imparted to matter by ionizing radiation per unit mass of irradiated material at the site of interest (ICRU, 1971; CGPM, 1975, 1976). But other dose quantities have been defined for specific purposes such as genetically significant dose, cumulative dose, dose equivalent, threshold dose, etc. (BEIR, 1972, 2005). In photobiology, dose sometimes refers to the quantity dose of ultraviolet radiation which has been defined as the energy per unit surface area applied to an object (Rupert, 1974). There has been much discussion regarding microwave dosimetry. Terms such as specific absorption rate, absorbed power density and specific absorption density and energy dose-rate have been used as a basic quantity to describe absorbed electromagnetic energy (Anderson, 1992; Bernhardt, 1992; Grandolfo, 1992; Leonowich, 1992; Sliney, 1992).

By comparison, the field of ultrasonic dosimetry has not developed nearly to the extent of ionizing radiation dosimetry. The most widely used quantity in ultrasonic bioeffect and biophysical studies is intensity in the unit of W/cm2. The principal reason for the use of intensity is, perhaps, convenience because it is understood how it is measured (IEEE, 1990; Lewin and Schafer, 1992). However, intensity represents many of the same problems as does the ionizing radiation quantity “exposure” in that it is not a measure of dose. Yet the majority of bioeffect and biophysical reports use intensity as the measured physical quantity of the ultrasonic field. This extensive literature documents the actions of ultrasound but, in most cases, lacks the necessary characterization of the field at the site of interest. An ideal situation would be to know the instantaneous particle velocity, the instantaneous acoustic pressure and the phase between these two field parameters at the site or sites of interest (O'Brien et al., 1972; O'Brien, 1978, 1986, 1992b).

There have been early ultrasonic dosimetric quantities that are noteworthy of comment in that they represent, in concept, the basic approach to dosimetry. The cataract-producing unit, CPU, was a quantity defined as the duration of exposure necessary to produce a grossly observable cataract and expressed in units of seconds (Purnell et al., 1964). The dosimetric concept damage ability index with the unit second is a quantity intended to describe the effect of ultrasound on spinal cord hemorrhage (Taylor and Pond, 1972). It has been suggested (Johnston and Dunn, 1976) that a universal dosimetric response to ultrasonic exposure may exist for different tissues but the response has only been demonstrated, in a limited manner, in mammalian brain tissue. The response was in terms of energy absorbed per unit volume for histologically observable lesions at superthreshold levels as a function of the delivered intensity. It was shown that at two different ultrasonic frequencies, 3 and 4 MHz, identical constant volume curves resulted even though there were two different threshold levels (Dunn and Fry, 1971). Later, a damage integral was defined to predict the occurrence and dimensions of thermally induced ophthalmic lesions (Lizzi et al., 1984).

In another category of ultrasonic dosimetric studies, in utero ultrasonic intensity in both the gravid and non-gravid human uterus has been estimated (summarized in Stewart and Stratmeyer, 1982; NCRP, 1983). In these early studies, a model of the tissue layers between the skin surface and fetal sac yielded a total attenuation in the range of 2–20 dB at frequencies between 2 and 5 MHz. The distances between the abdominal surface and the uterine cavity in early pregnancy ranged between 2 and 11 cm. In later work (Carson et al., 1989), similar distances were estimated to be 2.6 cm.

In human-based studies wherein direct in utero ultrasonic pressure measurements were made (Daft et al., 1990; Siddiqi et al., 1991), the average attenuation was reported to be 6.2±3.5 dB under full bladder conditions and 7.3±4.9 dB under empty bladder conditions. Applying the fixed-attenuation tissue model (attenuation dependent upon frequency and independent of distance), and normalizing to the center frequency of 2.4 MHz, the attenuation coefficient was estimated to be 2.56±1.47 dB/MHz for the full bladder condition whereas, with the overlying tissue model (attenuation dependent upon frequency and non-fluid distance), the attenuation coefficients were estimated to be 0.89±0.71 dB/cm MHz and 0.45±0.32 dB/cm MHz, respectively, for full and empty bladder conditions. The mean values for the fixed-attenuation tissue model's attenuation coefficient were about a factor of 3 greater than the values proposed to model the attenuation coefficient by Carson et al. (1989). FDA's Center for Devices and Radiological Health uses a value of 0.3 dB/cm MHz as a derating factor (a tissue attenuation coefficient) for manufacturers in their 510(k) process to estimate ultrasonic intensity quantities in tissue (FDA, 1985; Harris, 1992). The measured mean values for the overlying tissue model's attenuation coefficient were a factor of 2–3 greater than the values used by FDA suggesting that FDA's values error on the side of safety.

In general, it is necessary to determine a firm database from which various dosimetric modeling approaches can be explored. In one approach (Carson et al., 1989), a worst-case approach was employed to overestimate safety whereas, in another approach (Siddiqi et al., 1991), the mean values of the results and their distributions were reported so that the scientific community could make the appropriate safety judgments. But for ultrasonic dosimetry, neither an approach to monitor energy absorbed per unit mass nor to monitor the biological danger have yet been developed. The only adopted approach is that developed in the Standard for Real-Time Display of Thermal and Mechanical Indices on Diagnostic Ultrasound Equipment, commonly known as the Output Display Standard (ODS) and its revisions (ODS, 1992, 1998, 2004), and discussed below under thermal (Section 6) and non-thermal (Section 7) mechanisms.

5. A brief history

The development of the mechanisms responsible for ultrasound-induced biological effects is closely linked to the unique history of the development of diagnostic ultrasound applications in human medicine and the basis for the development of standards (O'Brien, 1998). A brief overview of the historical development of ultrasonic biophysics is thus provided. In many cases, the same pioneers whom development ultrasound applications also elucidated the ultrasound biophysical mechanisms.

More than three decades after the 1880 discovery of the piezoelectric effect by the brothers Paul-Jacques and Pierre Curie (Curie and Curie, 1880), a discovery which revolutionized the production and reception of high-frequency sound, the French scientist Paul Langevin developed one of the first uses of ultrasound for underwater echo ranging of submerged objects with a quartz crystal at an approximate frequency of 150 kHz (Hunt, 1982). Langevin was, perhaps, the first to observe that ultrasonic energy could have a detrimental effect upon biological material wherein he reported (Langevin, 1917) “fish placed in the beam in the neighborhood of the source operation in a small tank were killed immediately, and certain observers experienced a painful sensation on plunging the hand in this region.” Langevin also reported observing incipient cavitation in water when the source was active.

Another decade passed before a more detailed, experimental study was conducted (Wood and Loomis, 1927) to investigate Langevin's (1917) observation. Although the ultrasonic levels were not specified, their experimental studies showed that ultrasonic energy had a range of effects from rupture of Spirogyra and Paramecium to death of small fishes and frogs by a 1–2-min exposure, the latter also observed by Langevin with a Poulsen arc oscillator. Considerable work followed and in the earliest review paper on this subject, Harvey (1930) reported on the physical, chemical, and biological effects of ultrasound in which alterations were produced in macromolecules, microorganisms, cells, isolated cells, bacteria, tissues, and organs with a view towards the identification of the interaction mechanisms. The ultrasonic exposure conditions of these early works were neither well characterized nor reported, but the exposure levels were undoubtedly high.

Ultrasound-induced tissue heating was applied extensively as a therapeutic agent in the 1930s and 1940s. However, while it was clear that ultrasound could effectively heat tissue, and excess enthusiasm resulted in numerous clinical applications being proposed and tried, the inferior clinical experience caused this modality to fall into disfavor (see discussion of 1949 Erlangen resolution (Kremkau, 1979)). Thus, as Hill (1973) observed of this time, “it is perhaps unfortunate that the generation of ultrasound proved to be so relatively simple and cheap that a considerable practice was built up.”

Also, during the 1930s and 1940s, with an understanding that ultrasound at sufficient levels could have a dramatic effect on tissues, and produce large temperature increases, the potential for ultrasonic surgery was proposed. This ability to non-invasively burn focal tissue volumes deep in the body using ultrasound was first proposed in 1942 (Lynn et al., 1942, 1944) as a neurosurgery technique. Ultrasound surgery and its biophysical mechanism (heating) were further developed in the late 1940s and early 1950s (Fry et al., 1955). Also proposed in 1948 and applied in 1952 was the application of ultrasound surgery to destroy the vestibular function to treat the symptoms of Menie`re's disease (Sjoberg et al., 1963).

While ultrasound exposimetry was inferior in these early times to that possible today, these early bioeffect studies clearly demonstrated that ultrasound, at sufficient levels, could easily destroy biological material. From the earliest considerations that ultrasound might be a feasible energy source for producing images of the human body, it was known that high ultrasonic energy levels had the potential to be hazardous. This information must be kept in mind because many of the early pioneers who were developing ultrasound imaging devices were also cognizant that ultrasound had the potential for disrupting biological materials. Additionally, they were aware from the history of ionizing radiation to be concerned with the potential for hazardous effects. Thus, it must be presumed that there has been a continuing concern of the risk from ultrasound prior to and during the entire period of diagnostic instrumentation development.

During the 1950s and 1960s, the ability to quantify ultrasonic fields improved but only to a limited extent; there were still no nationally based ultrasound measurement standards or procedures. All of the improvements dealt with absolute procedures to quantify second-order quantities, and consisted of ultrasonic intensity via thermocouple probe (Fry and Fry, 1954a, b; Fry and Dunn, 1957; Dunn and Breyer, 1962) and electrodynamic method (Filipczynski, 1967), and ultrasonic power via radiation pressure and calorimetry (Wells et al., 1963) and radiation pressure balance (Newell, 1963; Kossoff, 1965).

It should be noted that as early as 1951 (French et al., 1951), individuals working on the development of diagnostic ultrasound equipment had an awareness of a potential risk to patients. Also in the early 1950s the same investigators active in image system developments were also investigating the nature of ultrasound-induced tissue damage (Fry et al., 1950, 1951, 1970; Hueter et al., 1956; Baum, 1956), thus demonstrating a concern for the safety of this diagnostic modality.

This period saw only a few advances in our understanding of how ultrasound interacted with biological materials. Perhaps the first major symposium on “Ultrasound in Biology and Medicine” was held at the University of Illinois in 1952 to examine phenomena of how ultrasonic energy interacted with and acted upon biological materials. Of the eight papers presented, six were published and dealt with the effects of high-intensity ultrasound (Fry, 1953; Wall et al., 1953; Wild and Reid, 1953) or the thermal mechanism of ultrasound (Fry and Fry, 1953; Herrick, 1953; Lehmann, 1953). Two additional symposia were held (June, 1955, 1962) to address similar issues (Kelly, 1957, 1965). This literature laid the basic foundation for the biophysical mechanisms by which ultrasound is known to affect biological materials, viz., thermal and cavitation.

There were scattered reports about ultrasound having an effect on biological systems during this period, but these reports in general did not deal with the types of exposure that would be expected from diagnostic ultrasound equipment, nor did they provide any kind of a consistent message. Two summaries identify a number of these early reports (Reid and Sikov, 1972; Stewart and Stratmeyer, 1982). However, during the 1950s and 1960s, as diagnostic ultrasound equipment started to be used in clinical medicine, there continued to be concerns about its safety, which could not be dispelled because of the paucity of well-conducted and comprehensive experimental studies.

One interesting observation made in the late 1960s that would wait more than two decades before it would become an active area of basic and clinical research was the identification of the production of cavitation at the tip of catheters when various liquids were injected through the catheter. The cavitation was a primary source of echoes in an echocardiograph image, and the first application of an injected contrast agent was identified (Kremkau, 1969; Kremkau et al., 1970).

The 15-year period between early 1970s and mid-1980s witnessed the greatest expansion with diagnostic ultrasound imaging capabilities, starting with bistable, static and ending with gray-scale, real-time capabilities. During this period, the ability to quantify ultrasonic fields improved considerably. Perhaps the first intercomparison (between two universities) to assess the absolute measurement of ultrasonic intensity was conducted (Breazeale and Dunn, 1974); the comparison was conducted with the elastic sphere radiometer (Dunn et al., 1977). A major breakthrough of earlier work (Brain, 1924; Fukada, 1968) occurred with Kawai's discovery in 1969 (Kawai, 1969) of the strong piezoelectric effect in polyvinylidenefluoride (PVDF) to measure the temporal characteristics of diagnostic ultrasound fields. Two types of PVDF hydrophones were developed, viz., needle (Lewin, 1981) and membrane (DeReggi et al., 1981; Bacon, 1982; Harris, 1982; Preston et al., 1983). The US National Bureau of Standards (now the National Institute of Standards and Technology, NIST) developed an ultrasound power transfer standard (Fick et al., 1984), and the UK National Physical Laboratory developed both a two-transducer reciprocity technique and an optical technique (Smith, 1986).

This 15-year period had its controversies relative to whether diagnostic ultrasound was safe. Two of the controversies are briefly reviewed to illustrate the level of concern the ultrasound community felt because these reports dealt with the potential of diagnostic ultrasound affecting chromosomes. In 1970 it was reported that ultrasound from a low-output commercial fetal pulse detector-induced aberrations in human lymphocyte chromosomes (Macintosh and Davey, 1970, 1972). Numerous laboratories throughout the world attempted to replicate the findings without success. The matter was brought to closure when, in 1975, the same lead author of the original report was invited to another laboratory in an attempt to duplicate the experiment; they were unable to reproduce the original findings (Macintosh et al., 1975). In the late 1970s, a report appeared (Liebeskind et al., 1979) which suggested that exposure from a diagnostic ultrasound system showed an increase in human lymphocyte sister chromatid exchange (SCE) frequency (an indication of chromosome damage). Like with the previous chromosomes aberration situation, numerous laboratories throughout the world attempted to replicate the findings. In 1984, the American Institute of Ultrasound in Medicine's Bioeffects Committee carefully and thoroughly reviewed a total of 14 ultrasound-SCE studies and concluded that these studies do not suggest a hazard from exposure to diagnostic ultrasound (Goss, 1984).

In the early to mid-1970s, there was great uncertainty with respect to the safety of ultrasound as discussed above and what authority and role the US Food and Drug Administration (FDA) would take in terms of regulating diagnostic ultrasound. Regulatory control of diagnostic ultrasound in the United States can be traced to passage of the 1976 Medical Device Amendments to the Food, Drug and Cosmetic Act, but for the several years prior to its passage, the FDA coordinated a classification scheme for all medical devices. There was apprehension among the public, patients, physicians, sonographers, basic scientists, manufacturers, and the government. One of the fundamental difficulties was the lack of an accurate and precise procedure to quantify diagnostic ultrasound equipment outputs. Because of these difficulties, the output levels from diagnostic ultrasound equipment were not well characterized, and human exposure levels could not be compared to results from laboratory experimental studies. In response to this uncertainty, and the lack of a suitable measurement scheme, the American Institute of Ultrasound in Medicine (AIUM), a medical and scientific professional society, and the National Electrical Manufacturers Association (NEMA), a trade organization that represented many of the ultrasound manufacturers, joined efforts in 1976 to develop a voluntary standard that would assure that sufficient information on the characteristics of diagnostic ultrasound equipment was supplied to allow medical personnel to make informed judgments regarding the application of this equipment to patients. The Safety Standard for Diagnostic Ultrasound Equipment (AIUM/NEMA, 1983) was developed over a 5-year period and set forth precise definitions, output measurement procedures and labeling requirements related to those characteristics of ultrasound equipment that were believed at that time to pertain to patient exposure and safety. The voluntary standard's labeling requirements called for manufacturers to make available (publicize) to the ultrasound community the maximum values of the following ultrasonic quantities: ultrasonic power; spatial peak, temporal average intensity (_I_SPTA); and spatial peak, pulse average intensity (_I_SPPA). The labeling requirements were based on the philosophy that there was a potential risk from diagnostic ultrasound exposure and included those quantities whose magnitudes were known or believed to be related to actual damage or to risk of damage to biological tissues; they were the quantities most often reported in the basic science literature to relate the strength of ultrasound to the production of biological effects in laboratory experimental studies. The voluntary standard did not specify upper limits.

In the early1980s, the pioneering observations by Apfel (1981a, 1982, 1986), Flynn (1982) and Flynn and Church (1988) from their mathematical models to describe the dynamic behavior of small bubbles (or microbubbles) in liquids suggested the strong possibility for transient cavitation (now termed inertial cavitation) to occur from microsecond pulses of ultrasound. This opened up the scientific debate that cavitation in tissue might occur from diagnostic ultrasound exposures.

Also in the early 1980s, the assessment of diagnostic ultrasound risk was addressed by two major activities. One of the activities was sponsored by the US National Institutes of Health consensus development conference processes (NIH, 1984) by convening an expert panel of physicians, basic scientists, epidemiologists, nurses, educators, and sonographers to provide answers to specific questions related to safety and efficacy of diagnostic ultrasound in obstetric practice. The document indicated that the increasing use of ultrasound during pregnancy is safe and effective for 28 medical conditions. This was, perhaps, the first time that the issue of diagnostic ultrasound efficacy was critically reviewed. Also, this process recommended against routine scanning of the embryo and fetus. Further, it was suggested that while diagnostic ultrasound does not appear to be associated with any known hazards, investigators should continue to evaluate risks.

The other activity was conducted under the auspices of the National Council on Radiation Protection and Measurements (NCRP, 1983). The document rigorously covered the basic physics of ultrasound with an emphasis on medical ultrasound fields and on the quantification of various ultrasonic field quantities to which humans were exposed during the course of an ultrasound examination. Also included were mechanisms by which ultrasound interacts with biological material and effects caused by ultrasound on biological materials such as plants, animals, and in vitro systems. Finally, this document set forth a number of recommendations that fell into the categories of research needs, industrial practices, education, and exposure criteria. It is interesting that a number of the recommendations were consistent with those put forth by the Workshop on the Interaction of Ultrasound and Biological Tissues (Reid and Sikov, 1972) a decade earlier.

When the FDA initiated the regulation of diagnostic ultrasound equipment in the mid-1980s (FDA, 1985), it set application-specific intensity limits in their 1985 “510(k) premarket notification” which manufactures could not exceed (Table 5). This notification is used by the FDA to determine if the new devices are substantially equivalent, in safety and effectiveness, to diagnostic ultrasound devices on the market prior to May 28, 1976, the date when the Medical Device Amendments were enacted. The exposure quantities required by FDA were, in part, similar to those in the Safety Standard for Diagnostic Ultrasound Equipment voluntary standard's labeling requirements (AIUM/NEMA, 1983). However, the limits were not based on safety considerations but rather on the maximum output limits of known diagnostic ultrasound equipment at the time when the Medical Devices Amendments were enacted, in May, 1976; hence the term pre-amendments levels (O'Brien et al., 2002).

Table 5

FDA's pre-amendments levels of diagnostic ultrasound devices (FDA, 1985)

Derated intensity values
_I_SPTA (mW/cm2) _I_SPPA (W/cm2) _I_m (W/cm2)
Cardiac 430 65 160
Peripheral vessel 720 65 160
Ophthalmic 17 28 50
Fetal imaging and othera 46 65 160

To emphasize the date-base regulation approach as opposed to the safety- and efficacy-base regulation approach of the FDA, the American Institute of Ultrasound in Medicine notified the FDA in mid-1986 (AIUM, 1986) that there existed prior to May 28, 1976 at least two diagnostic ultrasound devices (pre-enactment ultrasound devices) with intensity levels greater than those listed in Table 5. In early 1987, the FDA updated their diagnostic ultrasound guidance to higher intensity levels (FDA, 1987) to those listed in Table 6. The date-based regulation approach has been criticized by technical, scientific, and medical professionals, as well as by the diagnostic ultrasound industry (Merritt, 1989) because of the implication that these arbitrary limits are safety based, and because they could limit future clinical benefits by preventing the development of more advanced diagnostic ultrasound systems, and hence greater clinical benefit, that may require higher output levels. Further, it must be recognized that limited diagnostic ultrasound capabilities may, in fact, be responsible for greater risk to the patient due to either an inadequate diagnosis, or to the use of an additional diagnostic procedure with a defined risk.

Table 6

FDA's pre-amendments levels of diagnostic ultrasound devices (FDA, 1987)

Derated intensity values
_I_SPTA (mW/cm2) _I_SPPA (W/cm2) _I_m (W/cm2)
Cardiac 430 190 310
Peripheral vessel 720 190 310
Ophthalmic 17 28 50
Fetal imaging and othera 94 190 310

Within the past decade or so, there have been substantial improvements, through national and international efforts, in our understanding to identify the biological interactions diagnostic ultrasound exposures may have with soft tissues, including both thermal (NCRP, 1992; WFUMB, 1992; AIUM, 1993) and acoustic cavitation-like (AIUM, 1993, 2000; WFUMB, 1996) phenomena. These reports have provided, in part, the basis for the development of a new approach to regulate diagnostic ultrasound equipment (see Sections 6.2 and 7.1).

It had been long thought that diagnostic ultrasound exposures could not produce biological damage. However, in the early 1990s, based on initial observations in the late 1980s from lithotropsy experiments, diagnostic ultrasound-type pulses at diagnostic pressure levels have produced significant damage (lung hemorrhage) in mice, rats, rabbits, and monkeys, the first observation being in mice (Child et al., 1990). There are open questions yet about whether this effect occurs in humans.

Ultrasonic biological effect studies and biophysical research have shown that ultrasound can produce changes in living systems. The AIUM/NEMA Ultrasound Safety Standard for Diagnostic Ultrasound Equipment (AIUM/NEMA, 1983) and AIUM Acoustic Output Measurement and Labeling Standard for Diagnostic Ultrasound Equipment (AIUM/NEMA, 1992) labeling requirements were based on the philosophy that there is a possible risk from diagnostic ultrasound exposure. The specific labeling requirements of these and other (Harris, 1992) safety standards were selected to include those quantities whose magnitudes are known or believed to be related to actual damage or to risk of damage to biological tissues as a result of ultrasonic irradiation.

The basis for this rationale lies in an understanding of the mechanisms by which it is known that ultrasound can affect living systems. Such knowledge comes from fundamental laboratory studies (O'Brien, 1984, 1991; O'Brien and Withrow, 1985). These mechanisms are classified and discussed in terms of whether heat is or is not believed to be the principal cause for the biological effect. The applicable ultrasonic exposure quantities are identified during the course of this discussion. Both thermal and non-thermal mechanisms are considered.

6. Thermal mechanism

Whenever ultrasonic energy is propagated into an attenuating material such as tissue, the amplitude of the wave decreases with distance. This attenuation is due to either absorption or scattering. Absorption is a mechanism that represents that portion of the wave energy that is converted into heat, and scattering can be thought of as that portion which changes direction. Because the medium can absorb energy to produce heat, a temperature increase may occur as long as the rate at which heat is produced is greater than the rate at which the heat is removed (O'Brien, 1978; NCRP, 1983, 1992, 2002). The thermal mechanism is relatively well understood because increase in temperature produced by ultrasound can be calculated using mathematical modeling techniques (Robinson and Lele, 1972; Nyborg, 1975, 1981; Lerner et al., 1973; Cavicchi and O'Brien, 1984, 1985; Nyborg and Steele, 1983; Nyborg and O'Brien, 1989; Curley, 1993; Lubbers et al., 2003) and has been estimated for a variety of exposure conditions (NCRP, 1983, 1992).

In tissue, at the site where the ultrasonic temporal-average intensity is _I_TA, the rate of heat generation per unit volume is given by the expression (Nyborg, 1981; Cavicchi and O'Brien, 1984)

where

where α is the ultrasonic amplitude absorption coefficient which increases with increasing frequency, p and p* are the instantaneous ultrasonic pressure and its complex conjugate, respectively, ρ is density and c is sound speed. The product of p and p* is equal to the ultrasonic pressure amplitude squared, p02, at the specific location in the medium where Q. is determined and can be thought of as a temporal-average quantity.

The temporal-average intensity is not necessarily at the location where it is maximized, that is, at the spatial peak location. If it were, however, then _I_TA (Eq. (46)) would be the spatial peak, temporal peak intensity _I_SPTA, which would maximize Q. for that tissue site. AIUM's Statement on Mammalian In vivo Ultrasonic Biological Effects (Table 7), sometimes referred to as the 100 mW/cm2 Statement, is a generalization about the state-of-affairs with respect to an intensity–time limit (in terms of _I_SPTA) below which there have been no independently confirmed significant biological effects in mammalian tissues (AIUM, 1988).

Table 7

AIUM statement on mammalian in vivo biological effects (AIUM, 1988)

(Approved August, 1976. Revised and approved October, 1987)
A review of bioeffects data supports the following statement:
In the low megahertz frequency range there have been (as of this date) no independently confirmed significant biological effects in mammalian tissues exposed in vivo to unfocused ultrasound with intensitiesa below 100mW/cm2, or to focusedb ultrasound with intensities below 1W/cm2. Furthermore, for exposure timesc greater than one second and less than 500 s for unfocused ultrasound, or 50 s for focused ultrasound such effects have not been demonstrated even at higher intensities, when the product of intensity and exposure time is less than 50 joules/cm2.

For a given _I_TA, the maximum temperature increase Δ_T_max, under the assumption that no heat is lost by conduction, convection, or any other heat removal processes, is approximately described by (Fry and Fry, 1953)

where Δ_t_ is the time duration of exposure and Cv is the medium's heat capacity per unit volume. This formula is valid only for short exposure times; for longer exposure times, heat removal processes become significant. Nonetheless, as a “ballpark estimate,” using the intensities from the AIUM Statement in Table 7 of _I_SPTA = 0.1 and 1 W/cm2 at an ultrasonic frequency of 5 MHz, from Eq. (45), Q.=0.05 and 0.5 J/cm3 s (α≈0.25/cm at 5 MHz). Because the thermal properties of biological tissue can be approximated by water (Cv = 4.18 J/cm3 °C), the maximum rates of change of temperature are

ΔTmaxΔt=0.012and0.12°C∕s,

(48)

which means that for a 1 s exposure, Δ_T_max would be about 0.012 and 0.12 °C. If the exposure duration were longer than 1 s, the temperature would continue to increase but at a progressively slower rate, until the rate of heat generation was about the same as the rate of heat removal.

To estimate the temperature increase from a single pulse for clinical, diagnostic pulse-echo instrumentation, the local, single pulse intensity of Eq. (45) is considered to be the spatial peak value averaged over the duration of the pulse, that is, the spatial peak, pulse average intensity, _I_SPPA. For typical instrumentation, a maximum value of _I_SPPA may be as high as 500 W/cm2. Thus, the maximum time rate of change of temperature is

but, with a diagnostic pulse duration, Δ_t_, of approximately 2 μs, the maximum temperature rise, Δ_T_max≈120 μ°C. However, in the case of high-intensity focused ultrasound (HIFU) for which the pulse duration may be as long as 3 s (ter Haar, 2004), Eq. (49) clearly demonstrates that the temperature increase could exceed levels sufficient to ablate tissue.

In living tissue, heat transfer occurs partly by perfusion (i.e., blood flow) and partly by conduction (also called diffusion). A widely used thermal model for tissue is the bioheat transfer equation (Pennes, 1948), a formulation that includes a source/sink term that accounts for heat transfer from blood perfusion, that is,

where δ_T_/δ_t_ is the rate of temperature increase at a point, κ is the thermal diffusivity, T is the temperature, Δ_T_ is the temperature increase above ambient, and τ is the perfusion time constant. Even though the bioheat transfer equation has shortcomings (Weinbaum et al., 1984), it is easy to implement analytically and is perhaps the most widely used thermal model of living tissue. Analyses based on the bioheat transfer equation (NCRP, 1983; Curley, 1993) were used to develop some of the ODS's thermal indices (ODS, 2004).

There have been several studies to calculate the temperature increase in mammalian tissue from ultrasonic exposure and some of them have shown to compare favorably with experimental results (Pond, 1968, 1970; Robinson and Lele, 1972; Lerner et al., 1973; NCRP, 1983; Nyborg and Steele, 1983; Cavicchi and O'Brien, 1985; AIUM, 1988, 1993, 2000; Smith et al., 1995). These studies demonstrate that selected aspects of the theory are reasonably well understood. But there are still many unanswered concerns in terms of being able to assess in vivo temperature increase.

6.1. Thermal dose concept

Healthy cellular activity depends upon chemical reactions occurring at the proper location at the proper rate. The rates of chemical reactions and thus of enzymatic activity are temperature dependent. The overall effect of temperature on enzymatic activity is described by the relationship known as the 10° temperature coefficient, or _Q_10 Rule (Hille, 2001). Many enzymatic reactions have a _Q_10 near 3 which means that for each 10 °C increase in temperature, enzymatic activity increases by a factor of 3; a more physical description of rate-dependent temperature effects is the Arrhenius activation energy concept (Henle, 1983; Sapareto and Dewey, 1984; Dewey, 1994; Dewhirst et al., 2003). An immediate consequence of a temperature increase is an increase in biochemical reaction rates. However, when the temperature becomes sufficiently high (i.e., approximately ≥45°C), enzymes denature. Subsequently, enzymatic activity decreases and ultimately ceases, which can have a significant impact on cell structure and function.

If damage occurs during exposure of tissue(s) to elevated temperature, the extent of damage will be dependent upon the duration of the exposure as well as on the temperature increase achieved. Detrimental effects in vitro are generally noted at temperatures of 39–43 °C, if maintained for a sufficient time period; at higher temperatures (>44 °C) coagulation of proteins can occur. These effects have been documented in experimental studies of heat-induced cell death in cultures of normal and cancerous cell lines. The lethal (100% destruction) dose (LD100) for HeLa cells exposed to different temperatures for differing durations has ranged from 41 °C for 96-h duration to 46 °C for 30-min duration (Selawry et al., 1957; Hornback, 1984). These findings are comparable to the time-temperature relationship to destroy 50% (LD50) of sarcoma-180 tumor cells in mice (Crile, 1961; Hornback, 1984); from 42 °C for 2-h duration to 46 °C for 7.5-min duration.

These observations suggest a logarithmic relationship between time and temperature for death due to a temperature increase. Dickson and Calderwood (1980) have indicated a similar relationship for time vs. temperature for thermal-induced death of tumors and normal animal and human tissues. Important points addressed in this study are: (1) at 40 °C long-duration exposures (5–100 h) are required for thermal-induced cell death, and (2) at temperatures appreciably below 40 °C there are no irreversible adverse effects detected.

An empirical formula, based on a large number of studies involving the thermotolerance of cells and tumors, relates the time, t (in min), required to produce an isoeffect (e.g., a given amount of cell killing) to the time (_t_43) which would be required had the exposure occurred at a reference temperature of 43 °C, that is,

where R = 0:5 for _T_>43 °C and R = 0.25 for T≤43 °C (Henle, 1983; Sapareto and Dewey, 1984; Dewey, 1994; Dewhirst et al., 2003). Theoretical considerations based on reaction kinetics (thermodynamic Arrhenius analyses) led to the prediction that the temperature dependence of the rate of protein denaturation is determined primarily by the activation energy. The quantity R is an expression of the relative increase in reaction rate for a 1 °C increase in temperature. The rationale for there being two “_R_” values is based upon the empirical determinations of R for a number of biological systems and endpoints (Dewey et al., 1977; Sapareto and Dewey, 1984, 1994; Dewhirst et al., 2003). In these systems, _R_-values ranged from 0.4 to 0.8, with 0.5 being the most common value, for temperatures above 43 °C. The few studies performed at temperatures ≤43 °C indicate that the _R_-value is approximately one-half of the value obtained at the higher temperatures.

By using Eq. (51), the empirical relationship derived by Sapareto and Dewey (1984), an equivalent _t_43 can be ascribed to any combination of temperature and exposure duration. It also follows that any given biological effect due to hyperthermia can be characterized by that _t_43 value of the causative exposure. The lowest _t_43 value giving rise to some effect would be considered the threshold.

Miller and Ziskin (1989) estimated that the _t_43 was greater than 1 min for each teratologic observation in their study (the lowest _t_43 for any effect was 1.9 min for the production of exencephaly in the mouse (Webster and Edwards, 1984)). Rearranging Eq. (51), and assuming that R = 0:25 (for temperatures ≤43 °C), yields

Miller and Ziskin (1989) used _t_43 = 1 min for fetal tissues, that is,

and the March 26, 1997 AIUM Conclusions Regarding Heat statement (AIUM, 1997) used Eq. (53) (_t_43 = 1 min) to indicate that there have been no significant, adverse biological effects observed due to temperature increases less than or equal to the line definded by this equation (see Fig. 7); the applicable exposure duration ranged between 1 and 250 min.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0007.jpg

Temperature–time curves for 4 values of _t_43 (see Eq. (54) for which R = 0.5 for _T_>43 °C and R = 0.25 for _T_≤43 °C). The bolded _t_43 = 1 min line shows the lower exposure duration range (applicable to 1 min) of the March 26, 1997 AIUM Conclusions Regarding Heat statement (AIUM, 1997).

For non-fetal tissues a range of _t_43 values has been reported. Results for breast (Lyng et al., 1991) and other tissues (Dewey, 1994) are summarized in Table 8. It should be noted that some of the data were garnered using animal models, whose core temperatures are higher than 37 °C, implying that the temperature increase necessary to achieve a particular thermal dose would be lower than would be the case with humans (Miller and Dewey, 2003; Herman and Harris, 2003). Adjustments in the _t_43 as applicable to humans might have to be made.

Table 8

_t_43 thermal dose values for various tissues (Lyng et al., 1991; Dewey, 1994).

Tissue Species _t_43 (min)
Muscle, fat Pig 240
Skin Human, rat, mouse 210
Esophagus Pig 120
Cartilage Rat, mouse 120
Breast Human 100
Bladder Dog, rabbit 80
Small intestine Rat, mouse 40
Colon Pig, rabbit 30
Liver Dog, rabbit 30
Brain Cat, dog 25
Kidney Mouse 20

More generally,

where t is the time (in min) corresponding to the threshold for a specific bioeffect which results from exposure to a temperature T (in °C). Also, R = 0.5 for T>43 °C and R = 0.25 for T≤43 °C. This equation explicitly states the relationship between temperature and exposure duration on the boundary line.

Fig. 7 shows the temperature–time curves for 4 values of _t_43 (see Eq. (54)). The lower curve (_t_43 = 1 min) represents that estimated for fetal tissues for _t_>1 min (Miller and Ziskin, 1989; AIUM, 1997). The other three curves, based on Table 8 _t_43 values (10, 100 and 240 min), represent non-fetal tissues that are less sensitive to tissue damage from temperature. Note that the temperature values for an exposure duration of 1 min are 43.0, 44.7, 46.3 and 47.0 °C for _t_43 values of 1, 10, 100 and 240 min, respectively. Based on the values in Table 8, the _t_43 = 1 min plot represents a conservative, tissue non-specific boundary for assessing thermal safety for non-fetal exposures.

For very short exposure times, the hyperthermia literature shows only limited (Borrelli et al., 1990) _t_43 thermal dose data points for exposure durations of less than 1 min. Two aspects of single-burst in vivo threshold lesion studies in brain (Fry et al., 1970; Dunn and Fry, 1971; Lerner et al., 1973) and liver (Chan and Frizzell, 1977; Frizzell et al., 1977; Frizzell, 1988) are germane to the thermal dose issue for exposures less than a few seconds. The threshold lesion curve for cat brain is described by the expression It_0.5 = 350 Ws0.5/cm2 over for exposure durations between 0.3 ms and 300 s. The threshold lesion curve for cat and rabbit liver is described by the expression It_0.5 = 460 Ws0.5/cm2 over for exposure durations between 3 ms and 35 s. I is the spatial peak intensity (in W/cm2) and t is exposure duration (in sec). Thus the first aspect is that liver has a higher threshold than brain, consistent with the t_43 thermal dose trend for brain and liver in Table 8. The second aspect is that for the brain threshold studies, an estimate was made of lesion temperature increase Δ_T, yielding, at 6 MHz, Δ_T/I estimates (interpolated from Fig. 4 in Lerner et al., 1973) of 0.086, 0.13 and 0.16 °Ccm2/W for pulse durations of 1, 10 and 100 s respectively. Combining these Δ_T/I estimates with _It_0.5 = 350 Ws0.5/cm2, and assuming a cat core temperature of 39 °C (NCRP, 1992), yields three temperature–time data points (open triangles on Fig. 8; also see Table 9).

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0008.jpg

Temperature–time curves (see Fig. 7) plus the following threshold data: filled-in circle, cat brain; open triangle: extrapolated cat brain; filled-in triangle, rabbit brain; filled-in square, rat brain; open diamond, rabbit muscle; open circle, dog prostate; open square, BHK cells, dashed line, multiple tissue thresholds; shaded line (just above bolded _t_43 = 1 min line), multiple in vitro thresholds. Details listed in Table 9.

Table 9

Temperature-time threshold-based data for various biological materials

Fig. 8 symbol Time (s) Temp (°C) Material Reference
Open triangle 1 69.1 Cat brain in vivo Lerner et al. (1973)
Open triangle 10 53.4 Cat brain in vivo Lerner et al. (1973)
Open triangle 100 44.6 Cat brain in vivo Lerner et al. (1973)
Filled-in circle 1.4 65 Cat brain in vivo Lele (1977)
Filled-in circle 1.8 64 Cat brain in vivo Lele (1977)
Filled-in circle 2.5 63 Cat brain in vivo Lele (1977)
Filled-in circle 3 65 Cat brain in vivo Lele (1977)
Filled-in triangle 10 53 Rabbit brain in vivo Vykhodtseva et al. (2000)
Filled-in triangle 30 48 Rabbit brain in vivo McDannold et al. (2004)
Filled-in triangle 30 47.8 Rabbit brain in vivo Chen et al. (2002)
Filled-in square 9 60.2 Rat brain in vivo Pond (1968)
Filled-in square 3 63.7 Rat brain in vivo Pond (1970)
Open diamond 30 47.2 Rabbit muscle in vivo McDannold et al. (2000)
Open diamond 30 47.5 Rabbit muscle in vivo Cheng et al. (2003)
Open circle 180 51 Dog prostate in vivo Peters et al. (2000)
Open square 1 57 BHK cells in vitro Borrelli et al. (1990)
Dashed linea 0.1 64.5 Multiple tissue thresholds Lele (1983)
Dashed lineb 770 41.5 Multiple tissue thresholds Lele (1983)
Shaded linea 60 46.2 Multiple in vitro thresholds Henle (1983)
Shaded lineb 840 42.9 Multiple in vitro thresholds Henle (1983)

In addition, there have been a number of documents that have reported threshold-based data for single-burst exposure durations as low as 100 ms (Table 9). These data are graphically shown in Fig. 8.

These data (Table 9) suggest that for non-fetal soft tissue, and for scanning conditions consistent with conventional B-mode ultrasound exams for which the exposure durations at the same in situ location would be less than a few seconds, the allowable maximum temperature increase could be relaxed relative to longer exposures.

6.2. Output display standard: thermal indices

A diagnostic ultrasound educational activity was initiated with a workshop in June, 1988. This workshop set out certain principles which resulted in the initiation of a 3-year process involving numerous clinicians, scientists, engineers and government regulators from many organizations; this group finalized and approved, in 1992, the Standard for Real-Time Display of Thermal and Mechanical Indices on Diagnostic Ultrasound Equipment, commonly referred to as the ODS (ODS, 1992). The purpose of this voluntary standard was to provide the capability for users of diagnostic ultrasound equipment to operate their systems at levels much higher than previously had been possible in order to have the potential for greater diagnostic capabilities; the standard did not specify any upper limits. In doing so, the possibility existed for the potential to do harm to the patient. Therefore, two biophysical indices were provided so that the equipment operator has real-time information available to make appropriate clinical decisions, viz., benefit vs. risk, and to implement the ALARA (As Low As Reasonable Achievable) principle (NCRP, 1990), that is, the ODS provides for a real-time output display which gives the user information about the potential for temperature increase (the Thermal Index (TI)) and mechanical damage (the Mechanical Index; see Section 7.1)

In the early 1990s, FDA implemented the ODS (ODS, 1992; FDA, 1993, 1994, 1997). While the ODS (ODS, 1992) did not specify upper limits, FDA's implementation (FDA, 1997) of the ODS stipulated regulatory upper limits of 720 mW/cm2 for the derated (0.3 dB/cm MHz) spatial peak, temporal average intensity, _I_SPTA.3, and either 1.9 for the Mechanical Index, MI or 190 mW/cm2 for the derated (0.3 dB/cm MHz) spatial peak, pulse average intensity, _I_SPPA.3. There is, however, an exception for ophthalmic application for which _I_SPTA.3≤50 mW/cm2 and MI≤0.23 (FDA, 1997). In addition, FDA (1997) requires the manufacturer to justify TIs greater than 6. The ODS has been revised (ODS, 1998, 2004), but FDA's regulatory upper limits have not changed.

The TI provides information about tissue temperature increase. This development addresses the thermal mechanism and closely related to the thermal mechanism is the TI. The basic TI definition is (ODS, 2004)

where _W_0 is the source power of the diagnostic ultrasound system and _W_DEG is the source power required to increase the tissue temperature 1 °C under very specific and conservative conditions. Three different Thermal Indices were developed to address three different tissue models and two different scan modes, that is, the soft-tissue Thermal Index (TIS), the bone Thermal Index (TIB), and the cranial-bone Thermal Index (TIC). The unscanned-mode is typically used clinically for spectral Doppler and M-Mode where the ultrasound beam remains stationary for a period of time. Also, the unscanned-mode soft-tissue TI, as well as TIB, are the only TI quantities that attempt to estimate temperature increase at locations other than at or near the source surface. The others estimate temperature increase at or near the source surface.

The following evaluation addresses the accuracy of the unscanned-mode TIS for circular (O'Brien and Ellis, 1999) and rectangular (O'Brien et al., 2004a) sources.

For the circular sources (O'Brien and Ellis, 1999), 3 source diameters (1, 2 and 4 cm) and 8 transmit f-numbers (ROC/source diameter = 0.7, 1.0, 1.3, 1.6, 2.0, 3.0, 4.0 and 5.0) were evaluated at 8 frequencies (1, 2, 3, 4, 5, 7, 9 and 12 MHz), yielding 192 cases. For the rectangular sources (O'Brien et al., 2004a), 33 one-dimensional focused rectangular aperture cases were investigated (Fig. 9) for which the aperture's _x_-length direction is the axis that is focused (_y_-length direction not focused) to an appropriate ROC to yield 3 _f_-numbers (ROC/_x_-length = 1, 2, and 4). Six frequencies (1, 3, 5, 7, 9 and 12 MHz) were evaluated (99/frequency), yielding 594 cases.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0009.jpg

Dimensions of the 33 rectangular aperture cases investigated.

For all 786 cases, the medium was assumed to be homogeneous in terms of both acoustic and thermal properties). The attenuation coefficient (also referred to as a derating factor) and absorption coefficient were both 0.3 dB/cm MHz, density was 1000 kg/m3, propagation speed was 1540 m/s, tissue perfusion length was 1 cm and tissue thermal conductivity was 6 mW/cm °C. These were the values used in the ODS (ODS, 1992, 1998, 2004) and the values used herein for the evaluation of the unscanned-mode TIS. They were also the values used to compute the maximum steady-state temperature increase (Δ_T_max) and its axial location. Also, all results reported herein are based on the derated spatial-peak, temporal-average intensity _I_SPTA.3 of 720 mW/cm2, FDA's (1997) regulatory limit.

Figs. ​10 and ​11 directly compare the 192 circular source Δ_T_max-TIS pairs as a function of frequency and f/number, respectively. TIS generally underestimates (is less than) Δ_T_max for f-numbers ≤2, conditions for which Δ_T_max≤0.30 °C and TIS≤0.40. This suggests that for transmit f-numbers ≤2, TIS would not need to be displayed according to the ODS display requirements. With the exception of the longer-focus, larger-diameter, higher-frequency sources, TIS generally tracks Δ_T_max for _f-numbers_≥3. These exceptions suggest a breakdown of the ODS procedures for calculating TIS.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0010.jpg

Paired maximum steady-state temperature increase Δ_T_max vs. unscanned-mode soft-tissue Thermal Index (TIS) for circular sources grouped by frequency under the condition that the derated (0.3 dB/cm MHz) spatial-peak, temporal-average intensity _I_SPTA.3 is 720 mW/cm2.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0011.jpg

Paired maximum steady-state temperature increase Δ_T_max vs. unscanned-mode soft-tissue Thermal Index (TIS) for circular sources grouped by _f_-number under the condition that the derated (0.3 dB/cm MHz) spatial-peak, temporal-average intensity _I_SPTA.3 is 720 mW/cm2.

Figs. ​12 and ​13 directly compare the 594 rectangular source Δ_T_max-TIS pairs as a function of frequency and f/number, respectively. TIS overestimates (is greater than) Δ_T_max for all but one of the cases.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0012.jpg

Paired maximum steady-state temperature increase Δ_T_max vs. unscanned-mode soft-tissue Thermal Index (TIS) for rectangular sources grouped by frequency under the condition that the derated (0.3 dB/cm MHz) spatial-peak, temporal-average intensity _I_SPTA.3 is 720 mW/cm2.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0013.jpg

Paired maximum steady-state temperature increase Δ_T_max vs. unscanned-mode soft-tissue Thermal Index (TIS) for rectangular sources grouped by _f_-number under the condition that the derated (0.3 dB/cm MHz) spatial-peak, temporal-average intensity _I_SPTA.3 is 720 mW/cm2.

The ODS process does not specify the location of Δ_T_max. Figs. ​14 and ​15 show the relationship between Δ_T_max and its location (axial distance) for the circular and rectangular sources, respectively, as a function of _f_-number (the same 3 _f_-numbers are shown for direct comparison). And, Figs. ​16 and ​17 show the relationship between ROC and its axial distance locations of Δ_T_max for the circular and rectangular sources, respectively, as a function of _f_-number. For all 786 cases, the axial distance locations of Δ_T_max are less than the ROC, the geometric focus location. As frequency increases for the lower _f_-number cases, the axial distance locations of Δ_T_max move closer to the geometric focus. However, for the higher _f_-number cases in each frequency set for _f_≥3 MHz, the axial distance locations of Δ_T_max jumps to near the transducer surface. This behavior, also reported by Thomenius (1990), becomes more prevalent at higher frequencies and larger source diameters. Thus, the Δ_T_max location is not necessary near the skin surface.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0014.jpg

Paired maximum steady-state temperature increase Δ_T_max vs. location (axial distance) of Δ_T_max for circular sources grouped by _f_-number under the condition that the derated (0.3 dB/cm MHz) spatial-peak, temporal-average intensity _I_SPTA.3 is 720 mW/cm2.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0015.jpg

Paired maximum steady-state temperature increase Δ_T_max vs. location (axial distance) of Δ_T_max for rectangular sources grouped by _f_-number under the condition that the derated (0.3 dB/cm MHz) spatial-peak, temporal-average intensity _I_SPTA.3 is 720 mW/cm2.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0016.jpg

Paired location of the geometric focus (ROC) vs. normalized location (axial distance) of Δ_T_max (location of Δ_T_max/ROC) for circular sources grouped by _f_-number.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0017.jpg

Paired location of the geometric focus (ROC) vs. normalized location (axial distance) of Δ_T_max (location of Δ_T_max/ROC) for rectangular sources grouped by _f_-number.

In scanned modes, however, the focal energy is distributed over a large area and hence will tend to have lower spatial intensities than unscanned modes. With the recent introduction of real-time three-dimensional scanners, the focal energy will be distributed over two dimensions that should result in further reduction in relevant focal intensities. However, there will be an added concern of near-field heating. Near-field heating could be exacerbated by transducer self-heating (Duck et al., 1989). At this point it is believed that the development of a new index to cover these cases is not indicated in view of the regulatory mechanisms that are already in place for transducer surface temperatures.

The ODS process does not take into account the temporal dependency of the temperature increase. An evaluation of the temporal dependency of temperature increase was conducted for the circular sources of three source diameters (1, 2 and 4 cm) at two frequencies (2 and 7 MHz) for five f-numbers (f/1, f/2, f/3, f/4 and f/5); Fig. 18 shows _t_80%, the time when the temperature increase reaches 80% of its steady-state value at that axial distance, for two source diameters (1 and 4 cm). Each _t_80% profile follows the same general pattern as a function of axial distance. For each of the 7-MHz cases, the axial distances of the minimum _t_80% values are, in general, near the respective focal regions. However, for the 2-MHz cases, the axial distances of the minimum _t_80% values tend not to be near the respective focal regions but rather closer to the transducer. Also, the minimum _t_80% values increase as a function of increasing f-number, and decrease as a function of increasing frequency. The global minimum _t_80% values are 97.2 and 59.7 s for the fifteen 2-MHz and fifteen 7-MHz cases, respectively, both of which are for the 1-cm-diameter, f/1 cases.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0018.jpg

_t_80% profiles as a function of axial distance for the 2-MHz (gray) and 7-MHz (black) cases for circular sources under the condition that the derated (0.3 dB/cm MHz) spatial-peak, temporal-average intensity _I_SPTA.3 is 720 mW/cm2.

6.3. Output display standard: possibly better thermal indices

The TIs in use today were developed more than 14 years ago (ODS, 1992). While the ODS has been revised twice (ODS, 1998, 2004), the TI equations have not changed. Figs. ​10-​13 indicate that there could be improvements. There has been one effort to improve the unscanned-mode TIS. The database used was the same as that used to develop for the 594 rectangular source cases (O'Brien et al., 2004a), and yielded two new TIS expressions:

TISnew(1)=W00.85fc0.58169⋅Aaprt0.33,

56

TISnew(2)=W00.73fc0.62130,

57

where _W_0 is the source power in millwatts (in mW), _f_c is the center frequency in megahertz (in MHz) and _A_aprt is the active aperture area (in cm2) (see Fig. 19). In both models, only the source power and frequency need to be measured; they can adequately be measured (AIUM/NEMA, 1998). The only difference between TISnew(1) and TISnew(2) is the degree of agreement with Δ_T_max that might be required.

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0019.jpg

Paired maximum steady-state temperature increase Δ_T_max vs. the proposed unscanned-mode soft-tissue Thermal Indices (TISnew(1) and TISnew(2)) for rectangular sources grouped by _f_-number under the condition that the derated (0.3 dB/cm MHz) spatial-peak, temporal-average intensity _I_SPTA.3 is 720 mW/cm2.

7. Non-thermal mechanisms

Both first- and second-order quantities have been implicated in non-thermally produced biological effects (NCRP, 2002). The non-thermal mechanism that has received the most attention is acoustically generated cavitation, principally from ultrasound contrast agent microbubbles. Excellent reviews of cavitation have been published (Flynn, 1964, 1975a, b, 1982; Nyborg, 1965, 1975; Coakley and Nyborg, 1978; Apfel, 1981b; NCRP, 1983, 1992, 2002; Leighton, 1994; ter Haar and Duck, 2000; AIUM, 2000; Suslick, 2001; Hoff, 2001). Effects have been also attributed to radiation force. There is also a class of ultrasound-induced biological effects that have not yet been determined to be due to either thermal or a specific non-thermal mechanism.

7.1. Output display standard: mechanical index

The Mechanical Index, MI, is defined as

where _p_r.3 is the derated peak rarefactional pressure in MPa (megapascals) and f is the ultrasonic frequency in MHz. The derating factor (attenuation coefficient of soft tissue) is assumed to be 0.3 dB cm MHz. The development of the Mechanical Index was based on theoretical and in vitro experimentation by investigators (Holland and Apfel, 1989; Apfel and Holland, 1991) who discovered a simple relationship between acoustic pressure and the onset of cavitation under an assumption that the optimum bubble size is present. The theory assumed isothermal growth, adiabatic collapse, an incompressible host fluid and neglected gas diffusion into the bubble, and the experiments were conducted in an aqueous medium, not tissue. These observations were the basis for the adoption of the MI (ODS, 1992).

7.2. Radiation force

Within recent times, a number of ultrasound diagnostic techniques have emerged that use the static or dynamic acoustic radiation force to locally move and/or vibrate tissue including acoustic radiation force impulse imaging (Nightingale et al., 1995, 2001), vibro-acoustography (Fatemi and Greenleaf, 1998, 1999), supersonic shear imaging (Bercoff et al., 2004). Generally, these techniques utilize the ultrasound-induced temporal-average force on the medium (Section 3.2) to elicit a biophysical effect (tissue motion). The magnitude of the effect is proportional to the local temporal-average intensity. Whether there are other biophysical effects (risk-related temporary or permenant tissue responses) have not been extensively studied.

However, the ultrasound-induced temporal-average force has been implicated as a mechanism associated with tactile response (Gavrilov et al., 1977a, b; Gavrilov, 1984; Magee and Davies, 1993; Dalecki et al., 1995), auditory response (Foster and Wiederhold, 1978; Tsirulnikov et al., 1988), increased fetal activity (Arulkumaran et al., 1991, 1996; Fatemi et al., 2001), bone repair (Dyson and Brookes, 1983; Wang et al., 1994, 2001), cardiac changes in frogs (Dalecki et al., 1993, 1997a), movement of detached retinas (Lizzi et al., 1978) and macroscopic streaming to differentiate between cystic and solid tumors (Stavros and Dennis, 1993; Nightingale et al., 1995). Other than those responses related to macrostreaming (Nyborg, 1953, 1965), there is a limited association, possibly only speculative, between the response and radiation force.

7.3. Cavitation without injected microbubbles

When an ultrasound wave propagates in tissue, a mechanical strain is induced, where strain refers to the relative change in dimensions or shape of the body that is subjected to stress. The strain is significant near gas or vapor bubbles, hence the interest in ultrasound-induced cavitation. Cavitation, in a broad sense, refers to ultrasonically induced activity occurring in a liquid or liquid-like material that contains bubbles or pockets containing gas or vapor. These bubbles originate within materials at locations termed “nucleation sites,” the exact nature and source of which are not well understood in a complex medium such as tissue. Under ultrasonic stimulation, the bubble oscillates and at sufficiently high ultrasonic pressure levels, the bubble can collapse. Collapse cavitation is termed “inertial collapse” because the bubble motion is dominated by the inertia of the liquid. There are some excellent treatments of bubble dynamics and inertial cavitation (Gilmore, 1952; Flynn, 1975a, b, 1982; Keller and Miksis, 1980; Apfel, 1981b; Flynn and Church, 1988; Matsumoto and Watanabe, 1989; Prosperetti and Lezzi, 1986; AIUM, 2000; NCRP, 2002; Church 2002, 2005). Cavitation can affect a biological system by virtue of a temperature increase, a mechanical stress, and/or free radical production. Even so, this is traditionally referred to as a non-thermal mechanism.

The occurrence of cavitation, and its behavior, depends on many factors, including: the ultrasonic pressure; whether the ultrasonic field is focused or unfocused, or pulsed or continuous; to what degree there are standing waves (i.e., energy reflecting back onto itself); and the nature and state of the material and its boundaries. Experimentally, because cavitation would probably affect only a single or a few cells, it would be extremely difficult to detect an adverse biological effect, unless the cavitation events were widespread among a large volume of tissue. The latter has been shown to be the case when mammalian nervous system tissue was exposed to ultrasonic levels in excess of _I_SPTP of 1000 W/cm2 for a pulse duration of at least 1 ms (Dunn and Fry, 1971; Frizzell, 1988); these conditions are outside of the diagnostic equipment range.

Theoretical predictions suggested that inertial cavitation could occur in liquids by diagnostic ultrasound exposures provided that appropriate cavitation nuclei were present (Apfel, 1982, 1986; Flynn, 1982; Apfel and Holland, 1991). The minimum threshold for inertial cavitation was calculated for the diagnostically relevant frequency range (Apfel and Holland, 1991), and this theory served as the basis for the ODS's Mechanical Index (ODS, 1992). A theoretical prediction considered the possibility of spontaneous nucleation in tissue modeled as a homogeneous liquid, in the absence of preexisting nuclei, wherein a frequency-independent (1–15 MHz) rarefactional pressure threshold was on the order of 4 MPa (Church, 2002). However, soft tissue is a viscoelastic material, not a pure liquid, and recent theoretical predictions suggest that inertial cavitation thresholds are higher for soft tissue than liquids (Yang and Church, 2005). Thus, there has been no experimental evidence to suggest that inertial cavitation events occur in mammalian parenchymal tissue from exposure-like conditions employed with diagnostic ultrasound equipment, that is, endogenous cavitation nuclei are rare in most tissues.

7.4. Cavitation with injected microbubbles

The principal medical application of ultrasound contrast agents (UCAs) is to enhance the echogenicity of blood. Such ultrasound contrast agent microbubbles provide preexisting nuclei and thus the possibility for ultrasound-induced cavitation and potential for bioeffects. In the United States, Albunex® (Molecular Biosystems, Inc., San Diego, CA) was the first FDA-approved UCA in 1994. An improved version of Albunex®, Optison™ (then MBI, San Diego, CA), was FDA approved in 1997. The only other 2 FDA-approved UCAs are Definity® (Bristol-Meyers Squibb, NY, NY) in 2001 and Imagent® (IMCOR, San Diego, CA) in 2002. Other UCAs have been approved in Canada and/or Europe (Miller and Nanda, 2004), but since 2002 there have been no new UCAs approved by FDA. While there are no written documents, it appears that FDA is uncertain about UCAs' safety and appears to be waiting until more is known about the device risk of these agents.

Three ultrasound exposure modes are used with UCAs, viz., low MI, normal MI and high MI, where MI is the Mechanical Index (ODS, 2004), to create 3 imaging phenomena (Miller and Nanda, 2004). Low MI is considered around 0.1 and the UCAs undergo linear oscillations. Normal MI is considered between 0.2 and 0.7 and the UCAs undergo nonlinear oscillations. High MI is considered between 0.8 and 1.9 [FDA's regulatory limit (FDA, 1997)] and the UCAs undergo destruction. For normal MI and high MI, it is thought that the UCA signals are bubble-specific.

It is unfortunate that the MI has become the UCA exposure quantity, but the MI is an exposure quantity available to system operators. The concept is well founded, but the science is weak because the MI is a water-based measurement, not an in situ measure of ultrasonic pressure. For the same displayed MI, the in situ ultrasonic pressure varies as a function of the interposed tissue attenuation coefficient and thickness. Also, what ultrasonic pressures (thresholds) are necessary to collapse UCAs in vivo are unknown.

Using a passive cavitation detector, UCA rupture based on ultrasonic emissions from single Optison™ UCAs have been measured (Fig. 20) (Ammi et al., 2004, 2006; Wang et al., 2006). Ultrasonic pulses were transmitted into weak solutions of Optison™. Signals detected with a 13-MHz center-frequency transducer revealed post-excitation ultrasonic emissions (between 1 and 5 ms after excitation) with broadband spectral content. The observed ultrasonic emissions were consistent with the US signature that would be anticipated from inertial collapse followed by “rebounds” when an UCA ruptures and thus generates daughter/free bubbles that grow and collapse. The peak rarefactional pressure threshold for detection of these emissions increased with frequency (p<0.0001), but the apparent threshold decrease with increasing pulse duration was not statistically different. The estimated MI thresholds at the 4 frequencies of 0.9, 2.8, 4.6 and 7.1 MHz (Fig. 20) were 0.6, 0.7, 0.7 and 0.9, respectively, well within the diagnostic ultrasound exposure levels (FDA, 1997).

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0020.jpg

Peak rarefactional pressure rupture threshold of Optison™ as a function of frequency and pulse duration. PRF = 10 Hz.

Concerns about the potential bioeffects of inertial cavitation associated with the interaction of ultrasound with contrast agents in human beings have been addressed (AIUM, 2000; NCRP, 2002). These concerns were raised by studies documenting hemolysis of erythrocytes in vitro in cell suspensions that contained contrast agent and in vivo in mice injected with intravenous contrast agent that were exposed to pulsed ultrasound (Williams et al., 1991; Dalecki et al., 1997b; Miller et al., 1997; Miller and Gies, 1998a, b; Poliachik et al., 1999; Brayman and Miller, 1999). In vitro studies have reported damage to monolayers of cultured cells whose culture media contained contrast agent and were exposed to pulsed ultrasound (Brayman et al., 1999a; Ward et al., 1999; Miller and Bao, 1998; Miller and Quddus, 2000a). Hemorrhage in the vascular beds of the intestine and skin (Miller and Quddus, 2000b; Miller and Gies, 2000) and damage to cells in the heart (Skyba et al., 1998) were also demonstrated in mice and dogs, respectively, following intravenous injection of contrast agent and exposure to pulsed ultrasound. In vivo studies have reported induction of petechiae and hemolysis (Killam et al., 1998; Skyba et al., 1998; Miller and Quddus, 2000b; Wible et al., 2002; Hwang et al. 2005), damage to the intestinal wall (Miller and Gies, 1998a, b, 2000; Kobayashi et al., 2002, 2003), and alteration to the blood-brain barrier (Schlachetzki et al., 2002; Hynynen et al., 2003).

More medically significant have been the reports of arrhythmogenic changes that included premature ventricular contractions in healthy adult human beings during triggered second harmonic imaging of a UCA for myocardial perfusion (van Der Wouw et al., 2000); supraventricular tachycardia or non-sustained ventricular tachycardia in patients at risk for syncope, supraventricular tachycardia, or ventricular tachycardia after intravenous administration of UCA and exposure to therapeutic transthoracic low-frequency ultrasound (Chapman et al., 2005); and cardiac arrhythmogenesis in a rat model (Zachary et al., 2002). Cavitation has been suggested as the likely mechanism for microbubble-induced premature cardiac contractions (Dalecki et al. 2005).

In addition, there are also a number of possible therapeutic benefits of UCAs. A significant problem in cancer chemotherapy is the compromised quality of life experienced by the patient due to the side effects of these drugs. Delivery of drugs directly and specifically to solid tumors is often inefficient, and as a result, the patient's healthy cells and tissues are subject to the toxic effects of the drugs. Thus, it is important to develop therapeutic strategies that deliver drugs to the appropriate cells within the patient in a way that is specific, efficient, and safe. One such method involves the use of US to enhance cell permeability. With this method, it is possible via the vasculature of the tumor, to deliver therapeutic compounds placed within UCAs directly to tumor cells when the tumor is insonated with US and the UCAs are ruptured. In addition, when exposed to US, the membrane of a cell can become transiently permeable. This effect, termed sonoporation, is enhanced when UCAs undergo cavitation near the cells. Small compounds (Brayman et al., 1999b; Guzman et al., 2001; Kayhani et al., 2001), macromolecules (Guzman et al., 2002; Miller et al., 1999), DNA (Bao et al., 1997; Greenleaf et al., 1998; Wyber et al., 1997; Amabile et al., 2001; Lawrie et al., 2000; Miller et al., 2003; Miller and Song 2003), protein (Mukherjee et al., 2000; Weimann and Wu, 2002), and other therapeutic compounds (Harrison et al., 1996; Kayhani et al., 2001; van Wamel et al., 2004) have been delivered into cells using US. This membrane alteration can be transient (McNeil, 1989), leaving the compounds trapped inside the cell after US exposure. Thus, sonoporation is a viable delivery method, ensuring site-specific drug delivery or transfection.

Even though this section deals with non-thermal mechanisms, UCAs can have an effect on bulk tissue heating (Hilgenfeldt et al., 1998, 2000; Chavrier and Chapelon, 2000; Holt and Roy, 2001; Sokka et al., 2003; Umemura et al., 2005). Typically, there is at least a 2–4 times enhancement of tissue heating by cavitation, or, if the bioeffect were a lesion, the lesion volume was likewise enhanced. Also, single bubbles undergoing inertial collapse can cause plasma formation and temperature elevation (≈4300–5000 K) sufficient to induce thermal injury (Didenko et al., 1999; Suslick, 2001). Additionally, such high temperatures in an aqueous medium may result in the formation of chemically reactive free radicals (Verral and Sehgal, 1988) that can cause damage.

7.5. Ultrasound-induced lung hemorrhage

One of the most extensively studied ultrasound-induced biological effects is lung hemorrhage. However, the mechanism is not clear. While heating resulting from the absorption of ultrasound can cause tissue injury, heating has been experimentally excluded as the mechanism responsible for ultrasound-induced lung hemorrhage via thermocouple measurements (Hartman et al., 1992). A thermal mechanism has also been excluded via a pathological evaluation of laser-induced lung lesions (Zachary et al., 2006). Also, inertial caviation has been excluded as the mechanism using an overpressure procedure (O'Brien et al., 2000) and the injection of contrast agents (O'Brien et al., 2004b). Some general observations can be made, however, from the extensive number of ultrasound-induced lung damage threshold studies.

Fig. 21 shows all of occurrence threshold findings across four species using the same methodology [mouse (Zachary et al., 2001a), rat (Zachary et al., 2001a; O'Brien et al., 2001, 2003b), rabbit (O'Brien et al., 2006), and pig (O'Brien et al., 2003a)]. Pre-2000 occurrence threshold findings (Fig. 22) across three species [mouse (Child et al., 1990; Raeman et al., 1993, 1996; Frizzell et al., 1994; Dalecki et al., 1997d), rat (Holland et al., 1996), pig (Baggs et al., 1996; Dalecki et al., 1997c)] have also been reported (AIUM, 2000) for which experimental techniques and statistical approaches were different from the statistical analysis procedure (Simpson et al., 2004) used with Fig. 21 experiments. A principal difference between the two sets of studies lies in the exposure duration wherein Fig. 21 threshold findings used a 10-s ED while the pre-2000 threshold findings (Fig. 22) used typically longer EDs. The 10-s ED yielded occurrence thresholds typically higher (median: 3.0 MPa; range: 1.2–5.8 MPa) than the pre-2000 thresholds (median: 1.0 MPa; range: 0.4–2.5 MPa).

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0021.jpg

Summary of ED05 (5%) lesion occurrence thresholds in terms of _p_r(in situ) (MPa) (bars; left axis) and MI (lines, right axis). *denotes the studies that are evaluated under the same exposure conditions. Error bars are SEMs. These data, by groupings from left to right, are derived, respectively, from Zachary et al. (2001a) and O'Brien et al. (2001, 2003a, b, 2006).

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0022.jpg

Summary of lesion occurrence thresholds in terms of _p_r(in situ) (MPa) (bars; left axis) and MI (lines, right axis) from studies not using our experimental techniques and statistical approach. These data, by groupings from left to right, are estimated, respectively, from Child et al. (1990), Raeman et al. (1993, 1996), Frizzell et al. (1994), Holland et al. (1996), Baggs et al. (1996) and Dalecki et al. (1997c, d).

All of these occurrence threshold values (from Figs. ​21 and ​22) have been organized as a function of frequency (Fig. 23 relative to ultrasonic pressure and Fig. 24 relative to Mechanical Index). Virtually every ultrasound-induced lung hemorrhage threshold is less than the FDA regulatory limit (MI<1.9, FDA 1997), that is, within the diagnostic ultrasound frequency range. These thresholds are over a wide range of animal weights [mouse: 22–32 g, rat: 190–330 g, rabbit: 1.9–2.9 kg, and pig: 1–25 kg] and chest wall thicknesses [mouse: 1.9–4.9 mm, rat: 4.1–6.0 mm, rabbit: 5.0–6.6 mm, and pig: 2–22 mm].

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0023.jpg

Global summary of lesion occurrence thresholds in terms of in situ peak rarefactional pressure as a function of frequency for four species (see Figs. ​21 and ​22). The solid line is the threshold equation derived from pre-2000 data (AIUM, 2000).

An external file that holds a picture, illustration, etc. Object name is nihms-17904-0024.jpg

Global summary of lesion occurrence thresholds in terms of the Mechanical Index as a function of frequency for four species (see Figs. ​21 and ​22). The solid line denotes an MI of 1.9, the FDA regulatory limit (FDA, 1997).

The results of the studies using mice, rats, rabbits and pigs have shown two important facts relative to the biological mechanisms of damage: (1) there are no differences in biological mechanism of injury induced by ultrasound based on species and age; therefore, structural differences among mammalian species studied are independent of the biological mechanism that causes ultrasound-induced lung damage and (2) lesions induced by ultrasound are similar in morphology in all species and age groups studied, and the character of the lesions is independent of frequency, PRF, and beamwidth. These findings suggest that the mechanism of injury is similar in all species and age groups. It is thus plausible to speculate that the same biological mechanism causing ultrasound-induced lung hemorrhage in laboratory animals may be in play in human beings undergoing sonographic procedures with direct or incidental exposure of lung, especially patients with pulmonary disorders and “at risk” neonates. However, the extent of ultrasound-induced lung damage, and the ability of the lung to heal (Zachary et al., 2001b), lead to the conclusion that while diagnostic ultrasound can produce lung damage at clinical levels, the degree of damage does not appear to be a significant medical problem.